The dual beam spectropluviometer

N/A
N/A
Protected

Academic year: 2021

Partager "The dual beam spectropluviometer"

Copied!
2
0
0

Texte intégral

(1)

HAL Id: hal-01108783

https://hal.archives-ouvertes.fr/hal-01108783

Submitted on 23 Jan 2015

HAL is a multi-disciplinary open access

archive for the deposit and dissemination of

sci-entific research documents, whether they are

pub-lished or not. The documents may come from

teaching and research institutions in France or

abroad, or from public or private research centers.

L’archive ouverte pluridisciplinaire HAL, est

destinée au dépôt et à la diffusion de documents

scientifiques de niveau recherche, publiés ou non,

émanant des établissements d’enseignement et de

recherche français ou étrangers, des laboratoires

publics ou privés.

The dual beam spectropluviometer

Nawal Akrour, Laurent Barthès, Aymeric Chazottes, Jean-Yves Delahaye,

Peter Golé, Jacques Lavergnat, Jean-Paul Vinson

To cite this version:

Nawal Akrour, Laurent Barthès, Aymeric Chazottes, Jean-Yves Delahaye, Peter Golé, et al.. The

dual beam spectropluviometer. Journée Scientifique SIRTA 2014, Jun 2014, Palaiseau, France. 2014.

�hal-01108783�

(2)

Discrete view of rain :

Various studies allowed the modeling and

the simulation of rainfall as a raindrop

point process.

Lavergnat and Golé, 1998 et 2006.

The DBS also allowed some investigations

on the DSD.

Mallet and Barthès, 2009. Barthès and Mallet, 2013.

Continuous view of rain :

At the Latmos, various studies on the multifractal properties of rain are based

on measurements of the DBS.

De Montera at al., 2009 & 2010 Verrier et al., 2011 & 2013.

TECHNICAL CHOICES

MEASUREMENT GOALS METHODS USED

DIAMETER

Down to 0.1 mm

• 2 thin (2 mm) and closely spaced (2 mm) flat beams (L x l = 250 x 40 mm) at λ=0.8µm. • 16-bit digitization oversampled at 40 kHz and averaged to 10 kHz for noise reduction. • Software :

- slope detection

- baseline = averages before and after slope change

- correction for drops larger than beam thickness

- likelihood test by comparing 2 channels • Quality flag for small diameters.

FALL VELOCITY

Optimized

• Averaging of ∆t1 and ∆t2 (see diagram above).

• Interpolation for high velocities.

TIME STAMPING OF DROPS

1 millisecond

• Clocking: sensor’s sampling frequency used between PC times inserted at beginning and end of files.

ERROR REDUCTION

• Sunlight

• Water on internal glass windows • Dripping water drops

• Wind sensitivity

• External splash on sensor body • Internal splash

• Condensation

• Sunlight rejection filter. • Air circulation pump.

• Water draining slots above windows. • Very rigid cast iron support ring. • Anti-splash circular ribs and scouring rolls. • Internal vanes.

• Heated electronics and Moisture absorbers.

ERROR DETECTION

• Self-test for monitoring baselines, synchronization, and window cleanness

• Reciprocating 0.4 mm diameter rod within beam triggered daily or manually and motor driven. Induced signal variations recorded to self-test file.

A rain drop falls through both, upper and lower, beams disrupting the optical signal.

The processing algorithm allows the

simultaneous detection of the drop diameter and fall velocity at precisely defined arrival time.

Outdoor measurement signals with a small isolated drop: 0.2mm and 0.89 m/s

time Rain drop Receiver photodiodes Light source Dual slot

Upper beam signal Lower beam signal

Upper beam Lower beam Fresnel’s biprism ∆t1 ∆t2 10 ms/div

PRINCIPLE

50m cable Processing & recording software Self-test motor 16 bits A/D Converter Series/parralel interface 16 bits A/D Converter Series/parralel interface Self-test Clock Sync.

BLOC DIAGRAM

Successfully compared with other optical disdrometers.

Krajewski et al., DEVEX-disdrometer evaluation experiments : Basic results and implications for hydrologic studies, 2005, Advances in Water Resources.

FIELD EVALUATION : (DEVEX experiment)

2D homogeneity of beams and noise

TECHNICAL CHARACTERISTICS

2 mm Up Low 2 mm Up Low 2 mm Up Low 2 mm Up Low

AA

AA

AA

AA

AA

Emitter Receiver 250 mm Slots Ø 69 Ø 69 A B C D Measurement area Ribs

THE DUAL BEAM SPECTROPLUVIOMETER

N. Akrour, L. Barthès, A. Chazottes, J.Y. Delahaye,, P. Golé, J. Lavergnat and J.P. Vinson

LATMOS/IPSL/UVSQ/UPMC/CNRS

The Dual Beam spectropluviometer (DBS) measures the equivalent diameter, the fall

velocity and the time of arrival of raindrops. Its main advantage over previous optical

disdrometers is its ability to capture the whole measurement range of atmospheric

precipitating particles (down to 0.1mm raindrops).

Delahaye et al., A dual beam spectropluviometer concept, Journal of Hydrology , 2006.

Room for improvement and rejuvenating:

A 2014 IPSL funding is asked for in order to modernize the DBS (both hardware and software components).

The processing algorithm as well as the calibration process are under study to improve the quality of the measurements (and among others to reduce the noise for the small drops). Intercomparisons with other instruments at Sirta is in progress.

RELATED WORKS

Power spectra of the time series collected from july 2008 to july 2009 at the Sirta. This picture display the various scaling of the rain measurements

Références

Documents relatifs

The C++ version of the GUINEA-PIG program, guineapig++, was modified to properly track Bhabha events from dedicated external generators in the field of the colliding bunch, taking

L’archive ouverte pluridisciplinaire HAL, est destinée au dépôt et à la diffusion de documents scientifiques de niveau recherche, publiés ou non, émanant des

The technology of the tile edges matting helps to improve the situation as shown in Figure 12 for the light collection efficiency calculated for a plastic tile of the inner

Figure 1: drawing of the PHIL beamline, blue boxes are magnetic elements, orange is the RF gun, green boxes stand for vacuum pumps, charge is measured with Faraday cup after the

Shift of the vertical waist position as a function of the momentum offset of the incoming beam, measured at the Laser- Compton monitor. Squares are measured before the correction of

Many configurations composed of different injection or acceleration schemes, where different injection and accel- eration techniques have been applied, are studied in detail

L’archive ouverte pluridisciplinaire HAL, est destinée au dépôt et à la diffusion de documents scientifiques de niveau recherche, publiés ou non, émanant des

The Final Focus Test Beam (FFTB) is a prototype linear collider final focus, designed to reduce the 46.6 GeV SLAC beam to a size of 2 microns by 60 nanometers.. The FFTB has