Human Liver Tissue Characterization Using Backscattered Ultrasound Signals Mean Scatterer

Spacing

Ahmed BENYAHIA a, b, Abdelhafid KADDOUR a, Ahmed BAHI AZZOUOUM a, *

a University of Science and Technology of Oran Mohamed Boudiaf (USTO-MB), City, Oran 31000, Algeria

b Research Center in Industrial Technologies CRTI P.O.Box 64, Cheraga 16014 Algiers, Algeria

* Corresponding author. E-mail addresses: bahiahmed12@gmail.com

Abstract—Using the backscattered ultrasound signals, we studied the periodicity of in vitro healthy and pathological liver tissue, we utilized the mean scatterer spacing (MSS) as a parameter of tissue characterization, estimated by: Spectral Auto-Correlation (SAC), Spectral Correlation of the autoregressive model (AR-SAC) and the Discrete Wavelet Technique (DWT), in comparison between this three method, the result of simulation confirm that DWT with Fourier Transform together (DWT with FT) can provide a very effective tool to extract the MSS from such structure and characterize the tissue.

Keywords—Periodicity; Liver tissue; Mean scatterer spacing;

Backscattered Ultrasound Signals

I. INTRODUCTION

The arrangement in the soft organs such as the liver, kidneys and spleens is a systematic repetition of a basic structural unit (triads’ porticos in the liver, the nephrons in the kidney, and spleen trabecular). It is shown that this structure present a pseudo-periodic centers dispersed, resulting in a consistent periodicity in the data of the backscattered signal [1].

The distance between any two consecutive scatterers consistent in the ultrasonic signal is called "Inter Scatterer Spacing -ISS- ". The spatially ordered collection of all these individual ISS for a given ultrasonic signal is called the distribution of the ISS. The mean value of the distribution of the ISS is known "Mean Scatterer Spacing -MSS-" [2-4].

It is shown that the MSS can be used as a characteristic of tissues [4-7, 9-10]. The modification of the natural periodicity of liver tissue indicates pathologies thereof. However, there are important classes of pathologies where the periodicity is interrupted at the inner of a localized volume of a tissue.

Among these classes there is the metastatic liver tumor, liver cirrhosis …

The analysis by signal processing of biomedical data can give a true picture of the imbalances that occur in biological tissues. This analysis is based on classical techniques such as

Fourier analysis (autocorrelation, Cepstrum method, autoregressive model ...) [11-13] or relatively new techniques, wavelet technique (discrete and continuous).

In this present work, a liver tissue simulation model is first proposed and defined. The second section is dedicated to the theoretical presentation discrete wavelet DWT methods. Then, MSS is determined using simulation, finally; comparison between results obtained using classical signal processing and DWT method is given.

II. THE SIMULATION MODEL OF TISSUE

The mathematical model y (t) representing the structure of tissues is given as follows:

= − + − (1)

In this equation, the first summation represents the semi- regular structure with ai being the amplitude of the ith wave reflected by a reflector located at a temporal distance equal to τ1i, and L being the number of reflectors. The second summation is the random structure, with bi being the amplitude of the wave reflected by a diffuser located at a time distance equal to τ2i and M is the number of diffusers.

Some suggestions were made on the model given by equation (1) as follows:

1) The amplitudes ai and bi are random variables uniformly distributed in the interval [0, l].

2) The time between two reflectors of the semi-regular structure ∆τ1i1i 1i-1 follows a Gaussian distribution (equation 2) mean µ and variance σ.

= 1

√2 − −

2

(2)

3) The time between two diffusers of the irregular structure

∆τ2i2i –τ2i-1 follows a Rayleigh distribution given as follows:

(2)

= − 1

2 ! (3)

with u(t) is the unit step, the mean and variance are respectively given by:

=

" and

= #2 −

"

$

, m being a constant.

Fig. 1 represents different simulation structures, namely semi-regular, and irregular composite of a semi-regular and irregular structure with parameter of simulation are given in table 1.

TABLE I. SIMULATION PARAMETERS OF BACKSCATTERED SIGNAL MODEL.

Parameter Designation Value

MSS, µ Mean scatter spacing 2µs

N Number of samples 1024 points

µr the mean time spacing of the reflectors 2µs µd the mean time spacing of the diffusers 0.1µs σ standard variance of the temporal spacing µ/100, µ/50,

µ/30, µ/10.

SNR Signal to noise ratio 10dB

SNR_s Signal to noise ratio of the composite structure 5dB

ai Amplitudes of reflectors [0 :1]

bi Amplitudes of diffusers [0 :1]

Based on the equation (1), we cite two major issues that we can to work, the degree of regularity of the semi-regular structure and the power of the irregular structure, in this work we are interested in the degree of regularity of the backscattered signal.

The degree of regularity of the semi-regular structure is measured by the variance σ of the time interval between two reflectors. We shall see that the regularity of the structure affects the efficiency of the studied methods.

Thus, for the structure is semi-regular, the σ variance must be very small compared with the mean µ (mean spacing).

This factor (degree of regularity) plays a crucial role because the variation σ allows us to encompass the different real models.

III. DISCRETE WAVELET TRANSFORM FOR MSS ESTIMATION The discrete wavelet transform (DWT) is based on the continuous version, unlike the latter, the DWT uses a scaling factor and a translation discretized. Called dyadic discrete wavelet transform any base wavelet working with a scale factor u = 2i. It is clear that the discrete wavelet transform is implemented in practice in any digital system (PC, DSP, MAP has µP ...) [14].

The multi-resolution analysis is used to analyze a signal into different frequency bands, allowing a view of the finer to coarser.

The ultrasound signal is written as follows:

% =

&

% +

'

% + ( % ,

%* [1, -] (4)

N is the length of y(n). The term w(n) represents white Gaussian noise. The terms yr(n) and yd(n) represent the regular structure and the irregular structure (diffusion) of the backscattered signal respectively.

Using the property of linearity of the discrete wavelet transform, the equation (4) is given by,

/ 0 % = / 0 & % + / 0 ' % + / 0( % ,

1 = 1,2, … 3

(5)

Where the terms / 0 & % , / 0 ' % and / 0( % represent respectively the wavelet transform of & % ,

' % and ( % at the decomposition scale j.

The problem of the signal information extraction is equivalent to the problem of extracting / 0 & % from the sequence / 0 % in the presence of the component / 0 ' % + / 0( % .

We can apply the denoising algorithm and signal reconstruction in the field of Mallat wavelet [15-16] for the signal y(n), even in the presence of measurement and speckle noise.

(a) (b)

(c)

Fig. 1. Different structures of backscattered signal model: (a) semi-regular structure, (b) irregular structure and (c) composite structure.

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8

x 10-5 -1

-0.5 0 0.5 1 1.5

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8

x 10-5 -1.5

-1 -0.5 0 0.5 1 1.5 2

Amplitude Normalisée

0 0.5 1 1.5 2 2.5 3 3.5

x 10-5 -0.8

-0.6 -0.4 -0.2 0 0.2 0.4 0.6 0.8 1 1.2

(3)

The separation is performed in the wavelet scale space, exploiting the behavior of wavelet transform at different scales decomposition j.

Wavelets that are often used in processing of discrete one- dimensional signal are Daubechies wavelets. For fast wavelet transform (DWT), functions are defined by a set of indices which is designated as the "wavelet coefficients of the filters."[14].

Applying the decomposition of backscattered signal described above according to equation (1) by the wavelet transform.

For six scales j = 1, 2... 6, in left we show the approximation information (a1, a2, a3, a4, a5 and a6) and in right, detail information (d1, d2, d3, d4, d5 and d6). It is observed that the periodicity information becomes more and more obviously with increasing scales. This is because the regular structure yr(n) is the dominant low frequency bands.

Therefore, we are interested by the wavelet decomposition structures at high levels (low frequency) instead of directly measuring the MSS from the original signal.

We note that the frequency becomes clearer as the scale j increases, until level 6 and above, the information begins to lose from the highest levels.

We notice that the levels A4, A5, D5 and D6 inform well on the frequency information exists in regular structure of backscattered ultrasound signal, in the following, the ultrasound signal is decomposed (SNR_S = 5 dB, SNR = 20dB and µ/50) and calculating frequency spectrum for levels A4, A5, D5, and D6 of the wavelet transform to determine the MSS.

The calculation of frequency spectra of levels A4, A5, D5 and D6 allow us to extract the frequencies they contain, one notices the presence of a peak corresponding to the value of 0.5MHz in all spectra (common peak), the inverse of the value of 2µs period originally stated in the model of the ultrasound signal, so we conclude that with the presence of two types of noise signal, the decomposition of the wavelet transform allow us to extract the MSS.

Graphs (a), (b), (c) and (d) in fig.3 have a peak value of 0.5MHz for the four selected levels A4, A5, D5 and D6

respectively, and it is easy to decide the MSS value especially for level A4, it proves to be sufficient to extract the MSS value from this level for accuracy and not to miss the exact value of MSS proposed testimony other levels A5, D5 and D6.

IV. COMPARISON OF DISCRETE WAVELET TECHNIQUE AND OTHER METHODS FOR MSS ESTIMATION

We present in this section a comparison of the methods SAC-AR [8] and the method using the discrete wavelet technique while maintaining the same parameters of a suitable human tissue, which are shown in Table 1.

The figure (Fig. 4) shows a comparison between the SAC- AR and TOD method, we studied the effect changing the value of the variance in the ultrasound simulated signal tell in Table 1 and calculate the value of MSS and study the influence on the results of applying these methods.

The results prove that the estimation of MSS by the discrete wavelet technique is better in all models (for to σ = µ / 100, σ

= µ / 50, σ = µ / 30 and σ = µ / 10). The TOD can seem to be sensitive to the variation of σ. Table 2 shows the different values of the estimated MSS. The values estimated by the TOD method are close to the true value of MSS.

TABLE II. MSSCOMPARISON OF SAC-AR AND DWT METHODS (*105).

σ=μ/100 σ=µ/50 σ=µ/30 σ=µ/10 SAC-AR 5.03

± 0.03

5.04

± 0.04

5.14

± 0.07

5.50

± 0.43

DWT 5.03

± 0.01

5.003

± 0.01

5.013

± 0.02

5.03

± 0.06 V. CONCLUSION

We conclude that the discrete wavelet transform can be effectively used to indicate the value of the mean scatterer spacing, it offers a finer level of signal analysis and to better adapt to local properties. Therefore it become an efficient tool Fig. 3. signal y(n) and its components of Discrete Wavelet

Transform (details and approximations) using Db9 mother wavelet

Fig. 2. Frequency Spectrum of levels: (a) D5 (b) D6, (c) and A4 (d) A5.

100 200 300 400 500 600 700 800 900 1000

-0.50.51.501

100 200 300 400 500 600 700 800 900 1000

-10 1

a1

100 200 300 400 500 600 700 800 900 1000

-0.4-0.20.20

d1

100 200 300 400 500 600 700 800 900 1000

-1012

a2

100 200 300 400 500 600 700 800 900 1000

-0.50 0.5

d2

100 200 300 400 500 600 700 800 900 1000

-0.20.20.40.60

a3

100 200 300 400 500 600 700 800 900 1000

-2 0 2

d3

100 200 300 400 500 600 700 800 900 1000

0 0.2 0.4

a4

100 200 300 400 500 600 700 800 900 1000

-0.4-0.20.20.40

d4

100 200 300 400 500 600 700 800 900 1000

0.2 0.4

a5

100 200 300 400 500 600 700 800 900 1000

-0.2 0 0.2

d5

100 200 300 400 500 600 700 800 900 1000

0.2 0.3 0.4

a6

100 200 300 400 500 600 700 800 900 1000

-0.2-0.10.10

d6

0 0.5 1 1.5 2 2.5 3

x 106 -20

-10 0 10 20 30

(a)

Amplitude

0 0.5 1 1.5 2 2.5 3

x 106 -20

-10 0 10 20 30

(b)

0 0.5 1 1.5 2 2.5 3

x 106 -20

-10 0 10 20 30

(c)

Fréquence (Hz)

Amplitude

0 0.5 1 1.5 2 2.5 3

x 106 -20

-10 0 10 20 30

(d)

Fréquence (Hz)

(4)

in signal processing for diseases detection from biomedical signals, as in this case, the calculation of MSS of a liver tissue being to distinguish between a healthy tissue or pathological.

However, using the discrete wavelet technique with the Fourier transform together can provide a very effective tool to extract this parameter of such structure.

REFERENCES

[1] U.R.Abeyratne, X Tang, “Ultrasound scatter spacing based diagnosis of focal diseases of the liver,” Biomedical signal processing and contol2,science direct ,Elsevier,13 Mar.2007.

[2] B. Bouhadef, A. Benramdane, “Application du traitement de signal à la détermination de l’espacement moyen entre diffuseur d’un milieu hétérogène,” USTO PFE ELN ,2001/2002.

[3] J. U Quistgaard, “signal aquisition and processing in medical diagnostic ultrasound,” IEEE Signal process. Mag, pp67 75, 1997.

[4] V. Roberjot., S. Bridal. Lori., P. Laugier., and G. Berger., «Absolute backscatter coefficient over a wide range of frequency in a tissue- mimicking phantom containing two populations of scatterers», IEEE Trans. Sonic and Ultr., vol 43, no. 5, September, 1996.

[5] J. Ophir., T. H. Shawker., N. F. Maklad., J. G. Miller., W. Flax Stephen., P. A. Narayana., and J. P. Jones., "Attenuation estimation in reflection:

progress and prospects", Ultrasonic Imaging, vol 6, pp. 349-395, 1984.

[6] G. Berger., P. Laugier., M. Fink., and J. Perrin., “Optimal precision in ultrasound attenuation estimation and application to the detection of Duchene Muscular Dystrophy carriers”, Ultrasonic Imaging, vol 9, pp.

1-17, 1987.

[7] J. J. Faran, Sound scattering by solid cylinders and spheres. J. Acoust.

Soc. Am. 23: p. 405- 418, 1951.

[8] Abdelhafid Kaddour and al. “High-order statistic for mean scatterer spacing estimation for human tissues characterization”, Acoust. Sci. &

Tech. 32, 1, The Acoustical Society of Japan, 2011.

[9] F. M. Inssana and D. G. Brown., "Acoustic scattering theory applied to soft biological tissues", in Ultrasonic Scattering in Biological Tissues., Boca Raton, FL: CRC press. p. 75- 124, 1993.

[10] M. Meziri, W. C. A. Pereira, A. Abdelwahab, C. Degott, P. Laugier, « In vitro chronic hepatic disease characterization with a multiparametric ultrasonic approach », Ultrasonics 43 (2005) p.305-313, 2005.

[11] L. Fellingham ,F.G. Sommer, “ultrasonic characterization of tissue structure in the in vivo human liver spleen,” IEEE transasSonics ultrasonic, vol.31,pp.418 428,1984.

[12] R. Kuc., and M. Schwartz., “Estimating the acoustic attenuation coefficient slope for liver from reflected ultrasound signals “, IEEE Trans. Sonic and Ultr., vol SU-26, no5, September, 1979.

[13] K. A. Wear, B. A. Garra, and T. J. Hall., Measurements of ultrasonic backscatter coefficient in human liver and kidney in vivo. J. Acoust.

Soc. Am :p. 1852-1857, 1995.

[14] I. Daubechies « Ten Lectures on Wavelets ». Number 61 in CBMS-NSF Series in Applied Mathematics. SIAM, Philadelphia, 1992.

[15] Y. Meyer Ondelettes et Opérateurs, I: Ondelettes, II: Opérateurs de Calderon-Zygmund, III: (with R. Coifman), Opérateurs multilinéaires.

Hermann, Paris, 1990.

[16] S. Mallat, Une exploration des signaux en ondelettes. Les Edit. de

l'Ecole Polytechnique, 2000.

Fig. 4. MSS (in Hz) estimated using ACF-AR, TOD to (a) σ = µ / 100 (b) σ = µ / 50, (c) σ = µ / 30 and (d) σ = µ / 10.

0 5 10 15 20

3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5

(a)

Realizations

MSS*10e5

0 5 10 15 20

3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5

(b)

Realizations

0 5 10 15 20

3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5

(c)

Realizations

0 5 10 15 20

3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5

(d)

Realizations perfect

WDT SAC-AR

Références

Documents relatifs

La différence principale entre les deux est que la valeur du paramètre de forme de Nakagami, lié à la densité de diffuseurs, sature à 1 pour les concentrations élevées

Abstract : Using the backscattered ultrasound signals, we studied the periodicity of in vitro healthy and pathological liver tissue, we utilized the mean scatterer spacing (MSS) as

This algorithm iterates back and forth using the current estimated parameter in order to decompose the observed data and thus to increase the likelihood of the next

green and red highlights. The green and red highlights sur- round the fat and lymph-node tissue, respectively. Figure 1 reveals that the histology images are slightly smaller in

Analysis of the relationship between backscattered P-band radar signals and soil roughness... Analysis of the relationship between backscattered P-band

Over bare soils in agricultural areas, IEM simulates the backscattering coefficients (  ° HH ,  ° HV ,  ° VV ) using the characteristics of the sensor (incidence

While interictal and preictal discharges are transient events, epileptic seizures are complex dynamic events with variable oscillatory components (Baud et al., 2018; Eissa et

No double-membrane vesicles (DMVs) were detected in these liver tissues, but structures suggestive of a membranous web composed of small clusters of single-membrane vesicles