S. PAEME

1

, K. MOORHEAD

1

, J.G. CHASE

2

, C.E. HANN

2

, B. LAMBERMONT

1

, P. KOLH

1

,

M. MOONEN

1

, P. LANCELLOTTI

1

, P.C DAUBY

1

, T. DESAIVE

1 1 Cardiovascular Research Center, University of Liege, Belgium

2 Centre for Bio-Engineering, University of Canterbury, Christchurch, New Zealand

This research uses a simple dynamic model of the stiffness of the valve leaflets to characterize their fundamental effects on flow and pressure. The valve is described as a non-linear

rotational spring or a ‘hinge’ with the angle change under pressure driven flow being related to the stiffness and the damping of the valve.

This model is included in a CVS model consisting of 7 elastic chambers [1].

Structural model of the mitral valve

included in a cardiovascular closed loop model

This study shows that this valvular model included in a CVS closed loop model gives physiological results.

This proof-of-concept study demonstrates that this generic structural valve model could be used to investigate the hemodynamic repercussions of several pathological conditions, such as mitral regurgitation or stenosis.

Results

Methods

Fig. 1 : Closed loop CVS model consisting of 7 elastic chambers.

Fig. 2 : Mitral valve area evolution during one cardiac cycle Fig. 3 : Transmitral flow evolution

Conclusions

Acknowledgements

References

[1] Paeme, S. et al. Mathematical multi-scale model of the cardiovascular system including mitral valve dynamics. Application to ischemic mitral insufficiency. Biomed Eng Online, 10, 86, 2011

[2] Saito, S. et al. Mitral valve motion assessed by high-speed video camera in isolated swine heart. Eur J Cardiothorac Surg, 30, 584-91, 2006

[3] Flewitt, J.A. et al. Wave intensity analysis of left ventricular filling: application of windkessel theory. Am J Physiol Heart Circ Physiol, 292,6, 2817-23, 2007

[4] Nelson, M.D. et al. Increased left ventricular twist, untwisting rates, and suction maintain global diastolic function during passive heart stress in human. Am J Physiol Heart Circ Physiol, 298, 930-937, 2010

This work was financially supported in part by the FNRS (FRIA PhD grant), the University of Canterbury, the French Community of Belgium (Actions de

Recherches Concertées – Académie Wallonie-Europe). Sabine.paeme@ulg.ac.be tdesaive@ulg.ac.be

Figure 2 shows the evolution of mitral valve area during one cardiac cycle. The maximum mitral valve values reached during the E-wave and A-wave as well as timing of these two periods correspond to the physiological data [2,3]. Figure 3 shows the evolution of transmitral blood flow during one cardiac cycle with the E-wave amplitude nearly 1.5 times bigger than the A-wave amplitude as expected physiologically [3,4].

Introduction

A minimal cardiovascular system (CVS) model including mitral valve dynamics has been previously validated in silico [1]. However parameters of this model are difficult to link with structural and anatomical components of the valve.

This research describes the integration of a structural model of the mitral valve in an existing closed-loop cardiovascular system (CVS) model.

Figure

Fig. 1 : Closed loop CVS model consisting of 7 elastic chambers.

Références

Documents relatifs

Nous avons sélectionné plusieurs sujets susceptibles d’être abordés par les sages-femmes lors de l’examen clinique des seins : l’allaitement, la modification des seins pendant

Notre solution pour dépasser cette limitation des données brutes est de poser l’hypothèse H suivante : « les comportements humains les plus atypiques à un

La gloire éphémère de ce pêcheur de bars est ternie par ce gisant à moitié vivant, un pied dans la tombe, mais seulement un, pour un voyage sans retour vers l’archipel

The web interface also comprised detailed information for the selected subset of Pixel Sets with for instance, distributions of values and quality scores and a list of individual

2b reaction of end-group of P-1 with pendent group of P-2 in the presence of a condensing agent 2c reaction of end-group of P-1 with pendent group of P-2 in presence of a

Figure 12 Measurement made with 48 cm-diameter array, 800 Hz 1/3-octave band, first 50 ms of impulse response with a hole in both sides of the wall and absorption added to

Le traitement ambulatoire d’un accès palustre non compliqué survenant chez un adulte peut se concevoir avec certaines précautions pour garantir une bonne prise en charge

Ces collyres sont indiqués dans les formes sévères de sécheresse oculaire avec kératoconjonctivite sèche résistant à un traitement lubrifiant bien conduit et