Beamline Magnet Polarity Indication

Henry Nebrensky

Brunel University

(2)

Background

The MICE beamline can be used to work with either positively or negatively charged particles. The

changeover is achieved by reversing the (DC) current flow through ALL beamline magnets (Q1-9 and D1&2).

We thus need a way to ensure that all magnets have been wired correctly. The end result is specified by

“Convention on MICE Beam Line Magnet Polarities” (MICE Note 198). This defines the electrical input needed for the correct magnetic performance of each magnet, and identifies “positive” and “negative” terminals at the

magnets . However, it also mandates that changes are

implemented by swapping the leads at the power supplies

(not least because access is easier).

(3)

The problem

Hence in order to keep track of things, the Convention applies the labels “positive” and

“negative” (and associated colouring) to the cables ; thus for a “negative beamline” set up for negatively charged particles, the “negative cable” is attached to the positive terminal of the supply.

Unfortunately the two separate meanings of positive and negative (the action of the beamline, and the output of a DC power supply) are not readily

appreciated by the electricians (note the layers of

tape):

(4)

Q4 MPS (as it was before)

(5)

Hence…

It would seem preferable to have the cables labelled using a scheme that does NOT directly incorporate (or imply) the words “positive” and “negative”, or the associated colours of red and black.

Beyond this, the aim here is neither to contradict nor supersede MICE Note 198.

One possibility would be to assign a unique ID to each

cable. While this might be useful for other things,

it would be far too cumbersome when requesting or

verifying a “simple” change of beam polarity.

(6)

Current Solution

The aim is a scheme that makes it straightforward to open a cabinet in the MICE Hall and confirm that the wiring is correct with a simple visual check.

One (pair of) cables has been labelled with the word

“beamline”.

Then, for a positive beamline the “beamline” should be connected to the +ve supply terminal.

For a negative beamline, the “beamline” should be connected to the -ve supply terminal.

The label is applied over green tape, making it easy to spot.

(Proposal accepted verbally by Martin Hughes on

18/3/2010 and at MICO meeting on 12/4/2010)

(7)

Updated Power Supply End

(8)

Magnet end

Although MICE Note 198 specifies that the leads should never be disconnected at the magnet end, there is always the possibility of a mistake. It

would thus be prudent also to label the magnet end of the cables.

I suggest using labels with similar appearance but

titled “reference” applied to the cable connected to the “positive” magnet terminal.

(Can’t use “beamline” as it could get mistaken as indicating the beamline polarity or cables to swap, as at the power supply end. Can’t use e.g. “+” as this could be interpreted as being the cable carrying the +ve voltage.)

(9)

Updated Q9

(10)

Magnet Polarity Summary

At the MAGNETS:

The “reference” leads are permanently connected to the “positive” terminal

At the SUPPLIES:

If the “beamline” leads are connected to the positive output, have a “positive beamline”

If the “beamline” leads are connected to the negative output, have a “negative beamline”

Simples!

10

(11)

Status

Labels as described have been applied at all beamline magnet power supplies (Q1-9, D1&2).

Labels have been applied at all the magnets.

The beamline polarity has been swapped since Marco checked a couple of the magnets directly with the gaussmeter and it’s been a long time since all were checked – thus:

the polarity of ALL magnets needs to be

re-validated once all labels are in place.

(12)

Q7 (because it’s hard to get a picture of casually checking an MPS)

Polarity setting easily visible:

Références

Documents relatifs

Job burnout Job satisfaction.. With regard to the personal situation of the participants, 76% of them are married, 10% living together/civil union, 7% are single, 6% are divorced

Les articles (saunes comptes rendus de livres) doivent être accompagnés d'un résumé de 50 à 75 mots (si possible dans une autre langue) et d'une courte biographie de l'auteur.. Les

Recent in situ data obtained with balloon-borne mass spectrometers have revealed the nature of the most abundant ions in the stratosphere.. Proton hydrates,

Address: Los Angeles, California Country : United States of America Hobby : Cinema, sports. Doc.4 – Who

Vereecken, Efficient random walk particle tracking algorithm for advective-dispersive transport in media with discontinuous dispersion coefficients and water contents, Water Resour.

general form of the potential relief (or in the presence of a magnetic field) the particle overcoming the saddle point with small kinetic energy will not come back. So,

This theoretical curve is in agreement with the experimental values in the region that the ionization waves appear. le confirm that, when the ionization waves de- velop above B1,

With the aim of studying more precisely the influence of electrostatic interactions on the approach and the adsorption, we determine the kinetics of deposition under different