Empirical Research On The Optimization Of The Frequency Parameters Of " Chirp " Sequences Used In Contrast Ultrasound Imaging

N/A
N/A
Protected

Academic year: 2021

Partager "Empirical Research On The Optimization Of The Frequency Parameters Of " Chirp " Sequences Used In Contrast Ultrasound Imaging"

Copied!
4
0
0

Texte intégral

(1)

HAL Id: inserm-00820840

https://www.hal.inserm.fr/inserm-00820840v2

Submitted on 9 Mar 2014

HAL is a multi-disciplinary open access

archive for the deposit and dissemination of

sci-entific research documents, whether they are

pub-lished or not. The documents may come from

teaching and research institutions in France or

abroad, or from public or private research centers.

L’archive ouverte pluridisciplinaire HAL, est

destinée au dépôt et à la diffusion de documents

scientifiques de niveau recherche, publiés ou non,

émanant des établissements d’enseignement et de

recherche français ou étrangers, des laboratoires

publics ou privés.

Empirical Research On The Optimization Of The

Frequency Parameters Of ” Chirp ” Sequences Used In

Contrast Ultrasound Imaging

Amira Zaylaa, Jean-Marc Girault, Jamal Charara

To cite this version:

Amira Zaylaa, Jean-Marc Girault, Jamal Charara. Empirical Research On The Optimization Of The

Frequency Parameters Of ” Chirp ” Sequences Used In Contrast Ultrasound Imaging. International

Conference on Advances in Biomedical Engineering (ICABME 2012), Jul 2011, Tripoli, Lebanon.

pp.59-60. �inserm-00820840v2�

(2)

 

 

1

Empirical Research On The Optimization Of The

Frequency Parameters Of « Chirp » Sequences Used

In Contrast Ultrasound Imaging

Amira ZAYLAA1

, Jean-Marc GIRAULT2

and Jamal CHARARA1 1

Department of Physics and Electronics, Faculty of Sciences I, Lebanese University, Hadath, Beabda, Beirut, Lebanon 2

Université François Rabelais de Tours, UMR-S930, CNRS ERL 3106, Tours, France

amirazaylaa@hotmail.com, jean-marc.girault@univ-tours.fr, jcharara@ul.edu.lb

Résumé – L’imagerie de contraste est un domaine très actif dans lequel plusieurs techniques de codage d’excitation ont été proposées. Dans ce travail nous nous focalisons sur le codage continu du signal ultrasonore à transmettre au milieu à explorer. Nous montrons de façon empirique qu’il existe des valeurs particulières des paramètres fréquentiels de l’excitation de type « chirp » qui optimisent le contraste de l’image ultrasonore. Nous montrons aussi que pour des microbulles nonlinéaires que les indices de modulation ne sont plus symétriques.

Abstract – The contrast medical imaging is an active research domain in which several coding techniques of the transmit have been proposed to increase the contrast. In this paper, we focus on the continuous coding of the Ultrasound transmits, such as “chirp” sequences, and their insonation on the medium under exploration. We empirically show that there are particular values for the frequency parameters of the frequency coded excitation “chirp” which optimize the contrast in contrast enhanced ultrasound images. We also show that in the nonlinear response of micro-bubbles the modulation indices are no longer symmetrical.

Keywords: Empirical Optimization, Chirp Frequency Parameters and Contrast Ultrasound Imaging.

1. Introduction

Contrast Enhanced Ultrasound imaging (CEUS) is an imaging technique involving the flushing of a contrast agent (made up of micro-bubbles) into the bloodstream prior to examination. It is in a continuous effervescence and is applied in both blood flow and tissue harmonic imaging. The contrast imaging, i.e. the imaging of the medium infused with micro-bubbles is a diagnostic tool which cannot be discarded in any clinical routine nowadays due to its vast importance.

Recently, this field is undergoing profound changes in the way of carrying out images by encoding the ultrasound waves transmitted to the medium under exploration such as tissue etc.. The reason behind coding those “exploring” signals is to improve the contrast and tissue penetration. Several types of coding are available: continuous coding such as amplitude modulation and frequency modulation such as “chirps” and discrete coding such as impulse inversion.

In this paper we will focus on the transmission of "chirp" [1, 2, 3] and more precisely determine the optimal frequency parameters that govern the instantaneous frequency law.

2. Method

The principal of contrast coded imaging consists of transmitting, by means of an insonifying transducer, a signal (see Fig. 1) of type :

x(t)=exp(-(t-t0)2/2σ2)sin((2 π fi(t))t), (1)

where the instantaneous frequency is governed by a polynomial of the form :

fi(t)=f0 + β1 t + β2 t2. (2)

After propagation in the intermediate medium towards the medium infused with micro-bubbles, a signal of non-linear properties is backscattered towards the transducer.

3. Results And Discussions

So far, the international community believes that it

Figure 1 : Time-Frequency representation of the modulated signal transmitted to the micro-bubbles.

(3)

 

 

2

 

 

Figure 2 : Energy of the backscatter at different central frequencies and modulation indices β1 the (top image) shows

the linear response of the micro-bubbles and the (bottom image) shows the non-linear response.

is enough to transmit a « chirp » (figure 1) in order to enhance the contrast in CEUS images.

In our case we will compute the energy in our way to enhance the contrast since the energy is proportional to the contrast and feasible to compute.

In this province, we show that the transmission of a « chirp » into the medium infused with micro-bubbles is not always useful especially when the central frequency of the « chirp » (given by equation (2) for β1=0 and β2=0), is close to the resonant

frequency of bubbles (~1.5 MHz).

If the central frequency of the transducer is 3MHz and its -3dB bandwidth is 2.25 MHz, it is necessary according to figure 2 to choose (β1 ≠ 0). But if the

central frequency of the transducer is 1.5 MHz and its -3dB bandwidth is 1.125 MHz, it is not necessary to frequency modulate the signal or use chirps, so we choose (β1 =0).

We show as well in figure 2 that (β2=0).

Consequently the frequency modulation index β1 in

the linear response of the micro-bubbles does not contribute to the enhancement of the energy as the

 

 

 

Figure 3 : Energy of the backscatter at different modulation indices β1 and β2 the (top image) shows the linear response of the micro-bubbles and the (bottom image) shows the

non-linear response.

values indicated by the colorbar are so small, it merely plays this vital role in the non-linear response of the micro-bubbles where energy values are higher.

The same behavior is observed in figure 3 but for (β2≠0) and f0=3MHz. We note that the symmetry of

the modulation indices in the non-linear response of micro-bubbles is lost.

Assuming again the same value of the central frequency f0=3MHz (and its -3dB bandwidth is 2.25

MHz), in the non-linear response of the micro-bubbles we notice that the maximum backscattered energy occurs at β1=-0.5e+12 MHz/s and β2=-0.5e+19 MHz/s2.

We notice that this solution is not unique since there exist another maximum at β1=0.3e+12 MHz/s and β2

=-1.0e+19 MHz/s2. This figure indicates that the automatic search for the optimal solution via simple gradient

algorithm will not converge to the supreme, thus other techniques should be considered.

(4)

 

 

3

Our study shows that taking into account both the transducer and the size of the micro-bubble, it is necessary to investigate the frequency parameters that govern the behavior of time-frequency transmit signal to optimize the energy reflected by the medium infused with micro-bubbles and thus the contrast.

Currently we are developing a procedure capable of providing the optimal settings automatically.

References

[1] Y. Sun, D.E. Kruse and K. W. Ferrara, Contrast

imaging with chirped excitation. IEEE trans. on

Ultrasonics Ferroelectrics and Frequency Control, 54(3):520–529, 2007.

[2] J. Song, J.H. Chang, T.K. Song and Y. Yoo,

Coded tissue harmonic imaging with nonlinear chirp signals, Ultrasonics, 51, 516-521, 2011.

[3] E. Barlow, A. J. Mulholland, A. Gachagan and A. Nordon, A theoretical investigation on chirp

insonification of ultrasound contrast agents,

Figure

Figure 1 : Time-Frequency representation of the  modulated  signal transmitted to the micro-bubbles.
Figure 3 : Energy of the backscatter at different modulation  indices β 1  and β 2  the (top image) shows the linear response  of the micro-bubbles and the (bottom image)  shows the

Références

Documents relatifs

Reconstruction of a longitudinal image of the radius obtained, (a) if the anisotropic model for the wave-speed in cortical bone (equation (1)) and the optimal values (table 2)

Reprinted, with permission, from S´ebastien M´enigot, Jean-Marc Girault, Iulian Voicu and Anthony Novell, Optimization of Contrast to Tissue Ratio by Frequency Adaptation in

The first result indicated that the CTR had a global maximum whatever the pressure level A 0. These maxima were obtained neither with the two- thirds of the central frequency of

Results and discussions: We show through simulations and in vitro experiments that our adaptive imaging technique gives: 1) in case of simulations, a gain of acoustic contrast which

Burns, “Pulse inversion Doppler: a new method for detecting nonlinear echoes from microbubble contrast agents,” IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency

Figure 6: Simulation of short-term harmonic powers computed on the radiofrequency line when the transmitted signal was the best found non-binary stochastic command, the best

During imaging system investigation, the cavitation maps reconstruction achieved with different passive ultrasound techniques were assessed and the passive acoustic mapping

Figure 2 shows the frequency spectrum of the RF signal from one excitation pulse and that of summed signal from two excitation pulses with 90° phase shift in tissue region