Comparison of two automated assays for the determination of cobalamin

in serum

L. Vranken

1

, C. Bertholet

1

, P. Lukas

1

, E. Cavalier

1

Introduction

1 Department of Clinical Chemistry, University of Liège, CHU Sart-Tilman, B-4000 Liège, Belgium.

Cobalamin measurement is routinely performed for the screening of vitamin B12 deficiency. Unfortunately, there is no consensus on standardization and no consensus on thresholds used for deficiency. Hence, there is variability between the results of different Vitamin B12 methods leading to potentially different clinical interpretation of the results. In this study, we decided to compare automated assays for cobalamin determination within two different populations, normal and at risk of deficiency.

Material and Methods

Cobalamin was measured with two different automated immunoassay analyzers: Abbott Architect i1000sr and Roche Modular e602. We compared the two immunoassays in two separate cohorts. The first one consisted of 80 healthy people and the second one consisted of 108 metformin-treated type 2 diabetic patients. Serum cobalamin concentrations were classified as deficient (<200 ng/L), borderline (200-300 ng/L) and sufficient (>300 ng/L).

Statistical analysis:

MedCalc software, version 12.7.7.0 (Oostende, Belgium) was used to perform the Passing-Bablok regressions and Bland-Altman plot.

Conclusions

There was a systematic and a proportional bias between both

assays, found similarly in the two cohorts. Results obtained

with Abbott were lower than those obtained with Roche and

could potentially lead to earlier supplementation. Due to

variability

of

cobalamin

results

between

different

manufacturers, cut-off values for deficiency should be verified

for each laboratory to avoid misclassification.

Email: [email protected]

Diabetic population: Passing and Bablok was Abbott = -23,7 (95% CI -41; -6.6)+ 0,85 (95% CI 0.80; 0.90)x Roche

References

In the healthy population, Passing and Bablok regression was Abbott = -25.3 (95% CI -70; 20.7)+ 0.91 (95% CI 0.81; 1)x Roche. Concordance correlation coefficient (CCC), Pearson’s correlation (ρ) and Bias correction factor Cb (accuracy) were 0.9105, 0.9611 and 0.9574, respectively.

In the diabetic population, Passing and Bablok was Abbott = 23,7 (95% CI 41; -6.6)+ 0,85 (95% CI 0.80; 0.90)x Roche. CCC, ρ and Cb were 0.9190, 0.9785, 0.9391, respectively. In the diabetic patients’ cohort, 15 patients were borderline with Abbott vs. 3 patients with Roche. Three patients were considered as deficient with Abbott whereas the Roche assay did not classify any patient as being deficient.

Results

Annual Meeting of the Royal Belgian Society of Laboratory Medicine 2019

C5

• Aparicio-Ugarriza, R., Palacios, G., Alder, M. & González-Gross, M. A review of the cut-off points for the diagnosis of vitamin B12 deficiency in the general population. Clin. Chem. Lab. Med.

53, 1149–59 (2015).

• Vogeser, M. & Lorenzl, S. Comparison of automated assays for the determination of vitamin B12 in serum. Clin. Biochem. 40, 1342–5 (2007).

• Thorpe, S. J. et al. An International Standard for holotranscobalamin (holoTC): international collaborative study to assign a holoTC value to the International Standard for vitamin B12 and serum folate. Clin. Chem. Lab. Med. 54, 1467–72 (2016).

@

Diabetic population: Bland-Altman plot of Abbott and Roche total Vitamin B12 (ng/L). Mean difference (solid line) ± 2SD (dashed line) is shown.

Healthy population : Passing and Bablok regression was Abbott = -25.3 (95% CI -70; 20.7)+ 0.91 (95% CI 0.81; 1)x Roche Healthy population: Bland-Altman plot of Abbott and Roche total Vitamin B12 (ng/L).

Références

Documents relatifs

The wild-type, single mutant F290K, and double mutants (A289P − F290L, A289P−F290C, and A289P−F290I) of NpAS were subjected to molecular dynamics (MD) simulations (20 ns) in

The resulting articles (Tomm, 1987a, 1987b, 1988) laid the groundwork for Tomm’s later work on ethical postures: his concept that counsellors can more explicitly choose how

In einer Interviewstudie auf einer Intensiv- station und einer akutgeriatrischen Abtei- lung in Basel berichteten 86% der befrag- ten Pflegenden und 75% der Ärzte über Situationen,

With these methods (METM and FVTM) the value of the field at one characteristic point (a node or an integration point) of the new mesh is a function of the value of field on the

On the other hand, the terms computed by sensitivity analysis repre- sent the emission change of each individual source and their sum is equal to the combined impact of all sources

Using the electron-beam-induced current (EBIC) technique it is possible to determine in an experimental way the site of the space charge region and of the electrical p-n

after abortion (group 2A), the test detected more positive sera from infected ewes 16. lambing normally than any other assay (group 2B,

Parameters are computed here taking into account all the events at once, meaning that “climatic” (keeping in mind only 4 months of data are used) estimates are studied. An event