VALIDATION AND ACCURACY PROFILES OF THE LIAISON BAP_OSTASE (DIASORIN), A NEW METHOD FOR THE DETERMINATION OF BONE ALKALINE PHOSPHATASE.

E. Cavalier (1), I. Carlisi (1), JP. Chapelle (1), E. Rozet (2)

Departments of (1) Clinical Chemistry and (2) Analytical Chemistry, University of Liège, CHU Sart-Tilman, B-4000 Liège, Belgium.

Background. We wanted to validate the Liaison BAP_OSTASE and compare this new method with Access Ostase (Beckman-Coulter).

Methods. Determination of repeatability, reproducibility, functional sensitivity, recovery, linearity and establishment of the measurement uncertainty, accuracy profile and -expectation limits. Comparison with Beckman Ostase with Bland-Altman plot. Results. Intra and inter-assay CVs: <5% at 14 and 66 µg/L. Detection limit: 0.09 µg/L and functional sensitivity: <0.74 µg/L. Mean recovery and linearity estimated according the NCCLS Guidelines: between 85 and 115%. Relative expanded uncertainty: from 9.0 to 13.8% in the studied range (5.4-93.9 µg/L).

In this range, the accuracy profile shows that the risk that one result falls out of the 20% acceptance limits is <5%, indicating that the method is completely validated.

We compared 83 consecutive patients with the two methods and did not find any significant difference with the Wilcoxon test (Median (95% IC): Liaison: 12.8 µg/L(9.4–21.4) and Access: 14 µg/L (10.8–17.2)). The Bland and Altman plot shows no systematic bias between the two methods (mean difference =0.3 µg/L) but a quite important dispersion around the mean difference (mean SD:5.6 µg/L).The regression between the two methods is Access = 0.9123*Liaison + 2.0025 (R²=0.9006, p<0.001).

Conclusions. Liaison BAP_OSTASE is a robust method, with good CVs, linearity and recovery. The accuracy profile, a new approach in Clinical Biology, shows that the method is completely validated between 5.37 and 93.9 µg/L: in this range, we are sure that the values obtained are not at risk of being over 20% of the true values.

Références

Documents relatifs

The linear correlations between the anodic current of O 2 •− and the substrate concentration allowed the determination of antioxidant index values expressed by the

D’après les données disponibles au 3 janvier 2021, 2,9 millions de salariés sont couverts au moins un jour du mois de janvier 2021 par une demande d’autorisation préalable

After comparison with two methods used for asteroids, we show that the bootstrap appears a robust and pratical method for estimating the accuracy of predicted satellite positions..

The theoretical resonance curve R (9) is calculated with the hypothesis of a perfec- tly parallel and monochromatic light beam, whereas the real experimental beam has finite

As the utility of using zebrafish larvae as a model for studying tobacco product toxicity has not been evaluated, we have tested CSCs from two reference cigarettes and six

botulinum strain 109 785 (type C) or 105 256 (type D) in spiked liver samples after homogenization with either Pulsifier ® blender (Microgen, Surrey, UK) (in white) or paddle

These biochemical markers revealed a low genetic diversity (mean heterozygosity) over the screened area. The low level of diversity observed for the allozymes on a large

Interestingly, α -tocopherol may play an important role in preventing oxidative stress induced by cadmium exposure by scavenging free radicals.. Key words: HPLC, vitamin E,