• Aucun résultat trouvé

Superior flow for bridge to life with self-expanding venous cannulas

N/A
N/A
Protected

Academic year: 2021

Partager "Superior flow for bridge to life with self-expanding venous cannulas"

Copied!
5
0
0

Texte intégral

(1)

Superior flow for bridge to life with self-expanding venous cannulas

Ludwig K. von Segesser

1,

*

, Martens Kalejs, Enrico Ferrari, Sandra Bommeli,

Olaf Maunz, Judith Horisberger, Piergiorgio Tozzi

Department of Cardio-vascular Surgery, CardioMet, Centre Hospitalier Universitaire Vaudois, Lausanne, Switzerland Received 9 February 2009; received in revised form 25 April 2009; accepted 20 May 2009; Available online 16 July 2009

Abstract

Background: Recently, a compact cardiopulmonary support (CPS) system designed for quick set-up for example, during emergency cannulation, has been introduced. Traditional rectilinear percutaneous cannulas are standard for remote vascular access with the original design. The present study was designed to assess the potential of performance increase by the introduction of next-generation, self-expanding venous cannulas, which can take advantage of the luminal width of the venous vasculature despite a relatively small access orifice. Methods: Veno-arterial bypass was established in three bovine experiments (69 10 kg). The LifebridgeW

(Lifebridge GmbH, Munich, Germany) system was connected to the right atrium in a trans-jugular fashion with various venous cannulas; and the oxygenated blood was returned through the carotid artery with a 17 F percutaneous cannula. Two different venous cannulas were studied, and the correlation between the centrifugal pump speed (1500—3900 RPM), flow and the required negative pressure on the venous side was established: (A) Biomedicus 19 F (Medtronic, Tolochenaz, Switzerland); (B) Smart canula 18 F/36 F (Smartcanula LLC, Lausanne, Switzerland). Results: At 1500 RPM, the blood flow was 0.44 0.26 l min 1

for the 19 F rectilinear cannula versus 0.73 0.34 l min 1 for the 18/36 F self-expanding cannula. At 2500 RPM the blood flow was

1.63 0.62 l min 1for the 19 F rectilinear cannula versus 2.13 0.34 l min 1for the 18/36 F self-expanding cannula. At 3500 RPM, the blood

flow was 2.78 0.47 l min 1for the 19 F rectilinear cannula versus 3.64 0.39 l min 1for the 18/36 F self-expanding cannula ( p < 0.01 for 18/

36 F vs 19 F). At 1500 RPM, the venous line pressure was 18 8 mmHg for the 19 F rectilinear cannula versus 19  5 mmHg for the 18/36 F self-expanding cannula. At 2500 RPM the venous line pressure accounted for 22 32 mmHg for the 19 F rectilinear cannula versus 2  5 mmHg for the 18/36 F self-expanding cannula. At 3500 RPM, the venous line pressure was 112 42 mmHg for the rectilinear cannula versus 28  7 mmHg for the 18/36 F self-expanding cannula ( p < 0.01 for 18 F/36 F vs 19 F). Conclusions: The negative pressure required to achieve adequate venous drainage with the self-expanding venous cannula accounts for approximately 31% of the pressure necessary with the 19 F rectilinear cannula. In addition, a pump flow of more than 4 l min 1can be achieved with the self-expanding design and a well-accepted negative inlet pressure for

minimal blood trauma of less than 50 mmHg.

#2009 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. All rights reserved.

Keywords: Cardiopulmonary support; ECMO; Cardiopulmonary bypass; Emergency bypass; Cannulation

1. Introduction

Cardiopulmonary support (CPS) or extracorporeal mem-brane oxygenation (ECMO) have the potential for both, cardiac as well as pulmonary support, for various indications

[1—3]including emergency applications for cardiopulmonary resuscitation as well as temporary support for planned surgical procedures on the heart and the tracheo-bronchial tree as well as high-risk interventional procedures.

Recently, a compact CPS system designed for simplified application[4—6]has been introduced with a menu-guided set-up and priming procedure. This systematic system is particularly suited for emergency CPS where quick device readiness and safety are of prime importance. Connection of this device to the circulatory system of the patient is usually performed by remote cannulation of a peripheral artery and vein, respectively, typically a common femoral artery and vein by the means of traditional, rectilinear, percutaneous cannulas (catheters over the wire). However, there is a conflict of interest in remote access perfusion as the operator’s preference goes to relatively small cannula diameters, whereas larger cannula diameters are necessary to achieve high blood flows.

The present study was designed to assess the potential performance increase by the introduction of next-genera-tion, self-expanding venous cannulas[7—10]which can take advantage of the luminal width of the venous vasculature

www.elsevier.com/locate/ejcts

* Corresponding author. Address: Department of Cardio-vascular Surgery, CardioMet, CHUV, BH 10-275, Centre Hospitalier Universitaire Vaudois, Rue du Bugnon 46, CH-1011 Lausanne, Switzerland. Tel.: +41 21 314 22 79; fax: +41 21 314 28 79.

E-mail address:ludwig.von-segesser@chuv.ch(L.K. von Segesser). URL:http://www.cardiovasc.net/

1LKvS is founder and shareholder of Smartcanula LLC, Lausanne,

Switzer-land.

1010-7940/$ — see front matter # 2009 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. All rights reserved. doi:10.1016/j.ejcts.2009.05.026

(2)

despite a relatively small access orifice. In contrast to the previous reports about the self-expanding cannula, a ready-for-use centrifugal pump is an integral part of this set-up, hence providing, by definition, augmentation of venous drainage.

2. Methods

Following approval of the veterinary office of the State of Vaud, Lausanne, Switzerland, for the experimental evalua-tion of new cardio-pulmonary bypass equipment, veno-arterial bypass was established in three bovine experiments (mean bodyweight 69 10 kg) using the recently introduced LifebridgeWsystem (Lifebridge GmbH, Munich, Germany).

2.1. Pump set-up

The LifebridgeWsystem (Fig. 1) is basically a very compact

heart—lung machine with three pre-assembled modules: - the base station;

- the drive module; and - the patient module.

The base station weighs 18 kg, has a small footprint and can be folded for ease of storage and transportation. A touch screen allows for menu-guided set-up and provides back-up power for 120 min on batteries.

The drive module weighs 18 kg, and can be snapped into the base station described above, where it is interfaced automatically. It holds the basic electronics for driving the pump as well as the electromechanical tubing clamp, the bubble detector, the venting pump, etc., provides 20-min power back-up for emergencies and can be refurbished for the next application.

The patient module is basically a disposable cube (Fig. 1), which holds all the liquid-carrying components of the system,

including venous reservoir/bubble trap, a centrifugal pump, an integrated hollow-fibre/heat exchanger structure, an arterial filter, a bubble detector, a flow sensor, the necessary tubing as well as electromechanical clamps for the arterial line and the A—V shunt.

Once the three modules are connected, the pump set-up is realised semi-automatically in a menu-driven fashion, including systematic testing of the flow sensor, the level sensor, the pressure sensors, the bubble detectors, the RAM and flash memory modules, the electromechanical clamps, the drives of the systemic pump and the venting pump as well as all the interfaces and controls. The total priming volume including a 3/8 in. 3/8 in. table set accounts for approximately 1400 ml.

2.2. Cardiopulmonary bypass

Following general anaesthesia with volatile anaesthetics, the animals were instrumented in standard fashion (EKG, central venous pressure, arterial pressure) and fully hepar-inised (Liquemin, Roche, Basel, Switzerland: 300 IU kg 1 bodyweight) with 480 s as the target for the activated coagulation time (ACT, International Technidyne Corpora-tion, Edison, NJ, USA).

The LifebridgeW

system was connected to the right atrium in trans-jugular fashion with venous cannulas to be assessed and the oxygenated blood was returned through the carotid artery with a 17 F percutaneous cannula.

Two different venous cannulas were studied and the correlation between the centrifugal pump speed (1500— 3900 RPM), the flow achieved and the required negative pressure on the venous side was established:

(A) A 19 F rectilinear wire-wound percutaneous Biomedicus cannula (Medtronic, Tolochenaz, Switzerland)

(B) A Smart canula 18 F/36 F (Smartcanula LLC, Lausanne, Switzerland) as shown in Fig. 2 is inserted percuta-neously in collapsed fashion (<18 F) and expands in situ (up to 36 F or the available vessel lumen, respectively). We have previously reported the functional principles of the self-expanding venous cannulas mainly for applications with gravity drainage that is, open heart surgery and mini-invasive cardiac surgery[7—10]. Briefly, the self-expanding cannula can be stretched and collapsed with a hollow mandrel, inserted into the venous vasculature, for example, through a femoral vein, over a guide wire in open, semi-open or percutaneous fashion[11]. Once the cannula tip is located in the target zone by trans-oesophageal echocardiogram (TEE), for example, in the right atrium and the superior vena cava, the guide wire is removed first, and the mandrel second. As a result, the next-generation design cannula expands spontaneously up to its nominal diameter of 36 F or the available vessel lumen, respectively. The resulting increase in luminal width is the main reason for improved drainage and reduced pressure drop.

These venous cannulas were tested in a random fashion after insertion through a percutaneous orifice prepared with an 18 F dilator, and at least three measurements were made for each pump speed level (increments of 500 RPM from 1500 RPM up to the maximum motor speed of 3900 RPM).

Fig. 1. The LifebridgeW

system is a very compact heart—lung machine with three pre-assembled modules: the base station, the drive module, and the patient module. The latter (upper-right) holds all the liquid carrying compo-nents of the system including venous reservoir/bubble trap, a centrifugal pump, an integrated hollow-fibre/heat exchanger structure, an arterial filter, a bubble detector, a flow sensor, the necessary tubing as well as electro-mechanical clamps for the arterial line and the A—V shunt.

(3)

2.3. Statistics

Continuous variables are expressed as mean standard deviation and displayed after linear regression analyses where applicable. Analysis of variance (ANOVA) for repeated measures is used for comparison between the groups.

3. Results

Blood flow was successfully assessed with the LifebridgeW

system in all animals as a function of the pump speed (stepwise increase from 1500 RPM to 2000, 2500, 3000, 3500 and 3900 RPM) using the 19 F rectilinear wire-wound percutaneous Biomedicus cannula in a random fashion versus the 18 F/36 F self-expanding Smart canula.

Fig. 3depicts the blood flows achieved at various pump speeds for the two cannulas tested. At 1500 RPM, the blood flow was 0.44 0.26 l min 1for the 19 F rectilinear cannula

versus 0.73 0.34 l min 1 for the 18/36 F self-expanding cannula. At 2500 RPM, blood flow was 1.63 0.62 l min 1for

the 19 F rectilinear cannula versus 2.13 0.34 for the 18/36 F self-expanding cannula. At 3500 RPM, the blood flow was 2.78 0.47 l min 1

for the 19 F rectilinear cannula versus 3.64 0.39 l min 1for the 18/36 F self-expanding cannula ( p < 0.01 for 18/36 F vs 19 F).

Fig. 4shows the venous line pressure at the various pump speeds for the two cannulas tested. At 1500 RPM, the venous line pressure was 18 8 mmHg for the 19 F rectilinear cannula versus 19 5 mmHg for the 18/36 F self-expanding cannula. At 2500 RPM the venous line pressure accounted for 22 32 mmHg for the 19 F rectilinear cannula versus 2 5 mmHg for the 18/36 F self-expanding cannula. At 3500 RPM, the venous line pressure was 112 42 mmHg for the rectilinear cannula versus 28 7 mmHg for the 18/36 F self-expanding cannula ( p < 0.01 for 18/36 F vs 19 F).

The venous line pressure as a function of pump flow is plotted inFig. 5. It can be easily recognised that the venous line pressure measured for the 18/36 F self-expanding cannula stays above 50 mmHg even for blood flows as high

as 4.7 l min 1, whereas the maximal blood flow reached with

the 19 F rectilinear cannula in the same setting, not more than 3.7 l at a negative pressure of up to 230 mmHg. The results of the linear regression analysis confirm the superior performance expressed by the venous line pressure as a function of blood flow for the self-expanding cannula over the entire test range.

4. Discussion The LifebridgeW

system achieves up to 27% higher pump flows in conjunction with the self-expanding venous Smart canulaW

, which can be collapsed to 18 F for insertion and

Fig. 5. Flow and inlet line pressure with the 19 F rectilinear wire-wound percutaneous Biomedicus cannula versus the 18 F/36 F self-expanding Smart cannula (scatter plot and linear regression analysis).

Fig. 2. The self-expanding venous cannula (Smartcanula LLC, Lausanne, Swit-zerland) for adults is stretched and collapsed with a mandrel to 18 F and inserted over a guide wire, e.g., through the femoral vein. Once the cannula tip has reached the target region, e.g., the right atrium or the superior vena cava, the guide wire is removed first, the mandrel second and the smartcanu-laWexpands up to 36 F or the available vessel lumen respectively.

Fig. 3. Flow achieved as a function of RPM (revolutions per minute) with the 19 F rectilinear wire-wound percutaneous Biomedicus cannula versus the 18 F/36 F self-expanding Smart cannula (mean SD and linear regression analysis).

Fig. 4. Venous line pressure as a function of RPM (revolutions per minute) with the 19 F rectilinear wire-wound percutaneous Biomedicus cannula versus the 18 F/36 F self-expanding Smart cannula (mean SD and linear regression analysis).

(4)

expanded in situ to 36 F or the luminal width of the access vein, respectively. Simultaneously, the venous line pressure, which is also the pump inlet pressure, can be kept for the self-expanding cannula at less than 50 mmHg for flows up to 4.7 l min whereas negative pressures well beyond 200 mmHg are observed with rectilinear cannulas for flows up to a maximum of 3.7 l min 1.

Obviously, the man—machine interface, in this setting, the cannula that connects the extracorporeal circulation to the venous system of the patient, is of prime importance. As a matter of fact, the cannula performance is more of an issue with remote cannulation, the preferred access in the emergency setting [1—3], than for central cannulation. There are a number of reasons including the relatively small dimensions of the peripheral veins, typically the femoral vein, as compared to the caval veins, which bring the de-saturated blood naturally back to the heart. In addition, the veins are thin walled, not suitable for important negative pressures which induce them to collapse, a phenomenon that during perfusion can shut off venous drainage completely.

The self-expanding cannula assessed here allows for addressing both these issues. As reported previously, the self-expanding venous cannulas [7—10] can be stretched and collapsed with a hollow mandrel, inserted into the venous vasculature over a guide wire through an access orifice similar to the size of the peripheral vein or even smaller[11]

and expanded to its nominal diameter of 36 F within the vena cava (typical diameter in adults is >20 mm or 60 F) or to the available luminal width of the access vessel (typical diameter of the femoral vein is 8 mm or 24 F). Already at the iliac level, the vein diameter increases, and, therefore, the narrow segment of a well-positioned self-expanding cannula is relatively short. It has to be mentioned here that a cross-sectional area of 1 cm2(equivalent to <34 F) is sufficient to carry up to 6 l min 1with a 200-cm-long venous line relying on gravity alone[12].

The underlying physics provides further understanding regarding the improved flow achieved with the self-expand-ing cannulas as the resistance to flow within a rectilinear tube increases with its length in linear fashion whereas the flow increases with the diameter at the fourth power. The superiority of the self-expanding cannula design resulting in a short reduction of its diameter at the point of insertion has also been demonstrated by both computation fluid dynamics

[7,13]and bench tests[14]. For the set-up presented here, the negative pressure required to achieve adequate venous drainage with the self-expanding venous cannula accounts for approximately 31% of the 19 F standard rectilinear design at negative pressures reaching 200 mmHg and more. This explains also why a pump flow well above 4 l min can be achieved with the self-expanding design despite a negative inlet pressure for minimal blood trauma of less than 50 mmHg.

The second advantage of the self-expanding venous cannula is its ability to support the venous wall. We have termed this phenomenon temporary caval stenting as it allows the cannulated vein to stay open and prevents it from collapsing [15]. In contrast to cardiopulmonary bypass for open heart surgery, where gravity drainage is sufficient for achieving the target flow in conjunction with the self-expanding cannula [10], temporary caval stenting is even

more helpful, when higher negative pressures in the venous cannula can be expected similar to augmentation using centrifugal pumps or vacuum. The term ‘temporary’ refers here to the fact that the caval stenting is reversible as the self-expanding cannula can be re-collapsed by simple traction and therefore easily removed at the end of the extracorporeal circulation.

Additional design improvements coming with the self-expanding venous cannula evaluated here include the extremely thin cannula wall, which is practically non-existent over a major proportion of the cannula length due to its open wall design. The uncovered grid structure allows for drainage of the venous affluent from collaterals directly into the cannula lumen, an effect that further enhances the flow.

Finally, there is the option of positioning a self-expanding venous cannula in trans-atrial fashion into the pulmonary artery tree [16]. This procedure allows, at least in the experimental set-up, not only for efficient decompression of the left ventricle during veno-arterial bypass (e.g., in prolonged ventricular fibrillation), but also for recovery of large blood volumes from pulmonary vasculature, which in turn improves haemodynamics.

References

[1] Dembitsky WP, Moreno-Cabral RJ, Adamson RM, Daily PO. Emergency resuscitation using portable extracorporeal membrane oxygenation. Ann Thorac Surg 1993;55:304—9.

[2] Moreno-Cabral RJ, Dembitsky WP, Adamson RM, Daily PO. Percutaneous extracorporeal membrane oxygenation. Adv Card Surg 1994;5:163—79. [3] von Segesser LK. Cardiopulmonary support and extracorporeal membrane

oxygenation for cardiac assist. Ann Thoac Surg 1999;68:672—7. [4] Ferrari M, Figulla HR. Mechanische Herz-Kreislauf-Unterstu¨tzung in der

Kardiologie (Circulatory assist devices in cardiology). Dtsch Med Wochenschr 2005;130:652—6.

[5] Mehlhorn U, Brieske M, Fischer U, Ferrari M, Brass P, Fischer JH, Zerkowski HR. LIFEBRIDGE B2TW— a portable, modular, rapidly available

‘‘plug-and-play’’ mechanical circulatory support system. Ann Thorac Surg 2005;80: 187—92.

[6] Ferrari M, Poerner TC, Brehm BR, Schlosser M, Krizanic F, Schmidt R, Figulla HR. First use of a novel plug-and-play percutaneous circulatory assist device for high-risk coronary angioplasty. Acute Cardiac Care 2008;1—5.

[7] Mueller XM, Mallabiabarrena I, Mucciolo G, von Segesser LK. Optimized venous return with a self-expanding cannula: from computational fluid dynamics to clinical application. Interactive Cardiovasc Thorac Surg 2002;1:23—7.

[8] von Segesser LK, Jegger D, Mucciolo G, Tozzi P, Mucciolo A, Delay D, Mallabiabarrena I, Horisberger J. The Smart CanulaTM: a new tool for

remote access in limited access cardiac surgery. The Heart Surgery Forum 2005;8:E241—5.

[9] Jegger D, Chassot PG, Bernath MA, Horisberger J, Gersbach P, Tozzi P, Delay D, von Segesser LK. A novel technique using echocardiography to evaluate venous cannula performance perioperatively in CPB cardiac surgery. Eur J Cardiothorac Surg 2006;29:525—9.

[10] von Segesser LK, Ferrari E, Delay D, Maunz O, Horisberger J, Tozzi P. Routine use of self-expanding venous cannulas for cardiopulmonary bypass: benefits and pitfalls in 100 consecutive cases. Eur J Cardiothorac Surg 2008;34:635—40.

[11] von Segesser LK. Peripheral cannulation for cardio-pulmonary bypass. Multi-media Manual of Cardiothoracic Surgery,doi:10.1510/mmcts.2005.001610. [12] Ni Y-M, Leskosek B, Shi L-p, Chen Y-l, Qian L-f, Li R-y, Tu Z-l, von Segesser LK. Optimization of venous return tubing diameter for cardio-pulmonary bypass. Eur J Cardiothorac Surg 2001;20:614—20.

[13] Jegger D, Sundraram S, Shah K, Mallabiabarrena I, Mucciolo G, von Segesser LK. Using computational fluid dynamics to evaluate a novel venous cannula (Smart canulaW) for use in cardiopulmonary bypass

(5)

[14] Abdel-Sayed S, Favre J, Horisberger J, Taub S, Hayoz D, von Segesser LK. New bench test for venous cannula performance assessment. Perfusion 2007;22:411—6.

[15] von Segesser LK, Siniscalchi G, Kang K, Maunt O, Horisberger J, Ferrari E, Delay D, Tozzi P. Temporary caval stenting improves venous drainage

during cardiopulmonary bypass. Interact Cardiovasc Thorac Surg 2008;7:1096—100.

[16] von Segesser LK, Kwang K, Tozzi P, Horisberger J, Dembitsky W. A simple way to decompress the left ventricle during venoarterial bypass. Thorac Cardiovasc Surg 2008;56:337—41.

Figure

Fig. 1. The Lifebridge W system is a very compact heart—lung machine with three pre-assembled modules: the base station, the drive module, and the patient module
Fig. 5. Flow and inlet line pressure with the 19 F rectilinear wire-wound percutaneous Biomedicus cannula versus the 18 F/36 F self-expanding Smart cannula (scatter plot and linear regression analysis).

Références

Documents relatifs

Abstract: A group exercise in which students discuss predetermined questions which ask them to reflect upon their writing experience (prompts included below) with a partner, then

L’augmentation du nombre de demandeurs d’emploi, consécuti- ve à la dégradation de la conjonc- ture, s’est accompagnée d’une baisse de la part des femmes dans les

L’ensemble de l’œuvre, y compris son paratexte, met ainsi sur pied une profonde dynamique de la compassion – qui est l’un des noms que peut prendre le vivre

Simulations were compared to experimental phase contrast images acquired on the physical flow phantom under similar conditions.. The velocity encoding value Venc was

Exosomes have been proposed to be a novel way for intercellular communication contributing to the normal physiology of the nervous system and to the trans-synaptic transmission of

Given the pervasive and insidious nature of many forms of violence, including sexual harassment, all girls must be considered to be vulnerable and &#34;at risk.&#34; It is

Apres avoir note Ie caractere parfois contradietoire des resultats obtenus dans plusieurs etudes, ces auteurs font porter leur argumentation sur deux idees

4 Pour chaque match, les places du stade sont mises en vente dans les proportions suivantes :.