Polarized Helium to Image the Lung

N/A
N/A
Protected

Academic year: 2021

Partager "Polarized Helium to Image the Lung"

Copied!
11
0
0

Texte intégral

(1)

HAL Id: hal-00004040

https://hal.archives-ouvertes.fr/hal-00004040

Preprint submitted on 24 Jan 2005

HAL is a multi-disciplinary open access archive for the deposit and dissemination of sci- entific research documents, whether they are pub- lished or not. The documents may come from

L’archive ouverte pluridisciplinaire HAL, est destinée au dépôt et à la diffusion de documents scientifiques de niveau recherche, publiés ou non, émanant des établissements d’enseignement et de

Polarized Helium to Image the Lung

Michèle Leduc, Pierre-Jean Nacher

To cite this version:

Michèle Leduc, Pierre-Jean Nacher. Polarized Helium to Image the Lung. 2005. �hal-00004040�

(2)

ccsd-00004040, version 1 - 24 Jan 2005

Polarized Helium to Image the Lung

Mich`ele Leduc and Pierre Jean Nacher [email protected] - [email protected]

Laboratoire Kastler Brossel, ENS, 24 rue Lhomond, F75005 Paris1

AbstractThe main findings of the european PHIL project (Polarised Helium to Image the Lung) are reported. State of the art optical pumping techniques for polarising 3Hegas are described. MRI methodological improvements allow dynamical ventilation images with a good resolution, ultimately limited by gas diffusion. Diffusion imaging appears as a robust method of lung diagnosis. A discussion of the potential advantage of low field MRI is presented. Selected PHIL results for emphysema are given, with the perspectives that this joint work opens up for the future of respiratory medicine.

Introduction

In 1994 first images of Magnetic Resonance Imaging (MRI) were published showing pulmonary cavities of a dead mouse, inflated with optically polarised

129Xe gas [1]. Soon after, research teams in the USA and in Germany, associat- ing atomic and MRI physicists with radiologists, demonstrated the possibility to image human lung airways using inhaled3He [2, 3].

These gases are chemically inert, non radioactive noble gases, both of nuclear spin 1/2. They are spin polarised prior to inhalation by optical pumping meth- ods, which means that a large non equilibrium magnetization is obtained by orientation of the magnetic moments of the gas nuclei. Then the gas is inhaled in the lungs and imaged by MRI in conventional scanners tuned to the NMR frequency of the gas atoms. This method allows visualising the intrapulmonary airspaces filled with the gas, instead of the tissues examined by conventional proton MRI. The large polarisation of the gas (up to 80%) compensates for the lower density of nuclear spins in the gas.

Images of the gas in the lungs can thus be recorded during a breathhold with a high resolution. Furthermore, the MRI acquisition sequences can be fast enough to provide dynamical ventilation data. Mapping physical parameters such as local relaxation times or apparent diffusion coefficients (ADC) indirectly provide information on the lung function or may reveal subtle changes in the microstructure of the airways. It is generally recognized that this new imaging method has unprecedented potential for morphological and functional analysis of the lung. It is expected to be a novel addition to the arsenal of pulmonary diagnostic tests and could enable the development of MRI in the chest, tradi- tionally hampered by the low number density of protons and the magnetic field inhomogeneities that exist at the air-tissues boundaries in the lung.

1Laboratoire Kastler Brossel is a unit´e de recherche de l’Ecole Normale Sup´erieure et de l’Universit´e Pierre et Marie Curie, associ´ee au CNRS (UMR 8552).

(3)

PHIL: a European consortium to study emphysema

It was decided in 2000 to put forward a European consortium, coordinated by one of the authors (M.L.), to link the efforts of nine teams in five different countries. Participants were selected for their pre-existing know-how and com- plementary competences in atomic physics (Paris, Mainz and Krakow), MRI methodology and instrumentation (Orsay and Lyon), radiology and respiratory medicine (Mainz, Sheffield and Copenhagen) and animal model studies (Madrid and Lyon).

The PHIL consortium (“Polarised Helium to Image the Lung”, [4]) joined efforts to demonstrate the potential and the validity of the new non invasive method as a diagnostic and prognostic tool for given lung pathologies: emphy- sema and selected Chronical Obstructive Pulmonary Diseases (COPD), such as bronchitis and bronchiolitis. This choice was motivated by the frequent oc- currence of these diseases and the very high cost of their treatment for society:

10% of the population and 25% of the smokers suffer from COPD, which is the fourth cause of mortality in Europe. The core of the project was to perform a clinical trial on a large group of patients with the 3He MRI method and with conventional techniques: pulmonary function tests, High Resolution Computed Tomography (HRCT), Krypton scintigraphy. An important objective of the PHIL project was to provide new tools for the study of COPD, aiming at dif- ferentiation of various types of diseases, as well as their detection at an early stage, with expectation that in the long range the findings of the project could lead to monitoring therapeutic treatment.

Polarising 3He by metastability exchange optical pumping

The only candidates as inhaled gaseous tracers are3He and 129Xe, which are spin 1/2 noble gas atoms (isotopes with higher nuclear spins are disregarded, because they are too sensitive to nuclear relaxation processes). 3He has been chosen for several reasons: It provides much larger NMR signals, as its magnetic moment is nearly three times larger than that of129Xe. The3He atoms can be polarised in larger quantities and to higher polarisation rates. Helium does not cross the alveolar air/blood barrier and remains confined within the air spaces.

It is absolutely harmless and induces no side effects, whereas xenon is known as a potent anaesthetic at high concentrations.

3He can be polarised by two established methods which rely on optical pump- ing techniques. Optical pumping is a very efficient method to control the atomic spin state through interaction with a resonant light beam carrying angular mo- mentum. In the presence of a magnetic guiding fieldB and of weak relaxation processes, the net result of the repeated light absorption and re-emission cycles is a change in the relative populations of the involved atomic sublevels, i.e.

creation of spin orientation. For polarising3He, two routes are possible: spin exchange optical pumping (SEOP) [5], with indirect transfer of angular mo- mentum from a polarised laser beam to the3He nuclei via alkali metal atoms

(4)

(usually rubidium), and metastability exchange optical pumping (MEOP) [6], with direct transfer of angular momentum from resonant laser light to 3He atoms. Each technique has its own advantages and limitations. For instance, SEOP directly operates at high pressure but is a slow process (several hours are required to polarise helium), while MEOP is a much faster process (it only takes seconds to polarise 100 cm3 of gas at 1 mbar), but only operates at low pressure (∼mbar) and requires further compression for its application to lung MRI. SEOP has been used for several imaging experiments [7], but MEOP has been preferred for human lung imaging in the PHIL project because of its faster production rates and higher nuclear polarisations.

The principle of a3He optical pumping experiment using MEOP is illustrated in figure 1. A fraction of the helium atoms are excited by a plasma discharge to

Metastability exchange collisions 1083 nm

OP transition 21S0 21P0

11S0 Resonant

laser beam (1083 nm)

σσ++

Circular polariser

B Helium cell

RF

Figure 1: Left: schematic view of an optical pumping setup. In a low-pressure helium-3 gas cell, a weak RF discharge promotes a small fraction (106) of the atoms into the excited metastable state 23S, where resonant absorption of the circularly polarised 1083 nm light can occur. Repeated cycles of absorp- tion/emission of photons transfer a substantial polarisation in the 23S state.

Nuclear polarisation is transferred to the ground state atoms by metastability exchange collisions. Right: atomic levels of helium-3 and physical processes involved in the OP cycle.

a metastable energy state, where they can absorb resonant light to be optically pumped. For the3He isotope in this excited metastable state, an efficient cou- pling between the nucleus and the electrons (the hyperfine interaction) results in a strong entanglement of electronic and nuclear spins. Therefore optical ori- entation of the electronic angular momentum simultaneously induces nuclear orientation as well. This nuclear orientation is rapidly transferred to the atoms having remained in the ground state through metastability exchange collisions.

This process corresponds to a very short interaction between an excited atom and a ground state atom, which results in a fast exchange of the electronic excitations of the colliding atoms with no change in the nuclear orientations.

3

(5)

In actual gas polarisation devices, a uniform magnetic field (of order 1 mT) is applied over the whole setup to prevent magnetic relaxation. Powerful fibre lasers at 1083 nm are used for efficient optical pumping [8]. MEOP-based gas polarisation, which requires compression of the polarised gas, was developed for PHIL following two routes with complementary advantages:

- Centralised production of very high grade hyperpolarised gas and transport to the MRI centre. The Mainz group pioneered the development of dedicated systems for massive production of hyperpolarised 3He gas [9]. Sophisticated piston compressors lead to unrivalled performance in terms of output pressure, gas polarisation and production rate. Such facilities deliver remarkably large flux (tens of bar litres per day) of high grade polarised 3He: over 80% at the production unit, still reaching 60% after shipping by air (from Mainz to Copenhagen or to Sheffield) [10]. The transportation problems were overcome using special containers preventing magnetic relaxation and appropriate iron- free glass for the storage cells containing the gas.

- On-site production of gas with compact polarisers implemented next to the MRI scanner. The delivered gas may be of lower grade (3He is diluted in a neutral gas like nitrogen, or produced with lower polarisation, of order 40%), but problems with storage and transportation are avoided. The key element of the system developed by the Paris group is a compact polarisation-preserving peristaltic compressor that allows production of polarised gas next to or inside the MRI scanner [11]. Two different prototypes of such compact and robust polarisers were built, and are ready for possible commercial production.

Pulmonary ventilation images: resolution and dynamics

Figure 2 displays chest MRI images obtained at 1.5 Tesla with conventional proton MRI (left) and polarised helium MRI (right). On the proton scan, most

Figure 2: Chest MR images (1 cm thick slices) of the same normal subject.

Left: fast spin-echo proton image. Right: FLASH3He image of the gas in the lungs. The NMR frequency is changed from 64 MHz for protons to 49 MHz for

3He. Courtesy of J. Wild, Academic Radiology dept., U. of Sheffield.

(6)

tissues are visible, but the lung parenchyma does not appear: this is due partly to the low density of the lung tissues, partly to the shortening of the transverse relaxation time T2 by the non-uniform magnetisation of the inhomogeneous parenchyma in the high magnetic field of the scanner. On the helium scan, the gas-filled pulmonary air spaces are the only visible part of the chest. Shortening ofT2by susceptibility effects is less severe thanks to the rapid gas diffusion at the alveolar scale, a well known motional averaging and line narrowing effect. A set of 15-20 such slices at adjacent positions in the chest allows re-constructing a 3D map of gas distribution in the lungs.

Another attractive possibility offered by the3He-MRI technique is access to dynamical parameters of the lung ventilation. A typical temporal resolution of order 0.1 s is required to monitor the respiratory cycle. For ultra-fast imag- ing, a radial acquisition strategy combined with a sliding window method was developed. This method, referred as SPIRO (Sliding Pulmonary Imaging for Respiratory Overview), allows a regional quantification of dynamical ventila- tion parameters using a single breath of polarised gas. Figure 3 shows two series of helium lung images (here 2D projections) taken during a respiratory cycle. They show the gas flowing first down the trachea, then into the bronchi

Figure 3: Dynamic images collected using a sliding window radial sequence [12];

the time interval between displayed images is 0.5 s. Top row: healthy normal subject, from the first part of an inhalation of 300 ml of3He polarized to 40%.

Bottom row: COPD patient showing regions of ventilation obstruction in both lungs, particularly in the upper lobes, and a delayed emptying/depolarization of gas in the lower left lobe which could be indicative of air trapping. Courtesy of J. Wild, Academic Radiology dept., U. of Sheffield.

and finally in the peripheral lung. Clear differences are visible between a nor- mal subject (top row) and a patient suffering from COPD (bottom row), with wedge-shaped defects indicating that some pulmonary zones are poorly venti- lated. Key functional parameters can be extracted from the images, such as

5

(7)

gas arrival time, filling time constant, and gas volume. Regional maps of such parameters can be used to assess the degree of the disease severity.

The resolution of these helium images, even if superior to that of competing techniques, is however intrinsically limited and does not allow visualising the airways at the scale of the alveoli (of order 0.1 mm). Proton MRI can reach micrometric resolution for small objects if sufficient signal to noise ratio is achieved. The same resolution is not possible for helium MRI, due to the rapid diffusion of the gas inside the airways. MRI relies on frequency encoding of the position through the use of magnetic field gradients, sequentially applied along various directions in space. IfG is a field gradient applied along thex direction, the NMR frequency change across a distanceδxisδf=γGδx(γ=32 MHz/T is the gyromagnetic ratio of3He). To resolve details of sizeδx,Gmust be applied for a timeτ such that τ δf 1, hence δx 1/γGτ. When dealing with a gas, the observation timeτ is actually limited by the diffusion of the gas overδx, namelyτδx2/D(D80 mm2/s is the diffusion coefficient of3He in air). Thus the resolution is limited toδx(D/γG)1/3, of order 0.4 mm for a conventional scanner with typical gradients reaching 0.3 mT/cm.

Apparent Diffusion Coefficient

In spite of the limitations on image resolution with3He-MRI, measurements of apparent diffusion coefficients (ADC) can provide relevant information on the lung microstructure. When diffusing freely in air (in the absence of restrict- ing walls or barriers), helium atoms are displaced on average by an amount

, e.g. 0.9 mm after a timeτ=10 ms. In the lung, free diffusion occurs at such time scales only in larger airways, while collisions with the walls of smaller airways and of alveoli reduce displacements of helium atoms. A time- and scale- dependent ADC is introduced to characterise this restricted diffusion. It is not easy to relate this reduced value to a detailed description of the microstructure of the lung at the alveolar level due to its complexity. However measuring ADC values can be useful to assess pathological changes. For instance in emphy- sema, the restrictions to diffusion are reduced due to expansion of alveoli and modification of connectivity between them, which results in an increased ADC.

Several NMR techniques can be used to measure diffusion coefficients. They rely on applying a sensitising gradient to imprint a non-uniform magnetisation pattern during a given time, and on measuring the resulting magnetisation de- cay. The ratio of images recorded with and without such a diffusion-sensitising gradient is used to compute a diffusion map. Figure 4 displays ADC maps ob- tained in the lung with a standard technique (diffusion weighting by a bipolar gradient pair). One observes an increase by more than a factor two for the mean ADC value for a COPD patient compared to a normal subject. One also notes a significant difference between the ADC histograms recorded for an asymp- tomatic smoker and a normal subject; this indicates that ADC measurements have a potential for detection of mild COPD diseases.

(8)

Ventilation ADC map ADC Histogram (cm²/s)

Normal

Smoker

COPD

Figure 4: Left to right: gas distribution images, ADC maps and ADC his- tograms in 3 subjects (top to bottom). Courtesy of J. Wild, Academic Radiol- ogy dept., U. of Sheffield.

Low field versus high field for 3He MRI?

In conventional proton MRI, high magnetic fields are used to obtain sufficient equilibrium proton magnetisation and NMR signal. Noise arises from thermal (Johnson) noise in the patient’s tissues and in the NMR receiving coil. Usually, the former noise source dominates; for a given magnetisation, both the NMR signal and the noise are proportional to frequency (i.e. to the field). For3He- MRI with optically pumped gas, the magnetisation of the gas and hence the signal to noise ratio (SNR) do not depend on the field (this contrasts with the usual proton MRI for which the equilibrium magnetisation, and hence the SNR, increase linearly with the field). However this breaks down at very low field (below about 30 mT, depending on the coil size and technology), where the patient noise decreases below the thermal noise of the coil.

The PHIL Partner in Orsay demonstrated that there is no SNR penalty when using a low field MRI scanner at 0.1 T [13]. On the contrary, there is an important benefit of low-field operation: with reduced gradients arising from the magnetic susceptibility of tissues, transverse relaxation times are much longer at 0.1 T than at 1.5 T. Weaker field gradients can then be used for imaging or ADC measurements. This allows measuring diffusion coefficients at longer time scales, thus providing more information on the connectivity of lung air spaces.

A precise way of measuring ADC coefficients in this case is through the decay of spin echo trains.

7

(9)

However, shortening of transverse relaxation times can also be viewed as a useful contrast mechanism since it depends on the lung microstructure. It was for instance demonstrated to be more sensitive to the degree of lung inflation at 1.5 T than at 0.1 T [14].

A clear advantage of operating at low field is the reduced cost and increased flexibility of the imaging system: an open geometry can be designed, and stand- ing or sitting patients may be examined. Studies at ultra low field in a home made vertical scanner at 3 mT have started with the PHIL Partner in Paris [15]. This could open up possibilities of low-cost dedicated scanners for the screening or follow-up of lung diseases.

Emphysema studies with 3He MRI

The PHIL multicentre clinical trial, after upgrading the hardware and software of the involved MRI scanners according to the3He MRI standards, put forward a standardized protocol to study a significant number of patients and healthy volunteers for reference. Pre-existing sequences initially used in Mainz were adapted and optimized at the different sites: ventilation distribution (breath- hold), apparent diffusion coefficient (breath hold) and dynamic ventilation (sin- gle respiratory cycle). The protocol also included pulmonary function tests to assess severity scores. Data were recorded in a data base and finally submitted to a statistical analysis, which provided a wealth of information, still not fully exploited [16].

The subjects were divided in three groups: 79 patients with proven COPD, among whom 17 suffered from alpha-1-ATD (anti-trypsin deficiency) and 37 healthy volunteers. A score number referring to the degree of abnormality was attributed to each subject. Comparison between HRCT, MRI findings and abnormality scores showed that MRI correlates better with pulmonary function tests than HRCT. The vast majority of MRI findings consisted of wedged shaped defects, even subjects with normal HRCT showed ventilation defects. The comparison of ADC with emphysema index and mean lung density was also exploited. ADC measurements clearly separated the healthy subjects from those with emphysema of any type. 3He MRI proved also to be able to distinguish between different diseases by using a combination of ventilation distribution and ADC mapping. After assessing all the results, one general finding is that HRCT is usually more sensitive but MRI is more specific. A positive feature of the3He MRI method is that a single exam provides several unique complementary informations at the same time (morphology, ventilation defects, average of air spaces, even in-vivo oxygen concentration and uptake that can be derived from the local relaxation times).

The animal models studies of PHIL provided very complementary results [17]. Different models of lung diseases were induced in rats with chemicals such as elastase or cadmium chloride. To some extent these models mimic human emphysema. The rat groups treated with cadmium showed clear ventilation defects, corresponding to non-ventilated areas of the partially collapsed lung.

(10)

ADC and post-mortem morphometric measurements were compared. The most significant finding of this work is the excellent correlation between the ADC data and the morphometric ones, which allow direct quantification of alveolar size.

Complementary to experiments in clinical trials, in which the available popula- tion generally have suffered chronic diseases, emphysema in animals was mildly developed to evaluate the technique for early diagnosis. In that perspective, low field scanners could be highly useful, as the system successfully built in Krakow with a permanent magnet of 0.08 T based on a permanent magnetic material of new generation. Such low cost scanners are adapted to imaging small animal lungs and may be especially valuable for pharmaceutical research.

Conclusion

In this multidisciplinary work, atomic physicists played a prominent role at the early stage of the technique. The 3He polarisation methods were first devel- oped in the 80’s for the needs of polarised quantum fluids physics, for polarised targets in nuclear physics and more recently for neutron spin filters at reac- tors. Finally, in view of application to MRI, the metastability exchange optical pumping method was pushed to its limits and very appropriate polarisers were constructed by the PHIL consortium. They bring a satisfactory solution for clinical implementation of3He MRI, either by remote production of high grade gas or in-situ table-top polarisation. There is still room for improvement of fast dedicated MRI sequences and signal to noise enhancement. Great expectation is also related to dedicated low field scanners, with possible extension of the

3He MRI lung imaging method to developing countries.

The PHIL consortium has shown that3He MRI can be routinely performed on a significant number of patients and shows interesting new insights into common lung diseases, COPD and emphysema. It proves superior in attaining 3D lung function in the lungs, and is more closely correlated with routine lung function tests than HRCT. The method has already been tested in other lung diseases, and more studies are planned or have started. These include lung transplant assessment, cystic fibrosis (the most common genetic lung disease), asthma and lung cancer. The technology has proved of huge potential interest in these areas, because it is non-invasive and does not require ionising radiation, an extremely important feature, especially in longitudinal follow-up of children.

It could also be very beneficial for the development of novel therapies, reducing the size of trials and the number of sacrificed animals.

References

[1] M.S. Albert, G.D. Cates, B. Driehuyset al., Nature370, 199 (1994): Bi- ological MRI using laser-polarized129Xe

[2] J.R. MacFall, H.C. Charles, R.D. Blacket al., Radiology200, 553 (1996):

Human lung air spaces: potential for MRI with hyperpolarized3He

9

(11)

[3] H.U. Kauczor, D. Hofmann, K.F. Kreitner et al., Radiology 201, 564 (1996): Normal and abnormal pulmonary ventilation: Visualization at hyperpolarized3He MRI

[4] cf. http:// www.phil.ens.fr. The PHIL project has been supported by the EC contract QLG1-2000-01559 running from 12/2000 to 06/2004.

[5] T.G. Walker and W. Happer, Rev. Mod. Phys. 69, 629 (1997): Spin- exchange optical pumping of noble-gas nuclei

[6] P.-J. Nacher and M. Leduc, J. Physique46, 2057 (1985): Optical pumping in 3He with a laser

[7] T. Chupp and S. Swanson, Adv. At. Mol. Opt. Phys.45,51 (2001): Med- ical imaging with laser polarised noble gases

[8] G. Tastevin, S. Grot, E. Courtadeet al.,Appl. Phys. B78, 145 (2004): A broadband Yb-doped tunable fiber laser for He optical pumping at 1083nm [9] J. Becker, J. Bermuth, M. Ebertet al., Nucl. Instr. Meth. Phys. Res.402,

327 (1998): Interdisciplinary experiments with polarized3He

[10] E.J.R Van Beek, J. Schmiedeskamp, J.M. Wildet al., Eur. Radiol.13, 2583 (2003): Hyperpolarized3He MRI of the lungs: improving accessibility to gas using a central production facility and transport by air

[11] P.-J. Nacher, G. Tastevin, X. Maˆitreet al.,Eur. Radiol.9B18 (1999): A peristaltic compressor for hyperpolarised helium

[12] J. Wild, M.N.J. Paley, L. Kasuboski et al., Magn. Reson. Med. 49, 991 (2003): Dynamic radial projection MRI of inhaled hyperpolarized3He [13] E. Durand, G. Guillot, L. Darrasseet al., Magn. Reson. Med.47, (2002)

75: CPMG measurements and ultrafast imaging in human lungs with hy- perpolarized3He at low field (0.1 T)

[14] L. de Rochefort, A. Vignaud, X. Maˆitreet al.,Proc. 12th ISMRM meeting ISSN 1545-4436, 2724 (2004): Influence of lung filling on T2* values in human at 1.5 T with hyperpolarised3He

[15] C.P. Bidinosti, J. Choukeife, G. Tastevinet al., Magnetic Resonance Ma- terials in Physics, Biology, and Medicine (MAGMA)16, 255 (2004): MRI of the lung using hyperpolarized3He at very low magnetic field (3 mT) [16] K.K. Gast, A. Dahmen, C.P. Heusselet al., Proc. 12th ISMRM meeting

ISSN 1545-4436, 1673 (2004): 3He-MRI in a european multicenter trial in COPD and emphysema PHIL

[17] G. Peces-Barba, J. Ruiz-Cabello, Y. Cremillieuxet al., Eur. Resp. J.22, 14-19 (2003): 3He diffusion coeficient: correlation to morphometry in a model of mild emphysema

Références

Documents relatifs

The measured parameters of our subjects were compared with the predicted values for Caucasians of the American Thoracic Socc ciety [5].The equations were considered to be

The presence of a functional lung in fossil coelacanths is presumed based on the presence of an opened pneu- matic duct (in extant and fossil coelacanths); the presence of a

AR: Acute rejection; BIG: Bacterial infection group; BOS: Bronchiolitis obliterans syndrome; CG: Control group; CLAD: Chronic lung allograft dysfunction; FEV-1: Forced expiratory

This study aims to determine the mechanisms for supporting and developing of competitiveness of SMEs facing the challenges of globalization. The research is conducted through the

With Maxwell-Stefan’s laws, we solve the system given by the relation on the fluxes (6)–(7), the second being a consequence of the first, and (1) and (4) for two of the three

The relationship between infant lung function and the risk of wheeze in the preschool years... For

2 Michel Coste and Nestor Djintelbe The purpose of the introduction of a meaningful boundary for the group of rigid motions is to obtain information on the kinematics of the robots

L’archive ouverte pluridisciplinaire HAL, est destinée au dépôt et à la diffusion de documents scientifiques de niveau recherche, publiés ou non, émanant des