• Aucun résultat trouvé

BRING-UP PROCEDURE

Dans le document 8560/8561/8562 (Page 138-144)

BRING-UP PROCEDURE

The following bring-up procedure is intended to be used af-ter you have completed the troubleshooting procedure for catastrophic failure. This procedure enables you to fully check out the power supply and restore it to working order.

Preconditions

Before you perform the bring-up procedure, make sure that you have completed the following steps:

• The fan housing on the rear panel of the 8560 and the power supply top shield have been removed.

• The Inverter Board has been investigated for failures according to procedures outlined under ·Catastrophic Failures", and has been repaired as needed.

• The rear panel power switch is off, and the input capac-itors (the large cans below the Inverter Board) have discharged to a safe voltage. Discharging occurs within 5 minutes after line power is removed.

• The 8560's line voltage selector switch is properly set.

After verifying that all preconditions are met, refer to Fig.

12-7 and prepare the test set-up. Table 12-3 reflects the equipment required to perform the bring-up procedure.

Power Supply-8560/8561/8562 Service

Table 12-3

Equipment Required for Bring-Up Procedure

Equipment Recommended Type

Soldering iron/solder 8000 F tip temperature/63%

tin, 37% lead, rosin core Needle-nose pliers Sturdy, with plastic grips Screwdriver Phillips for #6 and #2 screws Voltmeter/diode meter Fluke Model 8030A or

equivalent

Oscilloscpe TEKTRONIX 465 or

equiva-lent

Isolation transformer Must have at least 8 A output capability

Variable autotransformer Variaca with at least 8 A out-put capability

External power supply Must be able to supply

+ /

-17 Vat 400 mA

Extender board kit Part No. 067-1058-01 a Variac is a registered trademark of the GenRad Corporation After installing the test equipment, check that all instrument fuses are good.

8560

I V.hm~' ps=

(4-25Vdc. 120 Vac rms)

ri7~~~~~~~~~~~

External Power Supply

> ± 17V. >400mA

12-10

Isolation Transformer

1=1 :::::8A Output

Capability

Variac :::::8A Output

Capability w/Wattmeter and Voltmeter

Regulator Board Secondary Board

Fig. 12-7. Test equipment set-up.

Probe

Vert. in Ext. Trig.

4759·32

Power $.ljpply-8560/8561/8562 Service

Procedure

This procedure contains enough detail to allow you to check out every primary supply characteristic except output ripple, input line frequency range, and overvoltage shutdown.

I

WARNING

I

Use insulated tools and probes at all times and ob-serve the supplementary cautions contained in the text. Stored charge in the large capacitors of the power supply can cause severe bums and/or elec-tric shock even when the supply is turned off.

1. Remove the Inverter Board from its slot and insert the Inverter Extender Board into the same slot. Then insert the Inverter Board into the extender.

2. Make sure the external power supply is off. Connect leads from the external supply through a 3-pin harmon-ica connector to the three square pins on the Regulator Board (J7075). The connections should be to + Volt-age, Ground, and -VoltVolt-age, from left to right (viewed from the rear of the 8560). The supply should be set between +17 V and +40 V (nominally +20 V).

3. Make sure that the rear panel power switch and the Variac power switch are off. Turn on the external power supply. Measure the voltage from the + 15 V

Bring-up Procedure

test point (TP7076) to ground (TP7025), and from the -15 V test point (TP7077) to ground on the Regulator Board. If any measured voltages are not within 5 per-cent of a test pOint's assigned voltage, you must repair the standby regulators (U1060, U1070) on the Regula-tor Board before further testing. Make sure that the external supplies are not operating in a current-limited mode. The external supply current limits should be at least 400 mA each.

4. Measure the voltage from Vref (TP7066) to Ground Sense (TP7056). Adjust the trimmer on the Regulator Board (R7067) until Vref is within 1 percent of +5 V.

5. Remove the undervoltage shutdown jumper (J7047) from the Regulator Board.

6. Connect the external trigger input of the scope to the trigger testpoint (TP7039) on the Regulator Board. Set the scope's sweep rate to 10 ,."s/cm. Be sure that the external trigger is selected.

7. Connect the vertical input of the scope through a probe to the base of Q3010 on the Inverter Board. Connect the ground clip of the probe to the emitter of the se-lected transistor. Set the vertical sensitivity to 2 V/cm.

8. Turn on the 8560 dc power switch. The scope should display the waveform shown in Fig. 12-8 or a waveform shifted from this by 180 degrees. If the waveform is correct, proceed to step 10.

Fig. 12-8. Q3010/Q2010 base-emitter voltage waveform.

12-11

Bring-up Procedure

9. If the voltage does not drop from 0.7 V to -2 V, but remains near ground or 0.7 V until the negative pulse, one of the driver transistors (05074 or 05075) is prob-ably bad. If the negative pulse is less than about 3 V from base to peak, and the transition time of the falling edge of the positive pulse is less than 1 I-LS, the other driver transistor is probably bad.

To check these transistors, change the scope's vertical sensitivity to 1 V/cm, and connect the probe to each of the drain tabs in turn while connecting the ground clip to TP6069. The scope should display the waveform shown in Fig. 12-9, or a waveform shifted from this by 180 degrees. If the large positive pulse has a flat top or is severely attenuated, the transistor is defective. Re-place any bad transistors with good ones. If, after you have checked these transistors, the circuit still does not seem to be working properly, proceed to step 11.

10. Check the base-emitter voltage waveform on 02010.

The waveform should have the same shape as that of 03010 but shifted 180 degrees in phase. If the wave-form does not meet this description, check for failures, as in step 9.

11. If the supply is being brought up after a catastrophic main inverter failure~ check the waveforms at the gates of 05074, 05075, and 05076. The waveforms at 05074 and 05075 should be square waves with a pe-riod between 50 I-LS and 70 I-LS, a high level at

+

15 V, and a low level at ground. These two waveforms should also be 180 degrees out of phase. The waveform at 05076 should be a 1 I-LS negative-going pulse from

+

15 V to ground every 25 I-LS to 35 I-LS.

If any voltage is more than a few volts away from the proper level, the transistor you are checking is probably defective even if its drain voltage is correct. However, if the transistor under investigation checks out good, U2030 on the Regulator Board is bad. If no voltage is should also be plugged in to provide a minimum load on the

+

12 V supply. However, do not bring up the supply with maximum load on the 5 V output.

13. Connect the vertical input probe to the emitter of 03010 and the ground clip to the end of C2030 closest to the rear of the instrument. (C2030 is the large, white plastic film capacitor parallel to the long side of the finned heatsink.) Set the vertical sensitivity to 50 V/cm.

14. Turn the Variac output voltage control to zero. Turn on both the Variac and the 8560 rear panel power switch.

WARNING

I

Use extreme caution when servicing circuitry on the . Inverter Board. Dangerous voltages exist on the Inverter Board when the 8560 is connected to line voltage.

15. Increase the Variac output voltage slowly to about 20 V. The scope should display the waveform shown in Fig. 12-10a. The wattmeter should read near zero. If so, slowly turn the Variac output voltage towards 90 V.

If the waveform seems to be unstable, slowly increase the Variac voltage while watching its wattmeter. The waveform should stabilize, eventually increasing in fre-quency. The wattmeter should not read much over 140 W if there is a minimum load in the instrument (see step 12). If anything indicates failure, turn the Variac down immediately. The supply may shut off at some line voltage below 90 Vrms, but if nothing else suspi-cious occurs, keep turning the Variac up until its output is 90 V. If the supply is operating correctly, when the Variac output is 90 V, the scope will display the wave-form shown in Fig. 12-10b.

Power Supply-8560/8561/8562 Service Bring-up Procedure

Fig. 12-9. Q5074/Q5075 drain voltage waveform.

Fig. 12-10. T1011 primary winding waveforms.

12-13

Bring-up Procedure

16. If the waveform doesn't stabilize or rise in frequency, a fault probably exists in the main inverter control cir-cuitry on the Regulator Board (U3030, U5030D, U5060B, U6010, U2040, U2030, and U3040B). If the wattmeter indicates excessive power consumption, a shorted diode probably exists on the secondary side of power transformer T1011. (See the Secondary Board schematics.) If the supply won't come on when the Variac output is 90 V, turn to "The Supply Won't Turn On" (ignore step 2 of that procedure).

17. Connect the negative lead of your voltmeter to Ground Sense (TP7065), and the positive lead to each of the testpoints listed in Table 12-4.

Table 12-4

Regulator Board Testpoints Testpoint

Testpoint Name Number Expected Voltage +5 V Sense TP7061 +4.85 V to +5.15 V

+24 V TP7057 +21.8 V to +23.2 V

+12 V Sense TP7043 + 11 .4 V to + 12.6 V

-12 V TP7015 -11.4 V to -12.6 V

Power Supply-8560/8561/8562 Service

18. If the + 12 V output is incorrect, see "Troubleshooting the + 12 V Secondary Regulator" in the Troubleshoot-ing Guide, earlier in this section. If either the +24 V or the -12 V output is incorrect, see "Troubleshooting the +24 V and -12 V Secondary Regulators", also in the Troubleshooting Guide.

19. Replace the undervoltage shutdown jumper (J7047) on the Regulator Board. Place the voltmeter leads on the +5 V Sense testpoint (TP7061) and Ground Sense testpoint (TP7056). Vary the Variac voltage between 132 Vrms and 90 Vrms. The waveform on the scope should remain stable, and neither the positive or nega-tive peak should go over 200 V. The voltmeter reading should remain within 1 percent of 5 V.

20. If the supply is being brought up after a catastrophic failure in the main inverter, and the supply fails to stay on over an input voltage range of 90 to 132 Vrms, the

+

12 V secondary regulator may be regulating improp-erly. To check the + 12 V secondary regulator, set the scope's vertical sensitivity to 50 mV/cm and change the coupling to ac. With the supply running, observe the waveform at pin 3 of U2010 on the Regulator Board. If this waveform does not match the waveform in Fig.

12-11 (especially if the sawtooth is smaller or nonexis-tent), U5030 on the Regulator Board is probably faulty and should be replaced.

Fig. 12-11. +12 V regulator reference waveform.

12-14

Power Supply-8560/8561/8562 Service

21. Turn off the external power supply. The 8560 supply should stay on without change. Turn the Variac voltage below 90 Vrms until the supply shuts off. It should shut off at approximately 80 to 85 Vrms. Turn the Variac voltage back up until the supply turns back on. This should occur below 90 Vrms. The supply should turn on cleanly, with only one chirp and no sputtering.

22. Check the overcurrent shutdown circuitry (U6040 on the Regulator Board, U3110A and U211 0 on the Sec-ondary Board). Connect a 20-ohm resistor from the -12 V test point (TP7015) to ground (TP7025). The supply should shut down immediately and chirp faintly every second or so until the resistor is removed. If the supply does not shut down, don't do any further overcurrent shutdown testing until the problem in the -12 V shutdown circuitry is repaired.

If the supply does shut down, connect a 0.75-ohm re-sistor from + 12 V sense (TP7043) to ground. Then, connect a 16-ohm resistor from + 24 V (TP7057) to ground. In both cases, the supply should shut down and chirp until the resistor is removed. Finally, connect

Bring-up Procedure

a 0.1-ohm resistor from the +5 V power bus to ground near the edge connector on the Regulator Board. The supply should again shut down and chirp until the resis-tor is removed.

23. Turn off the 8560 from the rear panel. Unplug the 8560 and wait 5 minutes for the input capacitors to discharge to a safe voltage. Then remove all probes from the sup-ply boards.

24. Remove the Inverter Board and its extender, then in-sert the Inverter Board into its edge connector. Be cer-tain that the Secondary and Regulator Boards are in the correct slots. Replace the top shield over the supply with its four screws. If the fan was disconnected, plug its connector into J7017, being careful to observe the correct polarity. Replace the fan housing with its eight screws. Reconnect all hard disk power connector(s) and replace any circuit boards removed during this bring-up procedure. Replace any outside covers and feet that were removed from the instrument. The sup-ply should now be in satisfactory working order.

12-15

8560/8561/8562 Service

Section 13

Dans le document 8560/8561/8562 (Page 138-144)

Documents relatifs