Automated Measurement of Total 25 Vitamin D

on the IDS-iSYS System

Introduction: Most health professionals know vitamin D for its critical role in increasing the efficiency of dietary calcium absorption and preventing rickets in children. In the past decade, emerging research has also uncovered other important relationships between vitamin D and bone health as well as new information indicating it could potentially play a significant role in strengthening the body’s defenses against chronic and autoimmune diseases and in controlling the growth of cells (both normal and cancerous). These new evidences have fuelled the global demand for 25-OH D testing. Clinical laboratories are under pressured to seek automated reliable platforms for 25-OH D analysis to keep up with the demanded volume. Immunodiagnostics System (IDS) has commercialized a direct, fully automated, co-specific immunoassay for 25-OH D for use on the IDS-iSYS system. We seek to assess the analytical performance of the IDS-iSYS 25-OH D assay, and to compare this method to LC-MS/MS, DiaSorin Liaison, and DiaSorin RIA methods. Methods: We determined the repeatability and reproducibility on 6 pools, ranging from 6.2 ng/mL to 71.2 ng/mL, the limit of quantification (LoQ) and the accuracy profile of the IDS-iSYS 25-OH D method. A total of 88 samples, selected to represent a wide range of 25-OH D concentrations [5.6-85.6ng/mL], was assayed by each method.

Results: Intra and inter-assay coefficients of variation (CV) were <11% at 6.2 ng/mL and <8% at 71.25 ng/mL. The LoQ was established at 6.2 ng/mL. The accuracy profile between 6.2 and 71.2ng/mL showed that the risk that one result falls out of the ±30% acceptance limit is <5%, indicating that the method is completely validated in this range. The actual assay working range is 5 to 140ng/mL. A comparison of IDS-iSYS 25-OH D with LC-MS/MS yielded the Passing Bablok fit equation: IDS-iSYS = 1.03 x (LC-MS/MS) + 1.50. The corresponding equation with Liaison was: IDS-iSYS = 0.94 x (Liaison) + 2.41, and with RIA: IDS-iSYS = 1.00 x (RIA) – 0.62.

E. Cavalier, A.Carlisi, AC. Bekaert, JP. Chapelle

Department of Clinical Chemistry, University of Liège, CHU Sart-Tilman, Liège, Belgium

Conclusion : The IDS-iSYS 25-OH D is a robust method,

presenting analytical performances in accordance with the

clinical expectations. The IDS-iSYS 25-OH D values were in

excellent agreement with those determined with LC-MS/MS,

DiaSorin Liaison and DiaSorin RIA. This method demonstrated

suitable characteristics as a high throughput, fully automated,

co-specific 25-OH D assay for clinical laboratories.

SA

490

Mean of the pools (ng/mL) Calculated β-expectation

tolerance limits (ng/mL)

Risk to fall of the limits (%)

6.240

[ 4.721 , 7.759]

1.982

9.590

[ 7.104 , 12.08]

2.879

15.57

[ 13.39 , 17.75]

0.1948

28.15

[ 20.79 , 35.50]

3.235

39.22

[ 31.81 , 46.63]

1.023

71.25

[ 55.11 , 87.39]

1.928

Références

Documents relatifs

En effet, bien qu’il n’y ait pas à ce jour d’études d’impacts accessibles sur les répercussions de l’exploitation des ressources naturelles sur l’environnement au

Comment nourrir 10 milliards de personnes en 2050 avec une seule terre ?.. Comment nourrir 10 milliards de personnes en 2050 avec une

Dès lors, c'est une logique de prudence qui devrait présider au développement et à l'utilisation des IP : loin d'une approche mécaniste et outrancièrement simplificatrice,

/ La version de cette publication peut être l’une des suivantes : la version prépublication de l’auteur, la version acceptée du manuscrit ou la version de l’éditeur. For

Rapport 4 - Guide de l'usager: Système informatisé SOCOUS (Base de données relationnelle et système d'information géographique sur les sources de contamination et

Cette section de la méthodologie se sub-divise en trois étapes: la production d'un indice normalisé de l'état (INN) de la neige, la caractérisation du fond de scène en

d’histoire de la tradition de ce texte que pour les difficultés liées à l’anonymat des auteurs, reconstruire la lecture qu’ont pu en faire les clercs médiévaux à l’origine

We are interested in the ways in which Indigenous ways of teaching and learning have been and continue to be communicated and constructed through adult-child interactions