VALIDATED QUANTITATIVE SIMULTANEOUS DETERMINATION OF

COCAINE, OPIATES AND AMPHETAMINES IN SERUM BY

UHPLC-MS/MS

N. Dubois1, B. Debrus2,3, Ph. Hubert2,3, C. Charlier1,3 1

Service de Toxicologie Clinique, Médicolégale, de l’Environnement et en Entreprise, CHU, Av. de l’hôpital 1, B-4000 Liège, Belgium, 2Laboratoire de Chimie Analytique, Département de Pharmacie, Université de Liège, CHU,

Av. de l’hôpital 1, B-4000 Liège, Belgium.3CIRM, Université de Liège, B-4000, Belgique.

Simultaneous determination of cocaine, opiates and amphetamines in serum by ultra-high performance liquid chromatography coupled to tandem mass spectrometry (UHPLC-MS/MS) allowed to replace favourably gas chromatography coupled to mass spectrometry (GC-MS) used until now in our laboratory. It had to answer to accreditation demand according to Belgian Accreditation (Belac). Twenty-one deuterated internal standards were added to 500µL of serum. Sample pre-treatment consisted of solid-phase extraction using Oasis MCX cartridges 1mL, 30mg (Waters, Zellik, Belgium). Chromatographic separation was done on an Acquity HSS T3 column (2.1 x 100mm, 1.8µm, Waters). Mobile phase consisted of pH 3 ammonium formate buffer and of methanol adjusted to pH 3 with formic acid and was delivered according to a gradient elution mode. Compounds were then analysed by tandem mass spectrometry operated in the multiple reaction monitoring (MRM) mode [1,2]. The method was validated using total error approach.

Twenty-seven drugs were separated in 19 minutes. The linearity of the method was acceptable in the validated range of concentrations. The biases were lower than 15% for all compounds and the relative standard deviations for repeatability and intermediate precision did not exceed 15%. Lower and upper

β-expectation tolerance intervals did not exceed the acceptance limits of 20% for concentrations upper than 20µg/L and 50% for concentrations lower than 20µg/L. The limits of quantitation were lower than 7µg/L for all compounds.

References

[1] Lacroix C. et coll., Quantification des opiacés, cocaïniques et amphétaminiques par chromatographie liquide haute performance / spectrométrie de masse en tandem après séparation en ligne de l’échantillon, Ann Toxicol Anal., 2008 ; 20(1) : 25-38

[2] Rook E. et coll., The quantitative analysis of heroin, methadone and their metabolites and the simultaneous detection of cocaine, acetylcodeine and their metabolites in human plasma by high-performance liquid chromatography coupled with tandem mass spectrometry, J. Chromatogr. B, 2005 ; 824 : 213-221

Références

Documents relatifs

The aim of this study is therefore to develop and validate a quantitative ultra-high performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method for determination

Plasma estrone sulfate assay in men: Comparison of radioimmunoassay, mass spectrometry coupled to gas chromatog- raphy (GC-MS), and liquid chromatography-tandem mass

of submitted results for Case 1 report a total duration estimate which is different by less than 1% compared to the true duration. Some participants arrived at the exact

The pulse shape, duration, amplitude and velocity are calculated roughly from Esposti Ongaro et al.’s analysis (2002), and in the present study are taken as plausible val- ues

On considère ces antiprismes qui possèdent deux bases identiques parallèles et des faces latérales en forme de

Validated method for the determination of the ethanol consumption markers ethyl glucuronide, ethyl phosphate, and ethyl sulfate in human urine by reversed-phase/. weak anion

On-line solid phase extraction coupled with liquid chromatography/tandem mass spectrometry simultaneous determination of opiates, cocaine and its metabolites,

We chose to increase analysis time to separate all major phenolic compounds present in apple juice, especially to avoid, as much as possible, co-elutions that may affect