Simultaneous Atomic Force Microscope and Quartz Crystal Microbalance Measurements:

Interactions and Displacement Field of a Quartz Crystal Microbalance

J.-M. FRIEDT, K. H. CHOI, L. FRANCISand A. CAMPITELLI IMEC, Kapeldreef 75, 3001 Leuven, Belgium

(Received January 22, 2002; accepted for publication February 25, 2002)

We analyze the interaction of two instruments often used in material science analysis, the atomic force microscope (AFM) and the quartz crystal microbalance (QCM), here combined in a single instrument for simultaneous measurements on a single sample. We show, using finite element analysis, that the in-plane displacement of a QCM oscillating in liquid with a quality factor of 2000 is 2 nm. The out-of-plane displacement is about one tenth of the in-plane displacement. This latter effect, due to the finite size of the electrodes, results in longitudinal acoustic waves launched in the liquid surrounding the QCM. If bounced against an obstacle, in our case the AFM cantilever holder, these longitudinal waves create standing wave patterns which cause frequency fluctuations of the resonator when it is moved, and thus decrease the QCM sensitivity.

[DOI: 10.1143/JJAP.41.3974]

KEYWORDS: QCM, displacement, resonance frequency, quality factor, AFM

1. Introduction

During the development of a combined quartz crystal microbalance1)(QCM) — atomic force microscope2)(AFM) instrument, we faced the challenge of estimating the displacement of the QCM resonator parallel to the electrode plane (which defines the lateral resolution of our AFM imaging) and normal to the electrode plane (which defines the topography resolution of the AFM scans).

Simultaneous AFM and QCM measurements of a sample allow the following:

. differents kinds of data to be gathered on different scales (cm2 scale in the case of the QCM, nm2scale in the case of the AFM) and of a very different nature (mass change/viscosity losses in the case of the QCM, topography/stiffness in the case of the AFM). These measurements are performed on a single sample and thus it is guaranteed that the same conditions are met during the data gathering, which might not be evident for successive experiments with various instruments.

. checking of the validity and assumptions in converting frequency shifts and quality factor (damping) variations as measured by the QCM to mass changes (using the Sauerbrey equation).

Although the combination of several measurement techniques has shown promising progress by obtaining complementary results,3) simultaneous measurements using AFM and QCM have not yet been performed to our knowledge.

We focused on precisely modeling the finite counter electrode size in order to accurately predict the out-of-plane displacement which also leads to frequency fluctuations when an obstacle (AFM cantilever holder) is brought close to the QCM surface because of the acoustic longitudinal waves generated in the liquid medium.4)

2. Simulation Tools

We used the open-source software Modulef,5) latest release as of September 2001, developed by INRIA (France) for simulating a cylindrical AT-cut quartz crystal resonator patterned with finite sized electrodes.

The piezoelectric properties of quartz were defined using the tabulated values found in ref. 6 and rotated to fit the AT

cut using the formula provided by ref. 7.

3. Static Simulations

Our first approach is to apply a DC voltage between the electrodes deposited on opposite surfaces of the quartz resonator. We monitor the spatial distribution of the displacement field while keeping the position of the sensing electrode fixed.

We simulate a 14-mm-diameter quartz resonator 400m thick, fitted with a 5-mm-diameter massless counter electrode on which the DC potential (0.5 V, Fig.2 left) is applied while the sensing electrode completely covers the top side of the QCM and is grounded. The total number of points used for this 3D simulation is 3320, which define 2040 pentaedral finite elements and 1280 hexaedral elements (Fig.1). The results of this simulation are as follows:

. the maximum static displacement in theXdirection is 1.075 pm, Fig.3, left,

. the maximum static displacement in theY direction is 0.126 pm, Fig.3, right, and

. the maximum static displacement in theZ direction is 0.102pm, Fig.2, right.

As expected from an analytical resolution for an infinite plate, the displacement field is mainly oriented in theXaxis direction (Fig.1, right).

Inside the volume defined by a vertical cylinder above the counter electrode, the displacement is mainly parallel to the Xdirection, while discrepencies in this behaviour appear on the frontier of the counter electrode. These results for theX direction are in close agreement with previous analytical results presented in refs. 8–10 and they bring new insight to the Z displacement of the QCM which cannot be analyzed using an analytical formula which assumes an infinite electrode surface. The displacement normal to the electrode plan, about one tenth of the in-plane displacement, is much larger than previously estimated9)and explains the origin of the large component of longitudinal waves generated by the QCM in the surrounding liquid.

4. Dynamic Simulations

With the intention of achieving conditions closer to the experimental conditions, a simulation of the dynamic oscillation of the quartz resonator should be performed in

Jpn. J. Appl. Phys. Vol. 41 (2002) pp. 3974–3977 Part 1, No. 6A, June 2002

#2002 The Japan Society of Applied Physics

3974

(2)

order to directly obtain the displacement field enhanced by the mechanical resonance. This approach presents two major difficulties:

. in a simulation without damping, the displacement field for any frequency other than the resonance frequency is the same as in the DC mode, and the displacement field is expected to diverge when the resonance frequency is met.

. when including the damping, the losses that are dependent on the experimental conditions and the imaginary partij(i;j2 ½1;6) of the stiffness constants (i.e., the damping coefficient) are extremely difficult to estimate. Using the relation7) ij¼Cij=ðQij!0Þ, where

Qij¼Qis the quality coefficient (estimated to be equal to the experimentally observed quality factorQfor alli andj),!0 the mechanical resonance pulsation and Cij the real part of the mechanical stiffness constants, leads to a conclusion similar to enhancing the static displacement field by a factor Q9) for estimating the dynamic displacement field.

Hence, we have not focused on obtaining additional results using dynamic mode simulations which are not expected to display any new results compared to the static mode simulation, and we now extend the results obtained from simulations with a DC electrical field to the dynamic oscillation by introducing the quality factorQ.

O

X Y

Z

2862 NODES 9684 FACES 2040 PENTAEDRA 1280 HEXAEDRA 20 4.9750E-04 19 4.7367E-04 18 4.4735E-04 17 4.2102E-04 16 3.9469E-04 15 3.6837E-04 14 3.4204E-04 13 3.1571E-04 12 2.8939E-04 11 2.6306E-04 10 2.3673E-04 9 2.1041E-04 8 1.8408E-04 7 1.5775E-04 6 1.3143E-04 5 1.0510E-04 4 7.8774E-05 3 5.2448E-05 2 2.6121E-05

1 -2.0539E-07 O

X Y

Z

2862 NODES 9684 FACES 2040 PENTAEDRA 1280 HEXAEDRA 20 1.0083E-07 19 9.1142E-08 18 8.0432E-08 17 6.9721E-08 16 5.9011E-08 15 4.8301E-08 14 3.7590E-08 13 2.6880E-08 12 1.6170E-08 11 5.4592E-09 10 -5.2511E-09 9 -1.5961E-08 8 -2.6672E-08 7 -3.7382E-08 6 -4.8092E-08 5 -5.8803E-08 4 -6.9513E-08 3 -8.0223E-08 2 -9.0934E-08 1 -1.0164E-07

Fig. 2. Display of the DC potential (voltage, in kV) distribution (equipotential lines range from 0.5 V to0:2mV) and the amplitude of the out-of-plane (Z) displacement (equidisplacement lines range from 0.1 pm to0:1pm).

O

X Y

Z

O

X Y

Z

Fig. 1. Final 3D structure after defining the mesh (left), and the displacement field after resolution of the problem (right).

O

X Y

Z

2862 NODES 9684 FACES 2040 PENTAEDRA 1280 HEXAEDRA 20 1.0698E-06 19 1.0177E-06 18 9.6006E-07 17 9.0247E-07 16 8.4488E-07 15 7.8728E-07 14 7.2969E-07 13 6.7210E-07 12 6.1451E-07 11 5.5692E-07 10 4.9932E-07 9 4.4173E-07 8 3.8414E-07 7 3.2655E-07 6 2.6896E-07 5 2.1136E-07 4 1.5377E-07 3 9.6180E-08 2 3.8588E-08 1 -1.9003E-08

O

X Y

Z

2862 NODES 9684 FACES 2040 PENTAEDRA 1280 HEXAEDRA 20 1.2457E-07 19 1.1232E-07 18 9.8787E-08 17 8.5251E-08 16 7.1715E-08 15 5.8179E-08 14 4.4643E-08 13 3.1107E-08 12 1.7571E-08 11 4.0351E-09 10 -9.5009E-09 9 -2.3037E-08 8 -3.6573E-08 7 -5.0109E-08 6 -6.3645E-08 5 -7.7181E-08 4 -9.0717E-08 3 -1.0425E-07 2 -1.1779E-07 1 -1.3133E-07

Fig. 3. Display of the amplitude of the displacement in the two in-plane directions (main displacement axesXandY). Lines of equal displacement are displayed inm.

Jpn. J. Appl. Phys. Vol. 41 (2002) Pt. 1, No. 6A J.-M. FRIEDTet al. 3975

(3)

Following these conclusions, we extend the static results to the dynamic behaviour expected to occur at resonance by using9) the formulaAdynamic¼AstaticQ, where Adynamic is the oscillation amplitude at resonance and Astatic is the previously computed DC displacement. Since we observe that our QCMs display quality factors in the 1000 to 3000 range when oscillating in liquid at the resonance frequency (4.7 MHz), the results are as follows (for Q¼2000):

. the maximum dynamic displacement in theXdirection is 2.1 nm,

. the maximum dynamic displacement in theYdirection is 0.25 nm, and

. the maximum dynamic displacement in theZdirection is 0.20 nm.

These results are in agreement, once normalized for the quality factor, with the measurements obtained by Borovsky et al. using an STM under vacuum.11)

5. Experimental QCM Frequency Fluctuations Mea- surements

We have designed a combined QCM-AFM instrument (Fig.4) based on the scanning probe microscope developed by molecular imaging (Phoenix, USA), with the QCM probing electronics provided by Q-Sense AB (Go¨teborg,

Sweden) and laboratory built quartz crystal resonators designed to oscillate at around 4.7 MHz by coating a 14- mm-diameter AT-cut quartz wafer with 50 nm Ti and 50 nm Au. The output amplitude of the Q-Sense instrument was measured to be in the 200 mVpp range.

We have experimentally (Fig.5) observed in the com- bined QCM-AFM instrument the frequency drift of the QCM due to the AFM cantilever holder approach, as well as frequency fluctuations of the quartz crystal resonator during image scans (11m2). We have observed that the first oscillation mode of the QCM (at 4.7 MHz) is the most sensitive to these environmental changes, as expected on the basis of its wider spread over the sensing electrode surface.12)

Figures5and6display the influence of the position of the cantilever holder on the standing wave pattern generated between the QCM and cantilever holder surfaces, in the first case during the tip approach and in the second case during a protein adsorption experiment. The first mode appears to be much more sensitive to the disturbances due to environ- mental variations than the higher order modes (the third and fifth overtones were also measured during this study). As expected,4) the peaks for the third mode are three times narrower and closer than the peak displayed for the first mode (since the wavelength of the standing wave is divided by 3 when the oscillation frequency is multiplied by 3).

6. Discussion

When used in combination with an AFM, the QCM will not visibly modify the image shape because of its oscillation.

The step range of a typical 1-m-wide, 256 pixel image is 4 nm, smaller than the in-plane oscillation of the quartz resonator. Similarly, the out-of-plane oscillation will not affect the resolution of the AFM for typical measurements in liquid. The main visible interactions are through the long- range longitudinal acoustic waves launched by the QCM which create standing wave patterns when reflected against the AFM cantilever holder. However, the tip displacement and position (tip-electrode distance) feedback then affect the frequency stability of the QCM and thus its sensitivity.

250 300 350 400 450 500 550 600

-40 -20 0 20 40 60 80 100

f1

time (s)

250 300 350 400 450 500 550 600

-10 -5 0 5 10 15 20

f3

time (s) approach

f d curve (141 um)

0 50 100 150 200 250 300 350 400

-60 -40 -20 0 20 40 60 80

time (s) f1

0 50 100 150 200 250 300 350 400

-15 -10 -5 0 5 10 15 20

f3

time (s)

withdraw approach

f d curve look for

resonance freq.

of cantilever

(141 um) (141 um)

Fig. 5. Graphs displaying the effect of longitudinal waves on the frequency stability of the resonator (top graphs show the evolution of frequency of the first mode, bottom graphs show the evolution of frequency of the third mode). Total displacement during cantilever approach is 141m, which is quite close to the theoretical half-wavelength of an acoustic wave propagating in water at 5 MHz (152m.4)).

tubing outlet

viton O-ring teflon liquid cell

QCM

Q

Sense QCM parameters measurement setup AFM cantilever

inlet

tubing glass prism

laser beam photodetector

Fig. 4. Experimental setup of the combined QCM and AFM instrument, based on a commercial AFM designed by molecular imaging and a QCM parameter analysis instrument built by Q-Sense AB (Go¨teborg, Sweden).

3976 Jpn. J. Appl. Phys. Vol. 41 (2002) Pt. 1, No. 6A J.-M. FRIEDTet al.

(4)

7. Conclusions

We have demonstrated the following using finite element analysis:

. the in-plane displacement of a resonator oscillating in liquid with a quality factor in the 1000 to 3000 range is at maximum in the nm range, typically 1 to 3 nm, when a 0.5 Vppsine wave voltage is applied to the resonator, and

. the out-of-plane displacement of the resonator is about one tenth of that of the in-plane displacement. Such a conclusion leads, for a resonator oscillating in liquid with a quality factor in the 1000 to 3000 range, to an out-of-plane displacement in the angstro¨m range. This out of plane displacement is a direct reflection of the finite size of the counter electrode.

The out-of-plane oscillation results in longitudinal waves being emitted in the liquid, which create standing waves when they bounce off a planar obstacle parallel to the electrode located a few microns to a few hundred microns above the surface (depending on the position of the AFM

cantilever holder during approach and scan). The interaction of the standing wave pattern and the crystal resonator is the source of the frequency fluctuations observed during the approach of the AFM cantilever holder towards the QCM surface.

1) A. Janshoff, H.-J. Galla and C. Steinem: Angew. Chem. Int. Ed.39 (2000) 4004.

2) B. Drake, C. B. Prater, A. L. Weisenhorn, S. A. Gould, T. R. Albrecht, C. F. Quate, D. S. Cannell, H. G. Hansma and P. K. Hansma: Science 243(1989) 1586.

3) P. A. Ohlsson, T. Tja¨rnhage, E. Herbai, S. Lo¨faas and G. Puu:

Bioelectrochem. & Bioenerg.38(1995) 137.

4) Z. Lin and M. D. Ward: Anal. Chem.67(1995) 685.

5) http://www-rocq.inria.fr/modulef/

6) D. Royer and E. Dieulesaint:Ondes e´lastiques dans les solides (tome 1)(Masson, Paris, 1996) pp. 129 and 146 [in French].

7) B. A. Auld:Acoustic Fields and Waves in Solids(John Wiley & Sons, New York, 1973) Vol. 1, pp. 75, 170.

8) B. A. Martin and H. E. Hager: J. Appl. Phys.65(1989) 2627.

9) B. A. Martin and H. E. Hager: J. Appl. Phys.65(1989) 2630.

10) K. K. Kanazawa: Faraday Discuss.107(1997) 77.

11) B. Borovsky, B. L. Mason and J. Krim: J. Appl. Phys.88(2001) 4017.

12) A. C. Hillier and M. D. Ward: Anal. Chem.64(1992) 2539.

Fig. 6. Sensitivity of the various modes (1, 3 and 5 from left to right) of the QCM to environmental disturbances including hydrostatic pressure changes when inserting a new sample using the peristatic pump, image scans and protein adsorption (here 2.5g/ml human plasma fibrinogen at date 4075 and 15 to 20g/ml human plasma fibrinogen at date 5530) on the gold-coated grounded electrode. Top graphs show the frequency variations while bottom graphs display the changes in damping (inverse of the quality factor). The first mode is obviously the one most affected by environmental disturbances.

Jpn. J. Appl. Phys. Vol. 41 (2002) Pt. 1, No. 6A J.-M. FRIEDTet al. 3977

Références

Documents relatifs

Donaldson Title Page Abstract Introduction Conclusions References Tables Figures J I J I Back Close Full Screen / Esc.. Printer-friendly Version

In this section, the limitation of L-band TomoSAR imaging in dense forests using TropiSAR data acquired over Paracou is illustrated; then, we report the results of forest

force spectroscopy measurements are performed automatically on hundreds of cells 32.. immobilized in

High-res- olution spiral computed tomography coronary angiogra- phy in patients referred for diagnostic conventional coro- nary angiography. Hausleiter J, Meyer T, Hadamitzky M,

As we expect the same friction force at the tip/surface interface during a usual scan as during a lateral scan, we can link the friction signal Vr (in

Image of the humid hood used for sample incubation on a sample stage (mica disk glued on the glass rod) attached to the Z-piezo of the scanner. The humid ambient is kept by a

probe chip and used for controlling the tip–sample separa- tion of cantilevers in a probe array AFM system, was proposed and the feasibility of the concept was demon- strated. A

Beyond the bulk acoustic resonator: SAW I Another strategy for reducing the mass of the vibrating element: confine the acoustic wave to the surface I SAW: Surface Acoustic Wave