Sensing of Oxygen Concentration in a Microfluidic Device mimicking Liver 3D Microarchitecture

N/A
N/A
Protected

Academic year: 2021

Partager "Sensing of Oxygen Concentration in a Microfluidic Device mimicking Liver 3D Microarchitecture"

Copied!
3
0
0

Texte intégral

(1)

HAL Id: hal-02387492

https://hal.archives-ouvertes.fr/hal-02387492

Submitted on 30 Nov 2019

HAL is a multi-disciplinary open access archive for the deposit and dissemination of sci- entific research documents, whether they are pub- lished or not. The documents may come from teaching and research institutions in France or abroad, or from public or private research centers.

L’archive ouverte pluridisciplinaire HAL, est destinée au dépôt et à la diffusion de documents scientifiques de niveau recherche, publiés ou non, émanant des établissements d’enseignement et de recherche français ou étrangers, des laboratoires publics ou privés.

Sensing of Oxygen Concentration in a Microfluidic Device mimicking Liver 3D Microarchitecture

Manon Boul, Satomi Matsumoto, Marie Shinohara, Yasuyuki Sakai, Teruo Fujii, Anne Dubart-Kupperschmitt, Eric Leclerc, Bruno Le Pioufle

To cite this version:

Manon Boul, Satomi Matsumoto, Marie Shinohara, Yasuyuki Sakai, Teruo Fujii, et al.. Sensing of Oxygen Concentration in a Microfluidic Device mimicking Liver 3D Microarchitecture. MicroTAS, Oct 2019, Basel, Switzerland. �hal-02387492�

(2)

SENSING OF OXYGEN CONCENTRATION IN A MICROFLUIDIC DEVICE MIMICKING LIVER 3D MICROARCHITECTURE

Manon Boul1,2*, Satomi Matsumoto3, Marie Shinohara3, Yasuyuki Sakai4, Teruo Fujii3,5, Anne Dubart Kupperschmitt2, Eric Leclerc5 and Bruno Le Pioufle1

1ENS Paris Saclay, Cachan, France

2UMR 1193, INSERM, Villejuif, France

3Institute of Industrial Science, The University of Tokyo, Tokyo, Japan

4Department of Chemical System Engineering, Graduate School of Engineering, The University of Tokyo, Tokyo, Japan

5LIMMS, CNRS - Institute of Industrial Science, The University of Tokyo, Tokyo, Japan

ABSTRACT

We designed a microfluidic structure which closely reproduces liver microarchitecture, constraining primary rat hepatocytes at a high density and in three dimensions (3D), and in which a gradient of oxygen can be generated. The device includes an oxygen sensitive membrane that could map the oxygen consumption of hepatocytes.

KEYWORDS: Microfluidics, Liver, 3D microtissues, Organ-on-a-chip, Oxygen gradient, Oxygen sensor

INTRODUCTION

Compared to classical two-dimensional cell culture, microfluidic devices or/and 3D culture conditions were evidenced to increase the period of time during which primary hepatocytes retain their functions [1]. Moreover, microfluidic techniques offer the opportunity to mimic the in vivo hepatocyte zonation, by subjecting hepatocytes to oxygen gradients [1-2]. Such oxygen gradients that can be estimated by numerical simulations, were recently experimentally assessed using an oxygen sensitive fluorescent membrane [3]. We proposed to include the oxygen sensitive membrane within a miniaturized fluidic device mimicking several hepatic cords in series, and inducing a gradient of oxygen on those. Moreover each of those hepatic cord units was inducing 3D organization of hepatocytes, due to the 72 µm height of culture chambers in which they can aggregate.

EXPERIMENTAL

The microfluidic structure is composed of hepatocyte chambers and medium-conveying channels, separated from each other by small channels (5x5x40 μm³) (Figure 1a, structure inspired from [4], with higher chambers to induce the 3D organization of cells). These elements respectively mimic hepatocyte cords, liver sinusoids, and the space of Disse. Compared to only one chamber devices proposed in [4], twenty hepatocyte chambers are arranged in series in our design, in a way to induce an oxygen gradient. Indeed, because of the consumption of oxygen in every chamber, an oxygen gradient can be induced between the first and last chamber reached by the medium.

Simulation of the oxygen mapping was performed using finite element analysis. An oxygen sensitive membrane [3] was incorporated at the bottom of the microfluidic structure and the polydimethylsiloxane (PDMS) surface was made impermeable to gas. The device outlets were then closed to prevent gas from entering. The evolution of cell morphology and oxygen concentration was followed by imaging the device in bright field and at 650 nm for 20h.

RESULTS AND DISCUSSION

A gradient of oxygen in the chambers could be evidenced, depending on the flow rate conditions (Figure 1b).

With a low flow rate (50 nl/min), a variation of 30% in oxygen concentration between the first and the last chambers can be observed, compared to 50% in vivo between periportal and pericentral hepatocytes [2].

By carefully choosing the flow rate, it is thus possible to reproduce in vivo oxygen conditions. The oxygen gradient was then characterized experimentally. The cells were loaded in the successive chambers (Figure 2a) and the membrane was shown to be sensitive enough to detect the oxygen consumption of primary rat hepatocytes after 20h on the device (Figure 2d). Such approach, here used successfully to detect the oxygen consumption in

(3)

b)

0 0.025 0.05 0.075 0.1 0.125 0.15 0.175 0.2

0 5 10 15 20

Oxygen concentration in the center of chamber (mol/m³)

Chambers, from n 1 to n 20 Oxygen concentration in hepatic chambers

2500 nl/min 1000 nl/min 500 nl/min 250 nl/min 50 nl/min

Hepatic

chamber n°1 n°2 n°3

a) b)

liver micro-architecture, is very promising for the characterization of oxygen concentration on a chip, for the reproduction of liver zonation purpose.

Figure 1: a) zoom on the microfluidic device structure, composed of 20 hepatic chambers in series, b) computational simulation of the oxygen concentration in the middle of every hepatocyte chamber, from chamber n°1, which is the first one reached by the medium, to chamber n°20, the last one reached by the medium, for different flow rates. Red line: hypoxic limit.

Figure 2: culture of primary rat hepatocytes in the microfluidic device containing the oxygen sensing membrane, in static conditions, and without gas exchange with the outside environment. In bright field at t = 0h a) and at t = 20h b), and at 650 nm at t = 0h c) and t = 20h d).

ACKNOWLEDGEMENTS

This work was performed thanks to the financial support of iLite, funded by the french National Agency for Research, in the investment for the future program and referenced as ANR-16-RHUS-0005.

REFERENCES

[1] Huh D., Hamilton GA, Ingber DE, “From 3D cell culture to organs-on-chips”, Trends Cell Biol., 2011, 21, 745.

[2] Kietzmann T., “Metabolic zonation of the liver: The oxygen gradient revisited”, Redox Biol, 2017, 11, 622.

[3] Matsumoto S., Leclerc E., Maekawa T., Kinoshita H., Shinohara M., Komori K., Sakai Y., Fujii T.,

“Integration of an oxygen sensor into a polymethylsiloxane hepatic culture for two-dimensional gradient characterization”, Sensors and Actuators B: Chemical, 2018, 273, 1062.

[4] Lee P.J., Hung P.J., Lee L.P., “An artificial liver sinusoid with a microfluidic endothelial-like barrier for primary hepatocyte culture”, Biotechnol Bioeng., 2007, 97, 1340.

CONTACT

* M. Boul; phone: +33-6-69-50-00-58; manon.boul@ens-paris-saclay.fr t = 0h

a)

100 µm

t = 20h b)

100 µm

c)

100 µm

t = 0h d)

100 µm

t = 20h

Références

Documents relatifs

To this end, the DSX:: EGFP transgenic line (thereafter, DSX) [18] was used, as it comprises a combination of fluorescent markers that allow the differentiation between male and

In KATO-III CCs, DNA demethylation with 5-aza still induced a 3-fold activation (Q-PCR) of MUC4 mRNA expression and inhibition of histone deacetylation with TSA induced a

Comme on vient de le remarquer, la sphère de dimension 3 se découpe en deux boules et le nœud de trèfle se restreint dans chaque boule à l’enchevêtrement de la figure ci-dessus..

As the oxygen concentration is decreased from 21 % to a value near the extinction limit of the fire, the fuel mass loss rate decreases by a factor of about 2.5 from 21 g.. The

The linear extrapolation of obtained points shows that the water masses associated with a relative phosphate minimum present characterisics of the mixing of

Fluxes of organic carbon from the sediment traps deployed in the vicinity of the channel or on the Angola margin are compared to mineralization rates calculated from benthic

We also include the possible diffusion limitation in oxygen transfer observed in extreme regimes involving parameters such as alveolar and venous blood oxygen partial

The characteristics of pressure profiles (arrival time, positive overpressure, positive impulse) could be expressed in terms of reduced variables according to the diffusion time