• Aucun résultat trouvé

Use of plastic phantoms for relative dosimetry

10. CODE OF PRACTICE FOR PROTON BEAMS

10.6.3. Use of plastic phantoms for relative dosimetry

The use of plastic phantoms is strongly discouraged as, in general, they are responsible for discrepancies in the determination of absorbed dose. Plastic phantoms should not be used for reference dosimetry in proton beams since the required water to plastic fluence correction factors, hpl, are not known. Nevertheless, when accurate chamber positioning in water is not possible or when no waterproof chamber is available, their use is permitted for the measurement of depth dose distribution for low energy proton beams (approximately below 100 MeV). In this case, the dosimeter reading at each plastic depth should be scaled using the fluence scaling factor hpl. It is assumed that hplhas constant value of unity at all depths.

The criteria determining the choice of plastic materials are discussed in Section 4.2.3. The density of the plastic, ρpl, should be measured for the batch of plastic in use rather than using a nominal value for the plastic type. Each

measurement depth in plastic zpl(expressed in g/cm2) must also be scaled to give the corresponding depth in water zwby

zw= zplcplg/cm2(zplin g/cm2) (42)

where cplis a depth scaling factor. For proton beams, cplcan be calculated, to a good approximation, as the ratio of csda ranges (in g/cm2) [118] in water and in plastic. The depth scaling factor cpl has a value of 0.974 for PMMA and 0.981 for clear TABLE 32. ESTIMATED RELATIVE STANDARD UNCERTAINTYa OF Dw,Q AT THE REFERENCE DEPTH IN WATER AND FOR A CLINICAL PROTON BEAM, BASED ON A CHAMBER CALIBRATION IN 60Co GAMMA RADIATION

Physical quantity or procedure

Relative standard uncertainty (%) User chamber type: Cylindrical Plane parallel

Step 1: Standards laboratory SSDLb SSDLb

ND,wcalibration of secondary standard at PSDL 0.5 0.5

Long term stability of secondary standard 0.1 0.1

ND,wcalibration of the user dosimeter at the 0.4 0.4

standards laboratory

Combined uncertainty in step 1 0.6 0.6

Step 2: User proton beam

Long term stability of the user dosimeter 0.3 0.4

Establishment of reference conditions 0.4 0.4

Dosimeter reading MQrelative to beam monitor 0.6 0.6

Correction for influence quantities ki 0.4 0.5

Beam quality correction, kQ 1.7 2.0

Combined uncertainty in step 2 1.9 2.2

Combined standard uncertainty in Dw,Q(steps 1 + 2) 2.0 2.3

aSee the ISO Guide for the expression of uncertainty [32], or Appendix IV. The estimates given in the table should be considered typical values; these may vary depending on the uncertainty quoted by standards laboratories for calibration factors and on the experimental uncertainty at the user’s institution.

bIf the calibration of the user dosimeter is performed at a PSDL then the combined standard uncertainty in step 1 is lower. The combined standard uncertainty in Dwshould be adjusted accordingly.

148

polystyrene. The procedure given in Section 10.6.1 should be followed to generate central axis depth dose distributions from the measure depth ionization distributions.

If a plastic phantom is used to measure the beam quality index, the measured quantity is the residual range in the plastic, Rres,pl. The residual range, Rres, in water is also obtained using the scaling Eq. (42).

10.7. ESTIMATED UNCERTAINTY IN THE DETERMINATION OF ABSORBED DOSE TO WATER UNDER REFERENCE CONDITIONS The uncertainties associated with the physical quantities and procedures involved in the determination of the absorbed dose to water in the user proton beam can be divided into two steps. Step 1 considers uncertainties up to the calibration of the user chamber in terms of ND,w at a standards laboratory. Step 2 deals with the subsequent calibration of the user proton beam using this chamber and includes the uncerainty of kQas well as that associated with measurements at the reference depth in a water phantom. Estimates of the uncertainties in these two steps are given in Table 32, yielding a combined standard uncerainty of 2% and 2.3% for the determi-nation of the absorbed dose to water in a clinical proton beam with a cylindrical and plane-parallel ionization chamber, respectively. Details on the uncertainty estimates for the various physical parameters entering in the calculation of kQ are given in Appendix II.

10.8. WORKSHEET

Determination of the absorbed dose to water in a proton beam

User: Date:

1. Radiation treatment unit and reference conditions for Dw,Qdetermination

Proton therapy unit: Nominal energy: MeV

Nominal dose rate: MU/min Practical range, Rp: g/cm2

Reference phantom: water Width of the SOBP: g/cm2

Reference field size: cm ×cm Reference SSD: cm

Reference depth, zref: g/cm2 Beam quality, Q(Rres): g/cm2

2. Ionization chamber and electrometer

Ionization chamber model: Serial No.: Type: ❏ cyl ❏pp

Chamber wall/window material: thickness = g/cm2

Waterproof sleeve/cover material: thickness = g/cm2

Phantom window material: thickness = g/cm2

Absorbed dose to water calibration factor ND,w= ❏ Gy/nC ❏ Gy/rdg

3. Dosimeter readingaand correction for influence quantities

Uncorrected dosimeter reading at V1and user polarity: ❏ nC ❏ rdg

Corresponding accelerator monitor units: MU

Ratio of dosimeter reading and monitor units: M1= ❏ nC/MU ❏ rdg/MU (i) Pressure P: __________ kPa Temperature T : _______ °C Rel. humidity (if known): ____ %

(ii) Electrometer calibration factorbkelec: ❏ nC/rdg ❏ dimensionless kelec=

150

(iv) Recombination correction (two voltage method)

Polarizing voltages: Vl(normal) = ____________ V V2(reduced) = _______________ V Readingsdat each V: M1= ___________________ M2= _______________________

Voltage ratio V1/V2= __________________________ Ratio of readings M1/M2= ___________

Use Table 9 for a beam of type: ❏ pulsed ❏pulsed-scanned

a0= ____________ a1= ____________ a2= ____________

_____________e,f

Corrected dosimeter reading at the voltage V1:

MQ= M1kTPkeleckpolks= ❏ nC/MU ❏ rdg/MU

4. Absorbed dose rate to water at the reference depth, zref Beam quality correction factor for user quality Q: kQ=

taken from ❏ Table 31 ❏ Other, specify:

Absorbed dose calibration of monitor at zref:

Dw,Q(zref) = MQND,w,kQ= Gy/MU

aAll readings should be checked for leakage and corrected if necessary.

bIf the electrometer is not calibrated separately, set kelec= 1.

cM in the denominator of kpoldenotes reading at the user polarity. Preferably, each reading in the equation should be the average of the ratios of M (or M+or M) to the reading of an external monitor Mem. It is assumed that the calibration laboratory has performed a polarity correction. Otherwise, kpolis deter-mined according to:

rdg at +V1for quality Qo: M+= ___________ rdg at –V1for quality Qo: M= ____________

dStrictly, readings should be corrected for polarity effect (average with both polarities). Preferably, each reading in the equation should be the average of the ratios of M1or M2to the reading of an external monitor Mem.

eIt is assumed that the calibration laboratory has performed a recombination correction. Otherwise, the factor k′s= ks/ks,Qoshould be used instead of ks. When Qois 60Co, ks,Qo(at the calibration laboratory) will normally be close to unity and the effect of not using this equation will be negligible in most cases.

f