Aiming at a disorder’s concept by 3D QCA vs.

FFR: a case of advanced ballistics

Christian Seiler

*

Department of Cardiology, University Hospital, CH-3010 Bern, Switzerland Online publish-ahead-of-print 9 September 2010

This editorial refers to ‘Three-dimensional and

two-dimensional quantitative coronary angiography, and

their prediction of reduced fractional flow reserve’†, by A.S.C. Yong et al. on page 345

‘Oculo-stenotic reflex’ is a term which ridicules the variably devel-oped attraction of interventional cardiologists to ‘seal’ coronary atherosclerotic bumps by balloon angioplasty. Its serious template relates to their daily business of estimating the functional relevance of a coronary stenosis by its structure. It is well known that even the experienced interventionalist may be ‘lost in translation’ when it comes to judging intermediate severity stenoses.1 Long before the development of intracoronary sensor guidewires for functional stenosis assessment, attempts were made to improve the structure– function translation by quantitative coronary angio-graphy (QCA) as obtained in two (2D), but also as reconstructed in three dimensions (3D).2,3 The two principles of any coronary structure – function translation are (i) statistical comparison of single, physically important structural stenosis elements with a functional reference; and (ii) the mathematical integration of steno-sis’ geometric hallmarks to a single measure with subsequent stat-istical reference comparison.

Ideally, both approaches would have to anticipate all the geo-metric patterns of a coronary atherosclerotic lesion, because it is the composite of all its finite structural elements which determines the drop in flow or perfusion pressure. Thus, the method would be aiming at a disorder’s concept (i.e. recognition of how exactly and variably the epicardial stenoses grow in coronary artery disease; CAD). In philosophical terms, ‘a disorder’s concept’ is an oxymoron and, therefore, pointing at it probably unsuccessful. In geometrical terms and as outlined below, it is challenging to model the structural elements of a stenosis accurately into a haemodynamic parameter.

Notwithstanding these considerations, Yong and co-workers have observed, on a purely statistical basis, 2D- and 3D-QCA par-ameters and compared them with coronary pressure-derived frac-tional flow reserve (FFR) as the funcfrac-tional reference (FFR ¼ Pd/Pao,

where Pdis the mean distal coronary pressure and Paois the mean aortic pressure).4The prime conclusion of the thorough work, for which the authors have to be commended is not unexpected: ‘the accuracy of QCA in predicting functionally significant FFR is limited . . . ’.

The present study

In 63 CAD patients undergoing elective percutaneous coronary intervention (PCI) of 63 atherosclerotic lesions, Yong et al. com-pared the 2D- and 3D-QCA-derived percentage diameter stenosis, minimal vessel lumen diameter (MLD), lesion length, minimal luminal cross-sectional area (MLA), and the percentage area steno-sis with FFR.43D-QCA reconstruction by a customized software was performed offline on the basis of two orthogonal angiographic views of the target lesion in the ECG-triggered, end-diastolic frame. 2D-QCA-derived MLD correlated best and directly with FFR at an r2linear regression coefficient of 0.34, and MLA obtained by 3D-QCA correlated best and directly with FFR at an r2linear regression coefficient of 0.40. Receiver operating characteristics

analysis showed that an MLD ≤1.25 mm (best cut-off) as

measured by 2D-QCA predicted a pathological FFR ,0.75 with a sensitivity of 3/4 and a specificity of 2/3; an MLA ≤1.60 mm2 (best cut-off) as measured by 3D-QCA predicted a pathological FFR ,0.75 with a sensitivity and specificity of 4/5. Statistically, the moderate linear regression coefficients for the association between structural (x-axis on their figure 4) and functional (y-axis) parameters were improved to an r2value of0.5 when applying a minus x21 function (best fit mathematical model), whereby—counter-intuitively—2D-QCA MLD was related more closely to FFR than 3D-QCA MLA.

Which are the sources of all that ‘noise’ in the relationship between structural coronary parameters and their functional refer-ence? Its level is considerable, i.e. 50 – 76% of the variability between structural and functional parameters relate to parameters others than MLD, MLA, and FFR. By statistical definition, FFR is the independent reference variable, assuming MLD and MLA as the

*Corresponding author. Tel: 41 31 632 36 93, Fax:+41 31 632 42 99, Email:christian.seiler@insel.ch

doi:10.1093/eurheartj/ehq259

The opinions expressed in this article are not necessarily those of the Editors of the European Heart Journal or of the European Society of Cardiology.

Published on behalf of the European Society of Cardiology. All rights reserved.&The Author 2010. For permissions please email: journals.permissions@oup.com.

European Heart Journal (2011) 32, 261–263 doi:10.1093/eurheartj/ehq314

(2)

suspects and culprits of the poor relationship. From a biophysical standpoint, such a notion has to be challenged by interrogating FFR as well as QCA.

Critique of QCA

Though they are quantitatively obtained parameters, each of the numerous individual factors such as MLD, MLA, percentage diam-eter stenosis, etc., does just partially account for the perfusion pressure drop across a stenotic lesion (DP). Even in the geometri-cally simple setting of a focal coronary stenosis created artificially by a circumferential balloon constrictor, the pressure drop is a quadratic function of pressure loss due to viscous friction (f) and flow (Q) separation (s):5 DP ¼ fQ+ sQ2. The viscous friction pressure loss coefficient f is determined by the stenosis length (L) and—inversely—by the minimal cross-sectional area (As; f ¼ L/As× 1/As). The flow separation pressure loss coefficient s is influenced by the difference between the luminal cross-sectional area of the stenotic and the non-stenotic (An) part of the vessel

[(1/As – 1/An)2]. Furthermore, the pressure drop is affected by the blood viscosity and its density, by the flow velocity profile, and by a geometric factor mostly not accounted for, i.e. the steno-sis entrance and exit angle. The contribution of all these factors would favour the description of their haemodynamic result (pressure drop) by a quadratic or even a cubic equation. Thus, the above-mentioned result of the study of Yong and co-workers4 that the best fit between MLD an MLA and FFR, respectively, is a minus x21 function does not—physically—describe the entire model. Particularly, it is unreasonable that 2D- and 3D-QCA data fit similarly well to FFR by the same minus x21function.

In the above context, a 3D view of the target lesion might be preferable over a 2D aspect, because factors such as the minimal cross-sectional area, the lesion length, and the entrance and exit angle, which vary according to the eccentricity of a stenosis, could be better accounted for. Figure 1 illustrates the value of two orthogonal views of a stenosis in the context of an eccentric lesion: the left main coronary artery dissection leading to a very severe eccentric lesion may be missed in one plane. However, it

Figure 1 Coronary angiogram with magnified views (middle rows) and schematic drawings (bottom rows) of the left coronary artery obtained at different time points during the exam (columns A, B, and C). The angiogram in column A is taken at the start of the exam. The AP caudal view shows a normal left main coronary artery. The angiogram in column B (AP view) is taken 5 min later following the attempt at coronary intubation by an Amplatzer AL2 catheter. The patient went into cardiogenic shock. The discrete haziness at the cranial border of the left main coronary artery is hardly visible; it is schematically illustrated in the drawing at the bottom of column B. The angiogram in column C (AP cranial view) is taken only after implantation of a left ventricular assist device; a tight distal left main stenosis becomes visible.

Editorial

262

(3)

has to be considered that the 3D coronary angiography employed in the present study is not ‘real-space’ or real-time 3D (with 15 images/s), but just a reconstructed 2D image on the basis of two image planes. The algorithm employed for the 3D reconstruction may be more decisive for the parameters obtained than the real stenosis geometry.

Critique of FFR

The haemodynamic reference used (FFR) is not as solid as suggested, but rather a moving target at which the ‘gun’ of QCA has been pointed. FFR may have substantially contributed to the moderate correlation found in the study of Yong et al. between structural and functional coronary parameters.4FFR is a coronary pressure-derived haemodynamic parameter, and the ‘F’ in ‘frac-tional Flow reserve’ can only be described by a pressure if certain conditions are fulfilled. On the basis of Ohm’s law (DP ¼ RQ, where R is vascular resistance), the most interesting, but diffi-cult to obtain, flow-rate Q is mirrored by the coronary pressure drop DP or by FFR only if R is constant. Theoretically, the con-stancy of R in the coronary microcirculation can be reached by maximal vasodilation. The vasodilator employed for that purpose is adenosine intravenously (i.v.) at a rate of 140 mg/min/kg. One of the first sources of varying FFR values in the present study is that the dose and route of application changed between i.v. at the above rate and intracoronary at a dose which is four times higher than that used in the literature. Also, DP can only be substi-tuted for Q if there is no change in R by the perfusion pressure in the microcirculation, i.e. if there is no coronary distensibility (higher resistance with diminished perfusion pressure). Both described assumptions related to a constant and minimal resist-ance6are probably erroneous,7 because in comparison with the experimental model, it is hard to achieve maximal vasodilation using the mentioned i.v. dose of adenosine, and because the cor-onary circulation actually seems to be distensible.8 Furthermore, a variable degree of collateral flow to the vascular area down-stream of a stenosis influences the distal coronary pressure with identical variability, e.g. the FFR obtained during vascular occlusion can be as high as 0.8 despite the occlusion.8 Finally, and for research puposes, FFR should be calculated by accounting for

central venous pressure (CVP) as (Pd – CVP)/(Pao – CVP). In the presence of a tight stenosis without collateral flow, the assumption that CVP ¼ 0 mmHg leads to an overestimation of FFR.

Clinical implications

For routine clinical and not for research purposes, the present study’s conclusion can be kept in mind: ‘the accuracy of QCA in predicting functionally significant FFR is limited . . . ’. In addition, it is probably reversible: ‘The accuracy of FFR in predicting a structu-rally significant stenosis by QCA is limited . . . ’. Thus and despite the findings of the FAME study,9we may retreat to a lean practice of PCI by taking into account the patient’s history, stress ECG, and an experienced estimate of the angiographic stenosis severity in several planes.

Conflict of interest: none declared.

References

1. White C, Wright C, Doty D, Hiratza L, Eastham C, Harrison D, Marcus M. Does visual interpretation of the coronary arteriogram predict the physiologic impor-tance of a coronary stenosis? N Engl J Med 1984;310:819 – 824.

2. Harrison D, White C, Hiratzka L, Doty D, Barnes D, Eastham C, Marcus M. The value of lesion cross-sectional area determined by quantitative coronary angiogra-phy in assessing the angiogra-physiologic significance of proximal left anterior descending coronary arterial stenoses. Circulation 1984;69:1111 – 1119.

3. Seiler C, Kirkeeide RL, Gould KL. Basic structure – function relations of the epicar-dial coronary vascular tree. Basis of quantitative coronary arteriography for diffuse coronary artery disease. Circulation 1992;85:1987 – 2003.

4. Yong ASC, Ng ACC, Brieger D, Lowe HC, Ng MKC, Kritharides L. Three-dimensional and two-Three-dimensional quantitative coronary angiography, and their pre-diction of reduced fractional flow reserve. Eur Heart J 2011;32:345 – 353. First published on 12 August 2010. doi:10.1093/eurheartj/ehq259.

5. Gould K. Pressure– flow characteristics of coronary stenoses in unsedated dogs at rest and during coronary vasodilation. Circ Res 1978;43:242 – 253.

6. Pijls NHJ, van Son JAM, Kirkeeide RL, de Bruyne B, Gould KL. Experimental basis of determining maximum coronary, myocardial, and collateral blood flow by pressure measurements for assessing functional stenosis severity before and after percuta-neous coronary angioplasty. Circulation 1993;86:1354 – 1367.

7. Spaan J, Piek J, Hoffman J, Siebes M. Physiological basis of clinically used coronary hemodynamic indices. Circulation 2006;113:446 – 455.

8. Seiler C. Collateral circulation of the heart, Vol 1. London: Springer; 2009. 9. Tonino P, De Bruyne B, Pijls N, Siebert U, Ikeno F, van’ t Veer M, Klauss V,

Manoharan G, Engstrøm T, Oldroyd K, Ver Lee P, MacCarthy P, Fearon W, FAME Study InvestigatorsFractional flow reserve versus angiography for guiding percutaneous coronary intervention. N Engl J Med 2009;360:213 – 224.

Figure

Figure 1 Coronary angiogram with magnified views (middle rows) and schematic drawings (bottom rows) of the left coronary artery obtained at different time points during the exam (columns A, B, and C)

Références

Documents relatifs

The errors in P 00 can be de- tected and corrected automatically using a simulation-based graph edit distance, as introduced in [4] to fix a faulty service to cooperate

The relexificationists argue that the slave populations spoke, on their arrival in the colonies, a variety of different African languages and that under the influence of the

L’archive ouverte pluridisciplinaire HAL, est destinée au dépôt et à la diffusion de documents scientifiques de niveau recherche, publiés ou non, émanant des

Thus each mathematical symbol (G, L,. .) has a single definite meaning inside each exercise, but might have different meanings from an exercise to another.. It is not required to

The winding number of a closed curve around a given point is an integer representing the total number of times that curve travels anti-clockwise around the point.. The sign of

In particular, studies taking into account only developed countries in the sample or adopting econometric methods different from ordinary least squares (OLS), instrumental

In Chapter 3, ‘ The syntax of dislocations and focus fronting ’, Lo´pez discusses the syntactic position of CLRD, CLLD and FF.. He argues for pragmatic features to be assigned at

Evaluation of size: the size of each project in the samples of the three categories (“active,” “dormant” and “inactive”) was evaluated using the sloccount 12 tool, that