• Aucun résultat trouvé

10

3 10

2 10

Figure 4.61: Left: selected proton events with the MIP trigger in the period (01/01/2016-30/04/2018) as a function of energy. Right: migration matrix for the selected MIP sample.

reconstructed events distribution provides a complete proton flux measurement at least until 10 TeV and the result compared to the standard HET measurement is shown in Fig. 4.62. The proton flux measured with the MIP sample is up to 5% higher than that measured with the HET sample until 5 TeV. Above the difference is reduced with respect to the standard HET proton flux. This difference can be due to the modeling of the interaction cross-section (see current inelastic cross-section measurements compared to different models used in various MC simulations in [97]), however it is not clear in which specific part it is due only on that.

4.7 Results

The previous steps allowed us to give a preliminary proton flux measured by DAMPE based on 1.59006 ·107 selected proton candidates that are shown in Fig. 4.63 and Fig. 4.64. Tab. 4.6 shows a summary of the observed events in two bins per decade.

The flux is measured from 50 GeV to 100 TeV and it is shown in Fig. 4.65. Tab. 4.7 The measured proton fluxes are presented in Tab. 4.7 as a function of the energy. In

CHAPTER 4. MEASUREMENT OF THE PROTON FLUX

Energy (GeV)

102 103 104 105

)-1 sr-2 m-1 s1.7 (GeV2.7 E×Flux 8000

10000 12000 14000

16000 DAMPE

DAMPE MIPT analysis

Figure 4.62: Central points of the proton flux multiplied by E2.7 measured by DAMPE with the MIP sample (empty circles) compared to the standard HET sample (filled circles).

BGO Energy [GeV]

102 103 104 105

)1.7 Number of Entries (GeV×2.7E

109

1010

1011

raw data events in 32 bin/dec raw data events in 2 bin/dec

Figure 4.63: Selected proton candidates as a function of the reconstructed energy multiplied by E2.7in 32 bin per decade (red) and in 2 bin per decade (blue).

these figure, the points are placed along the y axis with an energy calculated for a flux E2.7 following [95]. Tab. 4.6 and Tab. 4.7 also contain a representative value of the energy in the bin interval for a flux E2.7. The error breakdown is shown in Fig. 4.66.

The systematic error related to the preselection cuts will be added in the next iteration of the analysis.

The proton flux measurement performed in this thesis was compared with the pre-vious results in Fig. 4.67. The DAMPE data confirms the spectral hardening at ∼300 GeV observed by ATIC02 [41], CREAM [48], PAMELA [44] and AMS-02 [40], and

re-138

BGO Energy [GeV]

1 10 102 103 104 105 106

Number of Entries

102

103

104

105

106

107

Figure 4.64: Selected proton candidates as function of the reconstructed energy, with the helium background subtraction.

Energy [GeV] Emin [GeV] Emax [GeV] Events

24.9 20 31.6 (8.033±0.003)·106

54.2 31.6 100 (6.739±0.003)·106

171 100 316 (9.68±0.01)·105

542 316 1000 (1.354±0.003)·105

1714 1000 3162 (2.10±0.01)·104

5419 3162 10000 (3.16±0.06)·103

17135 10000 31623 (3.63±0.19)·102

54186 31623 100000 (29±5)

Table 4.6: Number of proton candidates reconstructed in the various energy ranges.

veals a spectral softening above ∼ 10 TeV that is much more clear now in the current measurements scenario. This “early knee” in the spectrum could be explained by the three component model in the acceleration of cosmic rays described in [99] (used already by CREAM in [48]) and [100]. As explained in [99], this feature in the spectrum after 10 TeV could depict the maximum energy of one type of particle accelerator source.

Only after an other type of source starts in contributing with particle acceleration. This softening should be rigidity dependent and be visible in all other nuclear species as well, like for the hardening [62], but only looking at the flux of other nuclei species above 10 TeV will give a clearer picture.

CHAPTER 4. MEASUREMENT OF THE PROTON FLUX

Energy (GeV)

102 103 104 105

)-1 sr-2 m-1 s1.7 (GeV2.7 E×Flux

6000 8000 10000 12000 14000 16000 18000

DAMPE, this work

Figure 4.65: Proton flux multiplied by E2.7 measured by DAMPE. The systematic error band is also drawn.

Energy [GeV] Emin [GeV] Emax[GeV] φ±σstat±σsys 49.8 39.8 63.1 (283±0.09+2117)·103 78.9 63.1 100 (80.8±0.04+5.6−4.9)·103

125 100 159 (23.2±0.01+1.61.4)·10−3 198 159 251 (66.4±0.04+4.44.1)·104

314 251 398 (19.0±0.03±1.2)·10−4

498 398 631 (54.8±0.08+3.53.3)·105 789 631 1000 (16.05±0.04+1.00−0.98)·105 1251 1000 1585 (4.69±0.02+0.290.28)·10−5 1983 1585 2512 (14.09±0.06+0.860.87)·106 3142 2512 3981 (4.36±0.03+0.27−0.26)·106 4981 3981 6310 (1.30±0.01±0.08)·106 7895 6310 10000 (3.81±0.06±0.24)·10−7 12512 10000 15849 (1.10±0.02±0.72)·107 19830 15849 25119 (2.99±0.08±0.18)·108 31429 25119 39811 (0.73±0.03±0.06)·10−8 49812 39811 63096 (0.19±0.01±0.02)·108 78946 63096 100000 (0.49±0.05±0.06)·109 Table 4.7: Results for the proton fluxes according to Eq. 4.1 in units of [GeV s m2sr]−1.

140

Energy [GeV]

102 103 104 105

Error [%]

2 4 6 8 10 12 14 16 18

Statistical Trigger (HET) Charge slection Track selection PSD smear MC until 251 TeV

Figure 4.66: Error breakdown for the proton flux measurement. The error on the background subtraction is included in the statistical error calculation.

Energy (GeV)

102 103 104 105

)-1 sr-2 m-1 s1.7 (GeV2.7 E×Flux

4000 6000 8000 10000 12000 14000 16000 18000 20000 22000 24000

ATIC02 (2009) PAMELA (2011) AMS-02 (2015) CREAM III (2017) NUCLEON (IC) (2018) NUCLEON (KLEM) (2018) DAMPE, this work

Figure 4.67: Proton flux multiplied by E2.7 measured by DAMPE compared with various experimens: ATIC02 [41], PAMELA [44] corrected according to the note on page 4 in [98], AMS02 [40], CREAM III [48] and NUCLEON [67].

CHAPTER 4. MEASUREMENT OF THE PROTON FLUX

142

The discovery of cosmic rays is quite recent compared to other kinds of physics, as shown in the introduction in chapter 1. Our understanding of the mechanisms that de-scribe the acceleration of cosmic rays and their origin is also relatively new (sections 1.1 and 1.2.2). The rather young age of this field of research makes it deeply interesting, but also extremely challenging. Many detectors on the ground (section 1.3), on bal-loons (introduction of chaper 2) and in space (section 2.1) already made a lot of efforts to improve cosmic-ray measurements, and several are currently continuing them. The DAMPE detector is designed to pursue this task, as described in section 2.2.

Only the best components were chosen during the construction of the DAMPE Sili-con Tungsten tracKer (STK), starting from selecting the tungsten plates (section 3.1.1) and measuring the planarity of the trays, evaluating also the effects of thermal cycles on them (section 3.1.2). All the silicon ladders were tested after the production and individually characterized, so that a final selection could be made to choose the best samples. Several distributions were recorded and evaluated to summarize their linearity and thickness (Fig. 3.14 and Fig. 3.15).

The signals of the read-out chips of the silicon ladders were calibrated with an analy-sis based on ground (section 3.2.3) and orbit data (section 3.2.5). We monitored the calibrated gain of the chips, at scheduled time intervals, by studying on-orbit data (Fig. 3.34). A charge-loss correction of the tracker response was performed on-ground (section 3.2.4) and on-orbit (section 3.2.6). Using on-orbit data, this calibration im-proved the preliminary tracker charge resolution for protons and helium by 13% and 21%, respectively (Fig. 3.39).

We performed an analysis based on 30 months of data from DAMPE, dedicated to calculate precisely the proton flux, and described in every detail in chapter 4. MC events were studied, simulated with several physics lists (summarized in section 4.3.1).

A systematic comparison between data and MC using all the studied distributions for protons allowed to select the list that matched most closely the data (Fig. 4.14, Fig. 4.34 and Fig. 4.27). We estimated the systematic uncertainties of all the cuts applied on the data by comparing the cut efficiencies between data and MC for various subdetec-tors, as detailed in section 4.4. The electron (section 4.3.7) and helium (section 4.3.8)

Conclusions and outlooks

contaminations were studied and taken into account in the final measurement. The sys-tematic error study was completed by estimating the MC related uncertainties, reported in section 4.5.1. The unfolded flux was obtained iteratively, as explained in section 4.5.2, and the method employed was cross-checked by injecting a known input spectrum (sec-tion 4.5.4). The proton flux measured with this analysis, allowed to improve the current knowledge of the energy dependence of the proton spectrum, which is the key element in understanding the origin, propagation and acceleration of cosmic rays in the Universe.

Looking at the flux of other nuclei species above 10 TeV will help to understand if this softening is rigidity dependent or if it is only visible in protons. Future space mis-sions, like HERD [101], plan to contribute to the cosmic-ray flux measurements, reaching energies up to 1 PeV.

144

Appendix A

Metrology of the STK

A.1 The tungsten plates

The ID number position with respect to the tray are shown in the left panel of Fig. A.1, the plates chosen for the three trays of STK are shown in the right panel of Fig. A.1 and in Fig. A.2. As completeness the information on the measurements regarding the

Figure A.1: Left: position of the tungsten plates in the STK tray. Right: ID of the plates used for the first STK tray containing W.

selected plates are reported in Tab. A.1, A.2, A.3, A.4, A.5 and A.6 respectively to the first, second and third layer with W in STK. For the first tray the 16 plates chosen fulfill all the requirements in Tab. 3.1. Among the other 32 plates, 6 do not fulfill completely the requirements. It was decided to place those plates at the edge of the trays.

APPENDIX A. METROLOGY OF THE STK

Figure A.2: ID of the plates used for the second (left) and third (right) STK tray containing W.

Plate ID tray Dx1(mm) Dx2(mm) Dy1(mm) Dy2(mm)

3 1 188.030 188.022 188.034 188.034

8 1 188.008 187.988 188.027 188.027

12 1 188.027 188.020 187.996 187.996

13 1 187.979 187.999 188.026 188.026

17 1 188.030 188.025 188.006 188.006

20 1 188.049 188.029 188.054 188.054

21 1 188.014 188.003 188.040 188.040

23 1 188.036 188.033 188.022 188.022

24 1 187.992 187.994 188 188

25 1 188.021 188.036 188.026 188.026

26 1 188.028 188.016 188.001 188.001

31 1 188.027 188.044 188.033 188.033

34 1 188.046 188.038 188.050 188.050

36 1 188.050 188.044 188.046 188.046

38 1 188.042 188.057 188.047 188.047

46 1 187.992 187.999 188.048 188.048

Table A.1: Sides information for the selected W plates for the first tray.

146

Plate ID tray area (mm2) thickness (mm)

3 1 35355,5 1.04

8 1 35348,5 1.03

12 1 35348 1.02

13 1 35346.6 1.03

17 1 35351.6 1.01

20 1 35360.2 1.03

21 1 35353.6 1.

23 1 35353.7 1.03

24 1 35346.3 0.99

25 1 35354.1 0.98

26 1 35349.1 1.02

31 1 35356.3 1.

34 1 35360.7 1.

36 1 35363 1.02

38 1 35362.8 1.02

46 1 35351.2 1.

Table A.2: Area and thickness information of W plates for the first tray.

Plate ID tray Dx1(mm) Dx2(mm) Dy1(mm) Dy2(mm)

2 2 188.014 188.024 188.017 187.988

5 2 188.022 188.026 188.028 188.042

6 2 188.013 187.996 188.031 188.027

10 2 188.044 188.031 188.011 188.016

29 2 188.050 188.026 188.021 188.015

33 2 188.040 188.052 188.048 188.055

35 2 188.056 188.056 188.036 188.05

42 2 188.047 188.046 188.060 188.053

44 2 188.008 188.011 188.027 188.043

47 2 188.052 188.044 188.045 188.056

48 2 188.036 188.047 188.049 188.052

49 2 188.047 188.056 188.050 188.053

54 2 188.048 188.041 188.023 188.026

14 2 188.036 188.034 187.997 188.006

16 2 188.040 188.051 188.056 188.045

30 2 188.045 188.053 188.063 188.053

Table A.3: Sides information for W plates of the second tray.

APPENDIX A. METROLOGY OF THE STK

Plate ID tray area (mm2) thickness (mm) Weight (g)

2 2 35348. 1.0 689.07

5 2 35355.1 1.01 699.09

6 2 35350.3 1.0 688.09

10 2 35353.6 1.03 691.87

29 2 35354.5 1.02 702.63

33 2 35362.3 1.0 693.81

35 2 35362.6 1. 683.45

42 2 35363.4 1.01 680.81

44 2 35352.4 1.02 700.37

47 2 35362.5 1. 687.78

48 2 35361.3 1. 686.35

49 2 35363.4 1. 695.7

54 2 35357. 1.02 699.46

14 2 35380.9 1.01 698.4

16 2 35350.9 0.99 684.68

30 2 35362. 1.02 699.26

Table A.4: Area, thickness and weight information for W plates of the second tray.

Plate ID tray Dx1(mm) Dx2(mm) Dy1(mm) Dy2(mm)

56 3 188.025 188.017 187.994 188.003

57 3 188.035 188.035 187.992 187.998

58 3 188.000 187.998 187.953 188.030

59 3 188.009 188.026 188.014 188.035

63 3 188.024 188.045 188.041 188.070

64 3 188.044 188.016 188.079 188.083

68 3 188.031 188.046 188.081 188.076

70 3 188.051 188.085 188.058 188.026

71 3 188.024 188.045 188.077 188.087

72 3 188.042 188.023 188.085 188.080

73 3 188.020 188.038 188.078 188.081

79 3 188.049 188.022 188.080 188.091

81 3 188.046 188.021 188.064 188.082

19 3 188.003 188.016 188.03 188.033

27 3 188.042 188.038 187.999 188.017

28 3 188.026 188.031 187.994 188.006

Table A.5: Sides information for W plates of the third tray.

148

Plate ID tray area (mm2) thickness (mm) Weight (g)

56 3 35347.7 1.02 702.88

57 3 35349.6 1.0 682.94

58 3 35342.2 1.0 687.61

59 3 35351.9 1.0 702.17

63 3 35360.9 1.04 691.47

64 3 35364.9 1.0 697.42

68 3 35366. 1.02 697.91

70 3 35364.7 1.03 698.56

71 3 35365.9 1.01 688.01

72 3 35365.6 1. 693.55

73 3 35364.4 1. 691.04

79 3 35366.8 1.0 689.08

81 3 35364. 1.0 688.01

19 3 35351.7 1.01 688.01

27 3 35353. 1. 691.94

28 3 35349.4 1.01 688.57

Table A.6: Area, thickness and weight information for W plates of the third tray.

A.2 The trays