Early failure secondary to noncoronary leaflet prolapse in a stentless aortic bioprosthesis

CHERIAN, Sanjay, et al.

Abstract

Structural degeneration is an important long-term disadvantage of biologic prostheses.

However, early failure of these prostheses is uncommon and is usually caused by rapid calcification. We report the successful management of a rare case of early failure of a stentless aortic bioprosthesis (within 4 months of implantation). The patient presented with severe noncalcific aortic regurgitation secondary to prolapse of the noncoronary leaflet. In consideration of the acute nature of failure in this new-generation bioprosthesis and its unclear cause, we believe that this report, albeit of a single case, warrants some attention.

CHERIAN, Sanjay,

et al

. Early failure secondary to noncoronary leaflet prolapse in a stentless aortic bioprosthesis.

Texas Heart Institute Journal

, 2012, vol. 39, no. 5, p. 676-9

PMID : 23109766

Available at:

http://archive-ouverte.unige.ch/unige:27649

Disclaimer: layout of this document may differ from the published version.

(2)

© 2012 by the Texas Heart ® Institute, Houston

Early Failure Secondary to Noncoronary

Leaflet Prolapse

in a Stentless Aortic Bioprosthesis

Structural degeneration is an important long-term disadvantage of biologic prostheses.

However, early failure of these prostheses is uncommon and is usually caused by rapid cal- cification. We report the successful management of a rare case of early failure of a stent- less aortic bioprosthesis (within 4 months of implantation). The patient presented with severe noncalcific aortic regurgitation secondary to prolapse of the noncoronary leaflet.

In consideration of the acute nature of failure in this new-generation bioprosthesis and its unclear cause, we believe that this report, albeit of a single case, warrants some attention.

(Tex Heart Inst J 2012;39(5):676-9)

A

ortic valve replacement using stentless biological prostheses offers the ben- efits of good hemodynamics,1 without the requirement of prolonged anti- coagulation.2 Although their major drawback—structural degeneration in the long term—is well documented,3 bioprostheses rarely fail early, and, when they do, rapid calcification of the valve leaflets is the usual cause.4

We report the successful management of a rare case of early failure of a stentless aor- tic bioprosthesis within 4 months of implantation. The patient presented with severe noncalcific aortic regurgitation secondary to prolapse of the noncoronary leaflet.

Case Report

In March 2009, a 74-year-old man underwent aortic valve replacement with a 23-mm Freedom Solo stentless bovine pericardial tissue valve (SORIN Biomedica Cardio S.r.l.; Saluggia, Italy) for severe aortic stenosis; concomitantly, he underwent coronary artery bypass grafting (saphenous vein graft to obtuse marginal artery). The right cor- onary artery (RCA) had less than 50% luminal stenosis, so it was not bypassed. Post- operative recovery and echocardiographic results were normal. However, 4 months later, the patient presented with acute chest pain, dyspnea, and ischemic electrocardio- graphic changes. Angiography showed that rapid progression of coronary artery dis- ease had resulted in significant RCA stenosis, with some luminal irregularity of the left coronary artery ostium. Despite successful balloon angioplasty and stenting of the RCA, he had ischemic symptoms and began to develop acute pulmonary edema.

Detailed transesophageal echocardiography (Fig. 1) and aortography (Fig. 2) revealed severe regurgitation of the aortic prosthesis. The unstable hemodynamic situation war- ranted emergency reoperation.

After the institution of cardiopulmonary bypass by femorofemoral cannulation, the dense pericardial adhesions were released. Aortotomy revealed an intact aor- tic bioprosthesis, without evidence of endocarditis, leaflet perforation, pannus for- mation, paravalvular dehiscence, abscess formation, or calcification. However, the noncoronary leaflet of the prosthesis appeared to be elongated and it prolapsed into the left ventricle, thereby preventing proper leaflet coaptation. There was severe in- flammation, tissue fragility, and fibrosis on the inner surface of the aorta along the sewing ring suture line, secondary to the healing process of the implanted stentless valve. The aortic prosthesis was well attached to the aortic wall and had not loos- ened, nor did the aortic wall changes alter the geometric relationships between the posts of the prolapsing cusp. The valve leaflets were excised along with the adjoin- ing aortic sewing ring and sent for bacteriologic and histologic analysis to rule out

Case Reports

Sanjay Cherian, MD, PhD Hajo Müller, MD

Afksendiyos Kalangos, MD, Mustafa Cikirikcioglu, MD, PhD

PhD

Key words: Aortic valve insufficiency/surgery; bio- prosthesis/failure; complica- tion; disease progression;

heart valve prosthesis/

failure; reoperation

From: Divisions of Car- diovascular Surgery (Drs.

Cherian, Cikirikcioglu, and Kalangos) and Cardiol- ogy (Dr. Müller), University Hospitals and Faculty of Medicine of Geneva, 1211 Geneva, Switzerland

Address for reprints:

Mustafa Cikirikcioglu, MD, PhD, Division of Cardiovas- cular Surgery, University Hospitals of Geneva, Rue Gabrielle Perret-Gentil 4, 1211 Geneva, Switzerland

E-mail:

mustafa.cikirikcioglu@ hcuge.ch

(3)

secondary endocarditis, although there were no macro- scopic signs of endocarditis. The intense intra-aortic fi- brosis had reduced the luminal diameter and permitted the implantation of only a 19-mm Carpentier-Edwards Perimount Magna bovine aortic prosthesis (Edwards Lifesciences LLC; Irvine, Calif ) in a supra-annular po- sition using interrupted mattress sutures. Weaning the patient from cardiopulmonary bypass was difficult, de- spite inotropic and intra-aortic balloon pump support.

Transesophageal echocardiography showed anterolat- eral akinesia of the left ventricle, for which reason we bypassed the left anterior descending coronary artery (LAD) with a vein graft. Subsequent recovery was un- eventful. Postoperative angiography confirmed occlu- sion of the left main coronary ostium, with patency of the RCA and of both vein grafts (to the left LAD and obtuse marginal branch). Histologic analysis of the ex- cised leaflets revealed numerous macrophages and giant cells with focal fibrin deposition. At the 2-year follow- up examination, echocardiography showed good func- tioning of the new aortic prosthesis, with well-preserved left ventricular function.

Discussion

Stentless bioprostheses provide good hemodynamic per- formance1 and favorable rates of morbidity and mor- tality. Although long-term structural degeneration is a major drawback,3 early complications of the bioprosthe- sis are very uncommon and usually relate to acute leaf- let mineral deposition4 or failures in surgical technique.

This case report describes a rare, but potentially fatal, early failure of the Freedom Solo stentless bioprosthetic valve in the aortic position.

The Freedom Solo is a new-generation stentless bo- vine pericardial bioprosthesis designed for supra-an- nular implantation using a single suture line.5 Proper implantation of this prosthesis permits good alignment of the valve orifice with the patient’s annulus, result- ing in low postoperative gradients and decreased cross- clamping time.6

The unique design of the Solo aortic prosthesis re- quires accurate determination of size, proper position- ing, and precise symmetry during implantation in order to attain optimal valve function.5 The operative tech- nique used in our patient was in absolute accord with the manufacturer’s instructions and was informed by our study of previous surgical experience with this pros- thesis,5-7 in order to ensure no technical errors during implantation. Although this stentless valve is pliable and therefore adaptable to asymmetric sinuses, subcor- onary implantation of the valve might distort the aortic root, reduce the normal expansion of the sinuses, and progressively lead to increased mechanical stress on the valve leaflets.4 Furthermore, there might be a “washout”

effect on the right and left coronary leaflets—where-

Fig. 1 Long-axis transesophageal echocardiographic images at 4 months after prosthesis implantation, before emergency operation. A) Two-dimensional image shows that the aortic valve leaflets are not prolapsing into the left ventricular cavity during diastole. B) Color-flow Doppler image shows severe aortic regurgitation.

A

B

(4)

by blood flow through the left and right coronary ostia might decrease tension on those leaflets during diasto- le. Because the noncoronary leaflet lacks that tension- decreasing washout mechanism, it could be subject to more stress during each diastole.

It is known that glutaraldehyde pretreatment devi- talizes cells, thereby leading to leaflet stiffening, calci- fication, and subsequent valve failure. In contrast, the Solo valve is detoxified with homocysteic acid, which has been shown to prevent these sequelae.5 Furthermore, there was no evidence of bacterial endocarditis, leaf- let perforation, pannus, paravalvular dehiscence, ather- oma, abscess formation, or calcification in our patient.

The cause of acute prosthetic failure might then be at- tributed to an unknown primary tissue failure or to a mismatch of pericardial leaflets of varying tensile prop- erties, which could explain the prolapse of only one leaf- let. Histologic analysis of the excised leaflets revealed numerous macrophages and giant cells with focal fi- brin deposition. However, because these findings were not localized to the prolapsed leaflet, it is difficult to correlate them with the mechanism of prolapse. In ad- dition to the above-mentioned considerations of cause, this acute prosthesis failure could not be attributed to any patient-related factor.

Despite initial echocardiography that revealed good prosthetic function after proper surgical implantation,

severe aortic regurgitation developed in our patient within 4 months of implantation. The right coronary disease that appeared to be of little significance at initial intervention required percutaneous coronary interven- tion during his 2nd hospitalization. This rapid progres- sion of coronary atherosclerosis within a very short time is a well-recognized phenomenon.8

Reoperation within 1 year of primary prosthesis im- plantation is an independent predictor of periopera- tive death, owing to the intense inflammatory reaction and fibrosis that occasionally requires additional proce- dures—such as patch repair of the aortic root, total aor- tic root replacement, or the use of interposition grafts.9 However, our patient benefited from explantation of the stentless bioprosthesis and received a new, stented bioprosthesis. Dense inflammation and fibrosis of the inner aorta along the previous suture line led to loss of inner aortic diameter, permitting only a 19-mm stented bioprosthesis. Similarly, the luminal irregularity of the left coronary ostium was a result of the intense inflam- matory and fibrotic reaction of the adjacent aorta; this, along with the implantation of a stented bioprosthesis on a fibrosed aortic annulus, led to left coronary osti- al occlusion, which was successfully managed by emer- gency bypass grafting.

To the best of our knowledge, this case recounts the earliest failure of a stentless aortic Freedom Solo prosthe-

Fig. 2 Aortograms during A) systole and B) diastole show severe aortic regurgitation after stentless aortic bioprosthesis implantation.

Arrows outline the left ventricular cavity distended by the regurgitant jet during diastole.

A B

(5)

sis. It underscores the fact that stentless aortic biopros- thetic valves can present not only with late degeneration, but with early complications. In consideration of the acute nature of failure in this new-generation biopros- thesis and its unclear cause, we believe that this report, albeit of a single case, warrants some attention.

References

1. Borger MA, Carson SM, Ivanov J, Rao V, Scully HE, Feindel CM, David TE. Stentless aortic valves are hemodynamically superior to stented valves during mid-term follow-up: a large retrospective study. Ann Thorac Surg 2005;80(6):2180-5.

2. ElBardissi AW, DiBardino DJ, Chen FY, Yamashita MH, Cohn LH. Is early antithrombotic therapy necessary in pa- tients with bioprosthetic aortic valves in normal sinus rhythm?

J Thorac Cardiovasc Surg 2010;139(5):1137-45.

3. Pavoni D, Badano LP, Ius F, Mazzaro E, Frassani R, Gelso- mino S, Livi U. Limited long-term durability of the Cryolife O’Brien stentless porcine xenograft valve. Circulation 2007;

116(11 Suppl):I307-13.

4. Caprili L, Fahim AN, Zussa C, Cristell DM. Very early mal- function of a large stentless aortic valve. Eur J Cardiothorac Surg 2009;36(2):417-8.

5. Aymard T, Eckstein F, Englberger L, Stalder M, Kadner A, Carrel T. The Sorin Freedom SOLO stentless aortic valve:

technique of implantation and operative results in 109 pa- tients. J Thorac Cardiovasc Surg 2010;139(3):775-7.

6. Beholz S, Dushe S, Konertz W. Continuous suture technique for freedom stentless valve: reduced crossclamp time. Asian Cardiovasc Thorac Ann 2006;14(2):128-33.

7. Pektok E, Arpaz M, Sierra J, Karaca M, Muller H, Biceroglu S, et al. Durability of supraannular, stentless pericardial aortic valve prostheses and their impact on left ventricular remodel- ling in the mid-term [abstract]. Br J Surg 2010;97(S4):S46-7.

8. Cherian S, Roffi M, Kalangos A, Cikirikcioglu M. Rapid pro- gression of left main coronary disease causing acute coronary syndrome. Ann Thorac Surg 2011;91(5):1625.

9. Borger MA, Prasongsukarn K, Armstrong S, Feindel CM, David TE. Stentless aortic valve reoperations: a surgical chal- lenge. Ann Thorac Surg 2007;84(3):737-44.

Références

Documents relatifs

Only nostocarboline showed inhibitory activity to BChE (Becher et al. 2005 ), while the other bioactive compounds were negative.. The high inhibitory activity of nostocarbo-

Dans un contexte de forte contraction de l’emploi salarié et d’une hausse du nombre d’inscrits à Pôle emploi (21) de 18 % entre fin 2008 et fin 2009, la dépense en faveur

Dans le district sanitaire de Kati, malgré la mise en œuvre de la vaccination de routine contre la poliomyélite et de la tenue de multiples campagnes d'information et de

Quand il sera cinq heures de l’après-midi combien de fois la grande aiguille sera-t-elle passée au- dessus de la petite aiguille.. Elle passera

Même si l’aprépitant est officiellement indiqué seulement pour les cycles de chimiothérapie incluant le cisplatine, considérant la gravité des nausées et des

[r]

[r]

TABLEAU 1A : Nouveaux cas de tuberculose active et cas de retraitement déclarés et taux d’incidence pour 100 000 habitants (tous les cas) pour le Canada, les provinces et