Investigating the variability of cardiac pulse artifacts across heartbeats in simultaneous EEG-fMRI recordings: A 7T study

JORGE, João, et al.

Abstract

Electroencephalography (EEG) recordings performed in magnetic resonance imaging (MRI) scanners are affected by complex artifacts caused by heart function, often termed pulse artifacts (PAs). PAs can strongly compromise EEG data quality, and remain an open problem for EEG-fMRI. This study investigated the properties and mechanisms of PA variability across heartbeats, which has remained largely unaddressed to date, and evaluated its impact on PA correction approaches. Simultaneous EEG-fMRI was performed at 7T on healthy participants at rest or under visual stimulation, with concurrent recordings of breathing and cardiac activity.

PA variability was found to contribute to EEG variance with more than 500 μV2 at 7T, which extrapolates to 92 μV2 at 3T. Clustering analyses revealed that PA variability not only is linked to variations in head position/orientation, as previously hypothesized, but also, and more importantly, to the respiratory cycle and to heart rate fluctuations. The latter mechanisms are associated to short-timescale variability (even across consecutive heartbeats), and their importance varied across EEG [...]

JORGE, João, et al . Investigating the variability of cardiac pulse artifacts across heartbeats in simultaneous EEG-fMRI recordings: A 7T study. NeuroImage , 2019, vol. 191, p. 21-35

DOI : 10.1016/j.neuroimage.2019.02.021 PMID : 30742980

Available at:

http://archive-ouverte.unige.ch/unige:129926

Disclaimer: layout of this document may differ from the published version.

1 / 1

(2)

−400 0 +400

Amplitude (μV)

−400 0 +400

Amplitude (μV)

0 0.2 0.4 0.6 0.8 1.0

−400 0 +400

Time (s)

Amplitude (μV)

ECG−based trigger estimation

EEG−based alignment (correlation maximization)

Outlier exclusion (Euclidean distance)

Références

Documents relatifs

The method is based on a truncated series expansion of the solution and it is suited to furnish directly the position of the magnetisation vector at the end of the resonance process

More precisely the present objective is to remove ocular (blinking) artifacts in simulated and real EEG signal using Independent Component Analysis and wavelet

The described method enables to spatially characterize the artifacts. However, artifacts can be defined as well by their frequencies. This is the reason why each of these steps can

Depuis 2008, la plate-forme eco3M est utilisée en Master Recherche Océanographie en options « Océanographie Physique Chimique et Biologique » (jusqu’en 2011) puis

_ Oui, la chaise longue, j'ai voulu m’asseoir dedans pour lire le journal et elle s'est refermée brusquement, je suis resté coincé, mais j'ai quand même pu lire l'article sur la

Pour améliorer la sécurité et la qualité de cette prise en charge, nous devons nous atteler, de façon interprofes- sionnelle, à mieux comprendre l’organisation patholo- gique

The proposed signal processing method is based on sub- band decomposition using the wavelet transform, and has been tested on human and small rodents ECG signals