ABSTRACT ESPR 2016 LIST MANNIKIN STUDY

DEVICES FOR LESS INVASIVE SURFACTANT THERAPY: A MANIKIN STUDY V. RIGO1, C. Debauche2, P. Maton3, I. Broux1, D. Van Laere4.

1CHU de Liège- CHR de la Citadelle, Neonatology, Liège, Belgium. 2UCL St-Luc, Neonatology, Brussels, Belgium.

3

CHC - St Vincent, Neonatology, Liège, Belgium.

4

Universitair Ziekenhuis Antwerpen, Neonatology, Antwerp, Belgium.

Background: “Less invasive surfactant therapy” (LIST), or tracheal instillation of surfactant through a small catheter in spontaneously breathing infants, is gaining popularity. Different catheters are currently used for this purpose: a nasogastric tube inserted with (LISA) or without (Take Care) Magill’s forceps, a 13 cm 16G adult angiocath (MIST), a 30 cm F4 angiography catheter (Stockholm). We developed a specific device by combining a F5 umbilical catheter and an intubation stylet (Liege). We aimed to compare those 5 devices using INSURE as a reference.

Methods: 20 neonatologists from 4 institutions supporting different surfactant instillation policies intubated 2 manikin heads with the 5 catheters and an endotracheal tube in a

predetermined random sequence. Water was flushed trough the catheter. Video review provided times between laryngoscope (T1) or catheter insertion (T2) in the mouth and water flowing from the trachea. Participants gave an ease of use score (range: 1-9) for each catheter.

Results: Procedural times were longer with the Take Care method and shorter with the Liège device (Table). Failure rates were higher for LIST procedures than for INSURE. Take Care and LISA were rated as more difficult, while Liège, Stockholm and INSURE were considered easier.

Conclusions: LIST procedures remain difficult, even on a manikin. The choice of catheter is

important. A device combining the rigidity of a stylet with the soft distal end of an umbilical catheter is associated with procedures of shorter duration and is considered easier by neonatologists.

median (IQR)

LISA

(a)

Take

Care (b)

MIST

(c)

Stockholm

(d)

Liège

(e)

INSURE

(f)

Friedman

ANOVA

T1:

blade> water

24s (19-38) 30s (22-58) 17s (10-31) 23s (11-37) 17s (13-27) 19s (13-26) p<0.0001 Wilcoxon test: p <0.05 vs ef vs cdef vs b vs b vs ab vs ab

T2:

device>water

14s (11-22) 21s (13-43) 10s (5-24) 15s (7-23) 9s (6-14) 10s (8-18) p<0.0001 Wilcoxon test:

p <0.05 vs bef vs acdef vs be vs be vs abcd vs ab Ease of use

score: (1-9) 5 (4-6) 3 (2-4) 6.5 (5-7) 7 (4-8) 8 (6.5-8) 7 (6-8) p<0.0001 Wilcoxon test:

p <0.05 vs bcdef vs acdef vs ab vs abe vs abd vs ab Failed attempts

(/40) 7 6 8 4 3

Cochran Q test: p=0.38

Références

Documents relatifs

Temperature  is  directly  related  to  kinetic  energy  and  therefore  velocity...    Saturation  vapor  pressure  versus

We prospectively identified cases performed over a 15-month period in which a Heartrail catheter was used to facilitate stent delivery following failure of conventional

Although no differences were detected in the overall intestinal microbiota structure, the relative abun- dance of one particular taxon, the genus Clostridium, was found to

These are the entries of a finite elements model, which returns the interaction force between the tool tip and the membrane at the contact point (from which the robot command

The aim of this study was to investigate the relationship between quality of life (QoL) and chronic inflammation assessed using C -Reactive Protein (CRP) in patients

A linear regression model, testing for an additive genetic model, was used to test the association between each of the 41 candidate SNPs and the total KL, JSN, and osteophyte

In individuals with a depressive symptomatology, those with higher levels of overall mindfulness were less likely to have a low HDL-cholesterol and a high level of fasting blood

cinerea inoculation, all plants were kept under humid conditions (n = 32; 6SE); the experiment was repeated twice with similar results.. By removing oxalic acid released by B.