• Aucun résultat trouvé

Detection of pollutants dissolved in water by Raman spectrometry probe

N/A
N/A
Protected

Academic year: 2021

Partager "Detection of pollutants dissolved in water by Raman spectrometry probe"

Copied!
3
0
0

Texte intégral

(1)

HAL Id: hal-00728946

https://hal.archives-ouvertes.fr/hal-00728946

Submitted on 20 Mar 2018

HAL is a multi-disciplinary open access

archive for the deposit and dissemination of sci-

entific research documents, whether they are pub-

lished or not. The documents may come from

teaching and research institutions in France or

abroad, or from public or private research centers.

L’archive ouverte pluridisciplinaire HAL, est

destinée au dépôt et à la diffusion de documents

scientifiques de niveau recherche, publiés ou non,

émanant des établissements d’enseignement et de

recherche français ou étrangers, des laboratoires

publics ou privés.

Detection of pollutants dissolved in water by Raman

spectrometry probe

M.D. Fontana, K. Ben Mabrouk, T.H. Kauffmann, Mario Marchetti

To cite this version:

M.D. Fontana, K. Ben Mabrouk, T.H. Kauffmann, Mario Marchetti. Detection of pollutants dissolved

in water by Raman spectrometry probe. IMEKO XX, Sep 2012, Busan, South Korea. �hal-00728946�

(2)

XX IMEKO World Congress Metrology for Green Growth

September 9−14, 2012, Busan, Republic of Korea

DETECTION OF POLLUTANTS DISSOLVED IN WATER BY RAMAN

SPECTROMETRY PROBE

Marc D. Fontana

*

, K. Ben Mabrouk, T. Kauffmann and M. Marchetti

*Laboratoire Matériaux Optiques, Photonique et Systèmes Unité de recherche commune Université de Metz – Supélec, 2 rue E. Belin 57070 Metz - France

Tel.: +33 (0)3 87 37 85 58; e--mail: fontana@metz.supelec.fr

Laboratoire Régional des Ponts et Chaussées de Nancy, 71 rue de la grande Haie 54510 Tomblaine, France

Abstract: The paper is devoted to the abilities of Raman probe to detect simultaneously several salts diluted in water.

It has shown that this probe can be used for in situ monitoring of pollutants in waste water.

Keywords: Raman sensor, optical detection, pollutants.

1. INTRODUCTION

The control of the quality of water is a growing need in our modern societies in order to insure the access of water in healthy conditions to an always rising number of people through the world. This leads to the obligation to detect all pollutants in flow water, rivers, basins, lakes… [1]. This challenge requires new and efficient techniques which are able to provide in situ and quick measurements.

Nitrates, sulphates, and phosphates have to be specially purchased since they are very water soluble and lead to negative incidence in the environment. They arise generally from fertilizers used in agriculture [2,3].

Numerous techniques have been used to detect anionsSO42,

NO , and 3 HPO42 such as electric conductivity [4] and calorimetric methods [5,6]. Despite the performances, these methods are unable to respond to the increasing conditions needed for the soil and water preservation. Indeed, technologies allowing long-term continuous monitoring are more and more required to detect in situ several contaminants within a fairly good accuracy and a high speed. So far, the analysis methods generally used are sensitive but off-lines and time-consuming which therefore need the prelevment of solutions.

Here our aim is to show the abilities of Raman sensors to detect the presence of several pollutants dissolved in water and to determine their concentration as well. Raman scattering is a well known spectroscopy technique useful to study the vibrational properties of molecules in a material.

The main advantages of the technique are the speed of measurement and the fact that it does not require any special preparation of the medium to be analysed. Moreover Raman systems involving diode lasers CCD detector and fiber- optics probes become more and more cheap, robust and portable. Only recently Raman sensors have been developed to provide some parameters linked to the change in the Raman spectrum. Thus was reported a new Raman sensor of NaCl in aqueous solution [7-9]. This probe is founded on the

detection of the change of OH stretching peak which is caused by the introduction of sodium chloride in water.

2. RESULTS

Here we performed Raman measurements in various dissolved salts of nitrates, sulphates, and phosphates and the peculiar influence of anion, and cation and H2O on the spectrum of each substance was analysed. Measurements have been done with a 532 nm exciting line using a spectrometer Kaiser RXN1 within a resolution of 2 cm-1. The main question is the ability for such a sensor to discriminate between the different substances. It is seen in Figure 1 that an own signature can be detected by Raman probe for each kind of anion in different spectral ranges.

This means that peculiar lines of the Raman spectrum can be used as a fingerprint-like of the presence of SO42,NO 3 orHPO42, as a consequence, the concentration of each salt in water can be deduced from an appropriate calibration.

960 980 1000 1020 1040 1060 1080 1100 200000

400000

960 980 1000 1020 1040 1060 1080 1100 500000

1000000 1500000

960 980 1000 1020 1040 1060 1080 1100 500000

1000000 1500000

ννννssss((((P OP OP OP O3333

) ) ) (9 8 9))

Raman Shift (cm-1)

(NH4)2HPO4 50 g/l νννν(1111(((S OS OS OS O4 4 4 4 ))))

(9 8 1 ) (NH4)2SO4 50g/l

νννν1111((((N ON ON ON O3333

) )) )

(1 0 4 7)

Intensity (a.u)

NH4NO3 50g/l

Fig. 1. Specific Raman signatures of nitrate, sulphate and phosphate anions.

Figure 2 shows the concentration dependence of the peak associated to the vibrational mode within the anion NO . 3 The calibration of the Raman probe was achieved by recording the integrated scattered intensity of this line as a function of the concentration varying between 5 and 0.5 g/l and prepared by titration. A linear law was obtained.

(3)

0 1 2 3 4 5 0.00

0.05 0.10 0.15 0.20 0.25

1020 10301040 1050 1060 1070 1080 0

5000 10000 15000 20000 25000

νννν(1111(((AAAA1 1 1 1 ))))

Intensity (a.u)

Raman shift (cm-1) 5g/l 3g/l 2.5g/l 2g/l 1.5g/l 1g/l 0.5g/l

N ON O N ON O

3333----

Anion concentration (g/l)

Normalized peak area

Fig 2. Nitrate content dependence of integrated Raman intensity. The insert shows the behavior of the relevant part of the Raman spectrum exhibiting the ν1 vibrational line as a function of concentration.

In a second step the complex mixed solutions in waste water including several pollutants have been studied and analysed from the data sources obtained in the pure substances. An example of spectrum is shown in Figure 3. Although the spectrum is more complicated an assignment of main peaks could be achieved so that a separation of the signature of each pollutant is nevertheless possible.

0 300 600 900 1200

800000 1000000 1200000 1400000 1600000 1800000 2000000

P

(8 9 0 c m- 1)

N

(7 8 0 c m- 1)

S

(9 8 2 c m- 1)

Raman shift (cm-1)

Intensity (u.a)

Fig. 3. Raman spectrum recorded in waste water. Arrows indicate the species N (nitrate), P (phosphate) and S (sulphate).

For each particular purchased pollutant, the influence of additional salts was studied and the sensitivity and the accuracy of our probe were evaluated. The influences of several measuring parameters such as laser source (power and wavelength) acquisition time, contact or contactless probe…were successively investigated

3. CONLUSION

We reported on the study of Raman probe of various pollutants like sulphates, phosphates, nitrates dissolved in water. Each salt was first separately and carefully investigated Then the abilities to get a poly-detection sensor was discussed. The various parameters allowing improving the resolution, the accuracy and the signal/noise were investigated. This probe was shown to be promising to get a simultaneous detection of several species in a mixture.

4. REFERENCES

[1] European Directive 98/83/CE.

[2] J. Pelley, "Is coastal eutrophication out of control?"

Environ. Sci. Technol., vol. 32, pp. 462A-466A, 1998.

[3] N. Berenzen, R. Schulz, M. Liess, "Effects of chronic ammonium and nitrite contamination on the macro invertebrate community in running water microcosm", Water Research, vol. 35, no.14, pp. 3478-3482, 2001.

[4] Thermo Orion, "Sensing the future", 2002.

[5] W. Yao, R. Byrne, R. W. Bury, "Determination of Nanomolar Concentrations of Nitrite and Nitrate in Natural Waters Using Long Path Length Absorbance Spectroscopy", Environ. Sci. Technol., Vol. 32, pp. 2646, 1998.

[6] D. M. W. Peat, I. D. McKelvie, G. P. Matthews, P. M.

Haygarth , P. J. Worsfold, "Rapid determination of dissolved organic phosphorus in soil leachates and runoff waters by flow injection analysis with on-line photo- oxidation. " Anal. Proc., vol. 32, pp. 437, 1995.

[7] R. Claverie, M. D. Fontana, I. Durickovic, P. Bourson, M. Marchetti and J-M. Chassot, "Optical Sensor for Characterizing the Phase Transition in Salted Solutions", in Proc. of Sensors, vol. 10, pp. 3815-3823, 2010.

[8] I. Durickovic, M. Marchetti, R. Claverie, P. Bourson, J- M. Chassot, and M. Fontana, "Experimental study of NaCl aqueous solutions by Raman spectroscopy: Towards a new optical sensor", Appl. Spectr., vol. 64, pp. 853-7, 2010.

[9] T. Kauffmann and M. D. Fontana, "Optical sensor of salt concentration: uncertainty evaluation." Sensors& Actuators:

B. Chemical.(to be published).

Références

Documents relatifs

In this study, we report a multi-receptor approach based on a panel of different reporter cell lines (Table 1) to establish toxicity profiles in environmental samples and its

Using synthetic fluid inclusions, namely closed vacuoles fabricated inside one solid host by hydrothermal processes, a Raman study was performed to examine how a superheated solvent

In order to built prediction models for these anions, a PLS analysis was applied on several salt solutions with different anion concentrations given in mol/l.. Moreover in order

Abilities of Raman spectroscopy to detect sulphates, nitrates and phosphates diluted in waterK. Ben

Raman probe of chemical species diluted in water: technique reliability and estimation of uncertainty.. Kawther Ben Mabrouk 1,2 ,

Here is shown how Cl - influence the OH stretching band of the water spectrum and how its concentration is determined using chemometrics methods applied on Raman spectra recorded

The most commonly used AOPs for the removal of organic pollutants from water is based on the Fenton’s reagent (an aqueous mixture of Fe 2+ and H 2 O 2 that produces

119 Un crible de protéines entomotoxiques réalisé avec le puceron du pois Acyrtosiphon pisum a permis de mettre en évidence que la plupart des inhibiteurs de protéases possèdent