CENTER FOR INTERDISCIPLINARY

RESEARCH ON MEDICINES

Laboratory for the Analysis of Medicines, Department of Pharmacy, CIRM, University of Liège

Proteomic analysis using capillary electrophoresis hyphenated with high resolution mass spectrometry:

comparison of two coupling interfaces

Marie-Jia Gou, Gwenaël Nys, Gaël Cobraiville, Alice Demelenne, Anne-Catherine Servais, Marianne Fillet

Conclusion

Fig.1: Effect of dynamic pH junction (DPJ) on peak intensity and peak width. The use of this online

preconcentration technique to analyze a bovine serum albumin (BSA) digest allowed the improvement of peak

intensity by approximately 100-fold. Moreover, we were able to inject a large amount of sample (≈ 14% of the

total volume of the capillary) and obtain satisfying peak width;

Fig.2: The composition of the sheath liquid greatly

influences the MS signal intensity. It is usually composed of a mixture of organic and aqueous solution. Effect of the sheath liquid composition on signal. This figure shows the influence of the percentage of methanol (MeOH) as well as formic acid (FA) on signal intensity (BSA digest 0.5 pmol/µL)

• The use of DPJ as preconcentration method allowed to improve 100-fold the sensitivity without sacrificing separation efficiency and peak width. • The composition of sheath liquid is of great importance to enhance the signal intensity.

• The use of EMASS-II interface (nanoflow) allowed the identification of more peptides and proteins.

• The comparison of LC-Chip and CZE showed a great complementarity resulting in a larger coverage of the proteome.

• An improvement of 4-fold and 6-fold in peak area and peak height were observed with the use of a nanoflow sheath liquid interface.

• Despite the delicate handling and lack of automatization of the EMASS-II interface, it can be considered as an useful tool in proteomics research in terms of coverage and sensitivity improvement. 20 25 30 35 40 0.00 4.50x104 Intens ity (counts ) Time (min) MeOH/H2O (50/50) + 0.1% FA MeOH/H2O (80/20) + 0.5 % FA

Fig.4: The number and uniqueness of the peptides and proteins identified by injecting 1 µg of E.coli digest was

evaluated. Two MS acquisition modes (DDA and DIA) were performed and the results obtained with Triple Tube

interface (microflow, left) and EMASS-II interface (nanoflow, right) are shown in the Venn diagrams. The complementarity of CZE with LC-Chip is also demonstrated.

Introduction

Capillary electrophoresis tandem mass spectrometry (CE-MS/MS) is gaining

interest in proteomics analysis and its use might bring complementary information in the analysis of complex proteome samples

CE and MS hyphenation is not

straightforward due to the need of two

independent electrical circuits to ensure

in-capillary separation and spray

formation.

Use of a sheath liquid to allow

separation and stable spraying

Aims of the study

Results

BSA digest (3 extracted ions) 50 pmol/µL

Injected 5s at 50 mbar

BSA digest (3 extracted ions) 0.5 pmol/µL

No DPJ, injected 500s at 50 mbar

BSA digest (3 extracted ions) 0.5 pmol/µL

DPJ, injected 500s at 50 mbar

Fig.3: Diagram of the two CE-MS coupling

interface namely the Triple tube as the

microflow sheath liquid interface (A) and the EMASS-II as the nanoflow sheath liquid

interface (B). Loss of sensitivity Dilution of the analytes from the CE effluent Flowrate of the sheath liquid (1-10 µL/min) High efficiency Small consumption of sample and solvent Orthogonality to LC

Sample enrichment

and signal intensity

• Dynamic pH junction (DPJ) as online preconcentration technique • Composition of the sheath liquid to enhance signal intensity

CE-MS coupling

interfaces

• « Triple tube » as a

microflow sheath liquid interface

• « EMASS-II » as a

nanoflow sheath liquid interface

• Use of two

complementary MS acquisition modes (DDA and DIA)

Comparison

• Number of identified peptides and proteins using a complex

proteome digest

• Gain of sensitivity due to the use of a

nanoflow interface

A

Références

Documents relatifs

Figure 2 Extracted ion electropherograms of 298.12 m/z for the analysis of RNase P1 BAP digested tRNA TrpCCA by CE-MS/MS in positive ion mode.. Figure 3 Multiple extracted

However, the better peptide distribution and the better separation efficiency achieved with the RP x IP-RP-HPLC setup permitted to identify significantly more peptides than with

Suitable interface for coupling liquid chromatography to inductively coupled plasma-mass spectrometry for the analysis of organic matrices.. 2 Comparison of Sample

L’archive ouverte pluridisciplinaire HAL, est destinée au dépôt et à la diffusion de documents scientifiques de niveau recherche, publiés ou non, émanant des

The aim of this research is to understand the chemistry of thiosalts, investigate capillary zone electrophoresis (CZE) with indirect detection and electrospray mass

The Amsterdam Preoperative Anxiety and Information Scale (APAIS) is a self-report questionnaire comprising six questions that have been developed and validated to evaluate

For the Rubber Hinge and Fabric Hinge designs, we placed the Toe, Inverted Heel Body, and Traditional Heel Body in their worst-case scenarios of static loading at

Two hypotheses might explain this: (1) spatio and/or tem- poral changes of tension wood quality (variations of a) during the gravitropic response are large and become more