ANALYTICAL PERFORMANCE OF TWO THYROGLOBULIN ASSAYS IN THE LOW

RANGE: IMPACT ON CLINICAL DECISION

Etienne CAVALIER

Department of Clinical Chemistry, University of Liège, CHU Sart-Tilman, Belgium

Introduction: Thyroglobulin (Tg) determination is used for follow-up and monitoring of patients with differentiated cancer of the thyroid. In these patients, Tg levels drop to undetectable (or very low) levels after total or near-total thyroidectomy and successful treatment with radio-labelled iodine. If a rise in Tg levels is observed during the follow-up, it can indicate a recurrence of the cancer. Thus, analytical performance of the different available Tg assays in the lower range of measurement is of importance. Unfortunately, this has been evaluated in very few studies.

Material and methods: We used two Tg automated methods:

-Beckman-Coulter Access -Roche Modular

These methods are claimed to be calibrated against the same material (CRM457).

Roche claims a limit of detection (LOD) <0.1 ng/mL and a functional sensitivity (FS) <1 ng/mL.

Beckman also claims a LOD of 0.1 ng/mL but is less clear on the FS, even if they claim a CV<10% for concentrations >1 ng/mL.

We thus defined the LOD (value significantly different from zero with a probability of 99.7%) and the FS (lowest value who gave a CV of 20% when low values samples were assayed in triplicates on 5 different days) of the two assays.

Then we defined the beta-expectations tolerance limits for 5 samples free of anti-Tg antibodies ranging from about 0.2 to 2 ng/mL by running them on triplicate during 5 different days.

Finally, we focused on 36 patients presenting low Tg levels (with no anti-Tg antibodies) to compare the results obtained with the two assays.

0.89 (0.80-0.98)

1.71 (1.57-1.86)

Pool 5

0.74 (0.69-0.78)

1.26 (1.04-1.48)

Pool 4

0.47(0.42-0.51)

0.86 (0.70-1.02)

Pool 3

0.30 (0.25-0.35)

0.40 (0.24-0.57)

Pool 2

0.22 (0.19-0.26)

0.22 (0.08-0.36)

Pool 1

Mean (upper-lower) beta-expectation

tolerance limits (ng/mL)

0.26

0.31

Functionnal

Sensitivity (ng/mL)

0.14

0.10

Limit of Detection

(ng/mL)

Beckman-Coulter

Access

Roche Modular

In other words, a patient presenting a Tg value of 0.30 with the Access could either range fro 0.25 to 0.35 with less than 5% of probability to fall out of this range. Run on Modular, this patient would have a value that could range from 0.24 to 0.57 with a chance greater than 5% to fall out of this range.

Conclusion:

The performance of the 2 instruments is acceptable in the lower range of Tg assays, but Access presents more robust results.

There is also a calibration problem with the Roche assay, even if claimed to be calibrated against CRM457. The first of the two

calibrators used to fit to the Master curve of the instrument is probably too high (around 4 ng/mL) whereas Access uses four calibrators,

the first around 1 ng/mL. Results obtained with the 2 methods are not transposable.

Références

Documents relatifs

after abortion (group 2A), the test detected more positive sera from infected ewes 16. lambing normally than any other assay (group 2B,

The collimation of the shower, the number of layers before the apex (triggering of the shower, used to probe for minimum ionizing particles), footprint (transverse lateral

2 Construis la pyramide c., en suivant la même logique, puis détermine le volume de chaque solide, en unités de volume (u.v.).. Complète le tableau, sachant que le dernier solide

L’archive ouverte pluridisciplinaire HAL, est destinée au dépôt et à la diffusion de documents scientifiques de niveau recherche, publiés ou non, émanant des

Le signe de V k est déterminé par le mode de fonctionnement de la GRV ; quand le système est excité (phase de magnétisation), nous avons +V k ; quand le système génère le

Figure 35.6 presents typical results of repeatability and linearity assays at high glucose concentration using a cellulose acetate enzyme membrane placed in a flow-through cell:

Effect of duration and temperature during the delivery of urine samples to the analytical laboratory on metabolic parameters assays... Nutrition &amp; Metabolic

From the theoretical analysis presented here, we have shown that, provided the satellite constellation does not have any degenerated geometrical properties, the dimension of