Structural characterization of isobar selenium containing

contaminant in a commercially available selenomethionine

standard using electrospray – Ion Mobility Time of Flight Mass

Spectrometry

J. Fara,b, G. Mazzucchellib, E. De Pauwb, G. Eppea,b;

a: Inorganic Analytical Chemistry (CART), Department of Chemistry, University of Liège b: Laboratory of Mass Spectrometry (CART), Department of Chemistry, University of Liège.

Author 1 accepts, if selected, to give an oral presentation: YES

Selenium (Se) is an essential and trace element. Benefic effects of selenium are reported in cancer prevention for supra-nutritional dietary intake of organic selenium like Se-enriched yeast. To understand its biological significance in toxicology or nutrition, selenium speciation (quantification and structural identification of Se species) is needed. However, the lack of authentic standard or reference materials prevent the use of retention time matching based method with an elemental, robust and specific detector like Inductively Coupled Plasma Mass Spectrometry (ICP MS) to be used. The problem can be overcome by the use of molecular mass spectrometry like electrospray MS. Selenium have 6 isotopes and a characteristic isotope pattern. This pattern makes easier the identification of a selenium containing compound in the mass spectra. Unfortunately electrospray ionization suffers from poor ionization efficiencies and requires a very high degree of purification of the samples (e.g. multi-dimensional liquid chromatography) prior to structural characterization using tandem mass spectrometry.

A major selenium containing compound identified in various selenium-rich supplement or biological sample is selenomethionine, the selenium analogue of the methionine amino-acid. Selenomethionine standard is commercially available and is used as internal standard in quantitation and quality control purposes.

In this work, we evaluated the purity of this standard with Ion Mobility Mass Spectrometry (IM-MS). IM-MS is an emerging tool in molecular mass spectrometry. The general principles of IM-MS is to separate ions into a slightly pressurized cell (mobility cell). The acceleration potentials are pushing the ion into the mobility cell. The velocity of ions will slow down according to their respective cross section.

The commercially available selenomethionine standard was investigated using the Waters© Synapt G2. Synapt G2 is High Resolution Time of Flight mass spectrometer equipped with an electrospray ion source. The high resolution mode (m/z: 196 Rm/Δm:10.000, RFWHM: 25.000) detected the presence of isobar compounds at m/z:

196. One was the 78Se selenomethionine, the other was the 80Se of a selenium containing contaminant. The two compounds were separated using the mobility cell. They produced fragment ions having almost similar mass at m/z: 181 but the identified contaminant fragmented more easily. This could explain why this impurity was not detected until now.

The structural characterization of this contaminant using the G2 in MSE mode is proposed in this work. In addition, in the view of using selenomethionine as internal

(2)

standard in analytical chemistry, we estimated the bias induced by this interference for selenomethionine speciation studies.

Références

Documents relatifs

We tested different protein extraction protocols based on previous applications of MALDI-TOF MS to fungi to produce mass spectra of 34 viable single-spore isolates belonging to

We report here the complementarity of native mass spectrometry (MS), ion mobility (IM) and collision induced unfolding (CIU) experiments for the structural characterization of

The present study have investigated positive and negative ion formation of radiosensitizers 1Me5NI, METRO, RONI, ORNI and NIMO using ESI mass spectrometry and quantum

B16F10 melanoma model, we found that a nontoxic dose of IL-2 immunocytokine synergized with tumor-specific anti- body to significantly enhance therapeutic outcomes compared

It was found that for a configuration without the aerodynamic shell, the FEM frequencies were within 4% of the experimental values for all of the modes

Les r´ esultats pr´ esent´ es dans ce chapitre montrent comment les processus m´ ecaniques et l’histoire tectonique d’un mod` ele peuvent ˆ etre int´ egr´ es ` a la construction

The aim of this project was to reduce the levels of OTA by supplementing the artificially contaminated solutions with seven strains of actinobacteria (AT10, AT8, SN7, MS1, ML5, G10

Second, using newly identified markers, we characterized the differential projection patterns of the distinct subclasses of type I neurons to the inner hair cells (IHCs), the