Analytical Performance of a

Bone Specific ALP Automated Immunoassay

Etienne Cavalier, Pierre Delanaye

Departments of Clinical Chemistry and Nephrology, CHU Sart-Tilman, Liège, Belgium

Background

Serum bone–specific alkaline phosphatase (bone ALP), a marker of bone turnover, has been proposed as an alternative biochemical target for clinical management of CKD-MBD patients. We assessed the analytical performance of the IDS-iSYS Ostase BAP (IDS, UK) automated assay and compared this method to the manual Ostase BAP Immunoenzymetric assay [EIA] (IDS, UK). The observed ranges for apparently healthy children and reference intervals for pre-menopausal,

Method Comparison

iSYS = 1.05 x (EIA) + 0.25 Pearson r = 0.994. Good correlation between the method with Mean difference of 6.2%

Reference Intervals

 Males: 5.7 – 32.9 µg/L (n = 120). reference intervals for pre-menopausal,

post-menopausal women and men were established.

Materials and Methods

• Accuracy profile: determined with serum pools. • Linearity: verified with 2 sets of high/low serum. • Method comparison: 116 samples [6.1-75.7µg/L]. • Reference intervals: males, pre-menopausal and

post-menopausal.

• Paediatric observed ranges: healthy <20 yrs. old.

Results

Precision profile

Total imprecision %CV between ranged from 4.0 to 6.3% in 4 pool levels with concentration between 10..2 - 62.7 µg/L. Linearity Observed = 1.05 x (Expected ) – 1.89; R2= 0.997.  Males: 5.7 – 32.9 µg/L (n = 120).  Pre-menopausal: 4.7 – 27.0 µg/L (n=54)  Post-menopausal:5.5 – 27.1 µg/L (n=79)

Paediatric Observed Ranges

Conclusion

The IDS-iSYS Ostase BAP assay performed well; it showed suitable characteristics to achieve acceptable precision and accuracy.

The method comparison between the IDS-iSYS and the EIA showed an excellent relationship.

The IDS-iSYS Ostase BAP is a suitable tool in clinical laboratories. Find this poster on ORBI: http://orbi.ulg.ac.be/handle/2268/157674

Références

Documents relatifs

A better integration of epidemiological and clinical surveillance with the virological surveillance would increase the usefulness of each of these systems and ensure an

The aim of this study was to compare the incidence of difficult tracheal intubation, by means of the IDS score, between lean and obese patients assessed with classical bedside tests

While finding duplicates in ORCID is, in general, as complex as finding au- thor duplicates in research literature (i.e., none or limited ground truth), we can leverage the

Dans ce papier, nous utilisons la décomposition modulaire afin de réduire la taille du graphe de flux utilisé pour générer les signatures.. Définition 1 Soit G = (V, E) un

exit Exit configuration mode and return to global configuration mode reset-signatures Reset signatures settings back to the default configuration show Display

SS-IDS is best suited for the newly installed web servers which doesn’t have large number of requests in the data set to train the IDS and can be used on top of currently used

We are interested in the ways in which Indigenous ways of teaching and learning have been and continue to be communicated and constructed through adult-child interactions

L'IDS peut alors interagir avec le firewall afin qu'il mette à jour ses règles d'accès pour empêcher tout contact avec l'éventuel pirate. Il faut cependant ce méfier de