On the accelerometer calibration onboard GRACE

C. Raizner, M. Weigelt, M.G. Sideris

Geomatics Engineering, University of Calgary mlbweige@ucalgary.ca

(2)

Satellite system: GRACE

Initial orbit height: ~ 485 km Inclination: ~ 89°

mission duration: 5+ years

GRACE = Gravity Recovery and Climate Experiment

source: GFZ-Potsdam

Key technologies:

• GPS receiver

• Accelerometer

• K-Band Ranging System

(3)

Data Processing

GPS

 position

2. principle of energy conservation

 gravity field along the orbit

3. spherical harmonic analysis

 

kin

   

T c E U Z fdx

(4)

Impact of Accelerometer

(5)

Calibration Approach

PROBLEM: accelerometer inaccessible

comparison to an a-priori model (e.g. EGM96):

uncalibrated disturbing potential

disturbing potential from a-priori gravity field

E U Z f t dx T 0 kin

c E U Z f

m

dx

T

kin

(6)

Error Model

true

non-gravitational accelerations

f t

measured

non-gravitational accelerations

f m error model

simple example:

caution: overmodeling for higher order models

b f

s

f

t

 

m

bias b

scale factor s

(7)

Calibration Steps

comparison to a-priori model T-T0

energy calibration

 no physical meaning

of estimated parameters

accelerometer calibration

 derivation of physical

meaningful parameters for each component of acc.

robust least-squares adjustment

(Huber estimator)

estimated parameters calibrated disturbing potential

(8)

Energy Calibration - Models

approach 1:

approach 2:

approach 3:

approach 4:

T c d t s f

m

dx

T

0 1

T c d t d t s f

m

dx

T

0 1 2 2

2 1

2 3

0 1 2 3

1 1

2 6

TT       c d t d    t d     t sf dx

m

3

0 1 3

1

TT       c d t 6 d     t sf dx

m

(9)

Energy Calibration - Results

approach 1: linear drift

(10)

Energy Calibration - Results

approach 2: linear drift + quadratic drift

(11)

Energy Calibration - Results

approach linear drift d1 quad. drift d2

cubic drift

d3 scale mean

of Δ RMS

1 426.9322 - - 0.0884 0.0050 1.6052

2 206.0507 25.2190 - 0.0444 0.0047 1.5560 3 -1535.5293 382.5055 -36.6489 0.0448 0.0047 1.5566

4 329.0073 - 2.5861 0.0444 0.0047 1.5566

statistics for differences between calibrated and a-priori disturbing potential

(12)

Acceleration Calibration

fixed scale factor!

approach 1: one bias component for each axis once per day

approach 2: one bias and one drift component for each axis once per day

T c b dx b dy b dz

T

0 x y z

T c b dx b dy b dz d tdx d tdy d tdz

T 0 x y z x y z

(13)

Acceleration Calibration - Results

approach 1: one bias component for each axis once per day

(14)

approach 1: one bias component for each axis 4 times per day

Acceleration Calibration - Results

(15)

approach 2: bias and drift model

Acceleration Calibration - Results

(16)

Conclusion

successful calibration on energy level second order model is recommended

problem: no physical meaning of the estimated parameters calibration on accelerometer level is problematic

no scale estimation at the moment

estimated parameters can currently not be interpreted

 problem needs further investigation

(17)

On the accelerometer calibration onboard GRACE

C. Raizner, M. Weigelt, M.G. Sideris

Geomatics Engineering, University of Calgary mlbweige@ucalgary.ca

Questions ?

Références

Documents relatifs

The UV-visible straylight will be corrected using the Zhong method [3] which consist in characterizing the instrument’s response toa set of monochromatic laser sources that cover

ASTEC (Automatic Selection of clinical Trials based on Eligibility Criteria) is a French national research project that aims to develop an CRSS designed on the 3 principles

la présente étude se fixe principalement deux objectifs : d’abord, ouvrir l’histo‑ riographie de la crise de l’exclusion à l’histoire plus large du continent et notam‑ ment

Conclusions: Results of the study suggest that the Vivago W wrist-worn accelerometer is a valid measure of PA at varying levels of intensity.. The study has also defined

ECD Fourier transform ion cyclotron resonance mass spec- trometry (FT-ICR MS) patterns of a complete series of model, single amino acid-substituted H-RAAAAXAAAAK- OH peptides

Model training and tests were performed offline. However, the classification task simulated online conditions since only the present data vector was known by the classifier. The

The importance of the IGF-I-PI3K-Akt- mTOR pathway signaling has been largely confirmed and the multi-protein complex mTORC1 emerged to play a fundamental role in the regulation

Selenium isotope and elemental systematics of recycled sediments, when interpreted within the previously established framework of marine Se cycle (7, 18–20) (section S1 and fig.