• Aucun résultat trouvé

Worldwide Interlaboratory Comparison on the Determination of Trace Elements in a Fish Sample | IAEA

N/A
N/A
Protected

Academic year: 2022

Partager "Worldwide Interlaboratory Comparison on the Determination of Trace Elements in a Fish Sample | IAEA"

Copied!
90
0
0

Texte intégral

(1)

IAEA/AQ/55

IAEA Analytical Quality in Nuclear Applications Series No. 55

Worldwide Interlaboratory

Comparison on the Determination of Trace Elements in a Fish Sample

INTERNATIONAL ATOMIC ENERGY AGENCY VIENNA

ISSN 2074–7659

W orldwide Interlaboratory Comparison on the Determination of Trace Elements in a Fish Sample

IAEA-MESL-ILC-TE-BIOTA-2017

Template has: 20 mm spine

please reset it to the corrected spine/

(2)

WORLDWIDE INTERLABORATORY

COMPARISON ON THE DETERMINATION

OF TRACE ELEMENTS IN A FISH SAMPLE

(3)

AFGHANISTAN ALBANIA ALGERIA ANGOLA

ANTIGUA AND BARBUDA ARGENTINA

ARMENIA AUSTRALIA AUSTRIA AZERBAIJAN BAHAMAS BAHRAIN BANGLADESH BARBADOS BELARUS BELGIUM BELIZE BENIN

BOLIVIA, PLURINATIONAL STATE OF

BOSNIA AND HERZEGOVINA BOTSWANA

BRAZIL

BRUNEI DARUSSALAM BULGARIA

BURKINA FASO BURUNDI CAMBODIA CAMEROON CANADA

CENTRAL AFRICAN REPUBLIC CHADCHILE CHINA COLOMBIA CONGO COSTA RICA CÔTE D’IVOIRE CROATIA CUBACYPRUS

CZECH REPUBLIC DEMOCRATIC REPUBLIC

OF THE CONGO DENMARK DJIBOUTI DOMINICA

DOMINICAN REPUBLIC ECUADOR

EGYPT EL SALVADOR ERITREA ESTONIA ESWATINI ETHIOPIA FIJIFINLAND FRANCE GABON GEORGIA

GERMANY GHANA GREECE GRENADA GUATEMALA GUYANA HAITI HOLY SEE HONDURAS HUNGARY ICELAND INDIA INDONESIA

IRAN, ISLAMIC REPUBLIC OF IRAQIRELAND

ISRAEL ITALY JAMAICA JAPAN JORDAN KAZAKHSTAN KENYA

KOREA, REPUBLIC OF KUWAIT

KYRGYZSTAN

LAO PEOPLE’S DEMOCRATIC REPUBLIC

LATVIA LEBANON LESOTHO LIBERIA LIBYA

LIECHTENSTEIN LITHUANIA LUXEMBOURG MADAGASCAR MALAWI MALAYSIA MALIMALTA

MARSHALL ISLANDS MAURITANIA

MAURITIUS MEXICO MONACO MONGOLIA MONTENEGRO MOROCCO MOZAMBIQUE MYANMAR NAMIBIA NEPAL

NETHERLANDS NEW ZEALAND NICARAGUA NIGER NIGERIA NORWAY OMANPAKISTAN

PALAU PANAMA

PAPUA NEW GUINEA PARAGUAY

PERUPHILIPPINES POLAND PORTUGAL QATAR

REPUBLIC OF MOLDOVA ROMANIA

RUSSIAN FEDERATION RWANDA

SAINT VINCENT AND THE GRENADINES SAN MARINO SAUDI ARABIA SENEGAL SERBIA SEYCHELLES SIERRA LEONE SINGAPORE SLOVAKIA SLOVENIA SOUTH AFRICA SPAIN

SRI LANKA SUDAN SWEDEN SWITZERLAND

SYRIAN ARAB REPUBLIC TAJIKISTAN

THAILAND

THE FORMER YUGOSLAV REPUBLIC OF MACEDONIA TOGOTRINIDAD AND TOBAGO TUNISIA

TURKEY

TURKMENISTAN UGANDA UKRAINE

UNITED ARAB EMIRATES UNITED KINGDOM OF

GREAT BRITAIN AND NORTHERN IRELAND UNITED REPUBLIC

OF TANZANIA

UNITED STATES OF AMERICA URUGUAY

UZBEKISTAN VANUATU

VENEZUELA, BOLIVARIAN REPUBLIC OF

VIET NAM YEMEN ZAMBIA ZIMBABWE The following States are Members of the International Atomic Energy Agency:

The Agency’s Statute was approved on 23 October 1956 by the Conference on the Statute of the IAEA held at United Nations Headquarters, New York; it entered into force on 29 July 1957.

The Headquarters of the Agency are situated in Vienna. Its principal objective is “to accelerate and enlarge the contribution of atomic energy to peace, health and prosperity throughout the world’’.

(4)

IAEA/AQ/55 IAEA Analytical Quality in Nuclear Applications Series No. 55

WORLDWIDE INTERLABORATORY

COMPARISON ON THE DETERMINATION OF TRACE ELEMENTS IN A FISH SAMPLE

IAEA-MESL-ILC-TE-BIOTA-2017

INTERNATIONAL ATOMIC ENERGY AGENCY VIENNA, 2018

(5)

COPYRIGHT NOTICE

All IAEA scientific and technical publications are protected by the terms of the Universal Copyright Convention as adopted in 1952 (Berne) and as revised in 1972 (Paris). The copyright has since been extended by the World Intellectual Property Organization (Geneva) to include electronic and virtual intellectual property. Permission to use whole or parts of texts contained in IAEA publications in printed or electronic form must be obtained and is usually subject to royalty agreements. Proposals for non-commercial reproductions and translations are welcomed and considered on a case-by-case basis. Enquiries should be addressed to the IAEA Publishing Section at:

Marketing and Sales Unit, Publishing Section International Atomic Energy Agency

Vienna International Centre PO Box 100

1400 Vienna, Austria fax: +43 1 26007 22529 tel.: +43 1 2600 22417

email: sales.publications@iaea.org www.iaea.org/books

For further information on this publication, please contact:

IAEA Environment Laboratories, Monaco Marine Environmental Studies Laboratory

International Atomic Energy Agency 4a Quai Antoine 1er, 98000

Principality of Monaco

WORLDWIDE INTERLABORATORY COMPARISON ON THE DETERMINATION OF TRACE ELEMENTS IN A FISH SAMPLE IAEA, VIENNA, 2018

IAEA/AQ/55 ISSN 2074–7659

© IAEA, 2018 Printed by the IAEA in Austria

November 2018

(6)

FOREWORD

The primary goal of the IAEA Environment Laboratories is to help Member States to understand, monitor and protect the marine environment. In this connection, the substantial impacts of large coastal cities on marine ecosystems are of primary concern for the IAEA and its Environment Laboratories in particular.

The marine pollution assessments needed to understand these impacts depend on accurate knowledge of contaminant concentrations in various environmental compartments. With the need to base scientific conclusions on valid and internationally comparable measurement results, and to provide policy makers with accurate and reliable information on the state of the environment, ensuring the quality of measurement results produced by individual laboratories is indispensible.

The IAEA’s Marine Environmental Studies Laboratory in Monaco, part of IAEA Environment Laboratories, serves as the analytical support centre for IAEA Member State laboratories and is a pillar of the IAEA’s quality assurance programme for determination of non-nuclear pollutants — trace elements and organic contaminants — in the marine environment.

Since the 1960s, the IAEA has been providing assistance to Member States in the field of data quality and quality assurance. To support Member States in their marine monitoring activities, as well as in the domain of food safety, the Marine Environmental Studies Laboratory has produced certified reference materials characterized for trace elements and methylmercury using samples of marine origin (biota and sediments), and has organized interlaboratory comparisons.

The IAEA Environment Laboratories interlaboratory comparisons and proficiency tests involve the comparison of participants’ results with an assigned value, usually a consensus value from the overall population of test results. The comparisons and tests are designed to allow participating laboratories to monitor and demonstrate their performance and analytical capabilities. At the same time, gaps and problem areas where further developments are needed can be also identified.

The present publication summarizes the results of the IAEA-MESL-ILC-TE-BIOTA-2017 interlaboratory comparison on the determination of trace elements and methylmercury in fish.

The IAEA is grateful to the Government of Monaco for the support provided to its Environment

Laboratories. The IAEA is grateful to the participants and laboratories that took part in this

intercomparison exercise. The IAEA officers responsible for this publication were S. Azemard

and E. Vasileva of the IAEA Environment Laboratories.

(7)

EDITORIAL NOTE

This publication has been prepared from the original material as submitted by the contributors and has not been edited by the editorial staff of the IAEA. The views expressed remain the responsibility of the contributors and do not necessarily reflect those of the IAEA or the governments of its Member States.

Neither the IAEA nor its Member States assume any responsibility for consequences which may arise from the use of this publication.

This publication does not address questions of responsibility, legal or otherwise, for acts or omissions on the part of any person.

The use of particular designations of countries or territories does not imply any judgement by the publisher, the IAEA, as to the legal status of such countries or territories, of their authorities and institutions or of the delimitation of their boundaries.

The mention of names of specific companies or products (whether or not indicated as registered) does not imply any intention to infringe proprietary rights, nor should it be construed as an endorsement or recommendation on the part of the IAEA.

The IAEA has no responsibility for the persistence or accuracy of URLs for external or third party Internet web sites referred to in this publication and does not guarantee that any content on such web sites is, or will remain, accurate or appropriate.

(8)

CONTENTS

1. INTRODUCTION ... 1

2. SCOPE OF THE INTERCOMPARISON ... 1

3. DESCRIPTION OF ILC TEST MATERIAL ... 1

4. EVALUATION OF ANALYTICAL PERFORMANCE ... 3

5. RESULTS AND DISCUSSION ... 4

5.1. Overview of the results ... 4

5.2. Laboratory performances ... 4

5.3. Analytical methods ... 12

5.4. Sample treatment, impact of water content, use of crm and recovery corrections ... 13

6. RECOMMENDATIONS ... 13

7. CONCLUSIONS ... 14

APPENDIX I ... 15

APPENDIX II ... 37

REFERENCES ... 67

LIST OF PARTICIPANTS ... 69

CONTRIBUTORS TO DRAFTING AND REVIEW ... 77

(9)
(10)

1 1 INTRODUCTION

The Marine Environmental Studies Laboratory (MESL) of the International Atomic Energy Agency’s Environment Laboratories (IAEA-NAEL) has the programmatic responsibility to provide assistance to Member States’ laboratories in maintaining and improving the reliability of analytical measurement results for trace elements and organic pollutants. This is accomplished through the provision of reference materials of marine origin, validated analytical procedures, training in the implementation of internal quality control, and through the evaluation of measurement performance of monitoring laboratories, via the organization of targeted worldwide and regional interlaboratory comparisons.

In order to assure reliability of analytical data for monitoring studies, one essential aspect of quality assurance and quality control is the periodic external assessments measurement performance of laboratories in charge of monitoring studies via Interlaboratory Comparisons (ILCs) or Proficiency Tests (PTs). The results from ILCs or PTs are of crucial interest for laboratories as they provide clear information of their measurement capabilities.

These exercises are designed to monitor the performance and analytical capabilities of the participating laboratoriesand to identify gaps and problem areas where further development is needed. Regular participation in the ILCs and PTs enhanced trust in results and methodologies and provides objective evidence for the accreditation purposes.

2 SCOPE OF THE INTERCOMPARISON

The present ILC was designed to evaluate the measurement performance of participating laboratories for the determination of trace elements in marine biota sample (fish).

Letters of invitation have been sent to 293 laboratories, previously participated or expressed the interest in the IAEA ILCs. Positive responses were received from 111 laboratories from 57 Member States. 61 laboratories from 38 IAEA Member States received the ILC test sample.

Each participating laboratory received one bottle of the test sample, accompanied by one information sheet and instruction for the use of the IAEA-NAEL on-line reporting system.

Participants were requested to determine as many elements as possible from the following list of elements:

Ag, As, Ca, Cd, Co, Cr, Cu, Fe, Hg, CH3Hg, K, Mg, Mn, Ni, Pb, Se, Sn, Sr, V and Zn,

using analytical procedures routinely applied in their laboratories.

In total 49 laboratories from 32 countries reported results back to the organizers. The data reported by laboratories, together with the technical and statistical evaluations of the results for investigated trace elements, are included in this report. All results were treated confidentially and each laboratory was identified with a unique confidential code number

3 DESCRIPTION OF ILC TEST MATERIAL

About 350 kg of mixed fish was collected in the eastern Irish Sea. After removal of the skin;

the fish filleted was freeze-dried and grounded to powder and sieved at 250µm. The portion above 250µm was reprocessed by micronisation. The obtained sample material with grain size

<100µm was further homogenized and bottled in plastic container.

Homogeneity test was performed at the MESL following the requirements ISO 35 guidelines

[1], using preliminary validated in MESL’s trace elements laboratories analytical methods.

(11)

2

All reported measurement results were evaluated by using Kernel density plots, as an appropriate method to represent the overall structure of a data set [2]. No bimodality structures linked to the sample preparation or analytical method were observed. Consequently, the assigned values and their measurement uncertainties have been calculated with all reported results.

Robust statistics was used for the determination of the assigned values. The robust mean and robust standard deviations were calculated as described in the ISO 13528, Annex C.21 [3].

The uncertainties associated with the assigned values were calculated according to the ISO standard 35 [1]. The combined uncertainty of the assigned value consists of uncertainty related to characterization (u

char

), between bottle heterogeneity (u

hom

) and long–term stability (u

stab

).

Above mentioned contributions were combined to estimate the expanded uncertainty using Eq.

(1).

𝑈 = 𝑘 × 𝑢 + 𝑢 + 𝑢 (1)

where:

k: coverage factor, k=2, representing level of confidence of about 95%

uhom

is the standard uncertainty, due to between unit inhomogeneity. u

hom

was calculated as the between unit standard deviation using ANOVA statistical approach [1].

u

stab

is the standard uncertainty, due to long term stability of the sample. As the raw sample was prepared more than 10 years ago

ustab

component was considered to have negligible contribution and was not further propagated during the estimation of the total combined uncertainty.

uchar

is the uncertainty of characterization, estimated according to the recommendations of the ISO 13528 [3] using Eq. (2).

𝑢 = 1.25 ×

(2)

Where: s* is the robust standard deviation and n the number of measurement results.

Assigned values and expanded uncertainties are presented in Table 1. For Co, Cr, Ni and Sn

expanded uncertainty was beyond 25%, therefore those values are given for information only

and will not be used for the evaluation of measurement performances of laboratories,

participating in this interlaboratory comparison.

(12)

3 TABLE 1. ASSIGNED VALUES AND UNCERTAINTY IN PT SAMPLE

Element Unit Assigned value Expanded uncertainty (k=2)

Ag mg kg-1 0.066 0.008

As mg kg-1 19.9 1.1

Ca mg kg-1 27.4 × 103 4.6 × 103

Cd mg kg-1 0.032 0.004

CH3Hg mg kg-1 as Hg 0.531 0.092

Co mg kg-1 0.121 0.020

Cr mg kg-1 3.45 0.86

Cu mg kg-1 2.40 0.34

Fe mg kg-1 138 19

Hg mg kg-1 0.593 0.032

K mg kg-1 14.0 × 103 2.4 × 103

Mg mg kg-1 2.11 × 103 0.37 × 103

Mn mg kg-1 15.4 1.5

Ni mg kg-1 4.09 1.18

Pb mg kg-1 0.606 0.064

Se mg kg-1 2.17 0.20

Sn mg kg-1 0.182 0.054

Sr mg kg-1 136 23

V mg kg-1 0.405 0.085

Zn mg kg-1 52.1 3.0

4 EVALUATION OF ANALYTICAL PERFORMANCE

The individual laboratory performance was expressed in terms of z-scores and Zeta-scores, in accordance with the ISO 17043 [4].

The determination of target standard deviation was based on the outcome from the previous ILCs, organized by the MESL with similar sample matrices for the same population of laboratories. The standard deviation for the proficiency assessment, σ

p

, was fixed to 12.5 % of the assigned values. The appropriateness of this level of tolerated variability of results was confirmed by calculation of the robust standard deviation of the participants’ results and the uncertainty of the assigned values for the respective measurants.

z-score, calculated following the Eq. (3), effectively expresses the difference between the mean

of the laboratory and the assigned value in the units of the target standard deviation (

p

).

𝑧 =

(3)

Zeta-score, calculated following the Eq. (4), states, if the participant result agrees with the

assigned value within the respective uncertainties. The denominator in the Eq. (4) is calculated

from the combined uncertainty of the assigned value and the measurement uncertainty reported

by the respective participant.

(13)

4

Zeta =

(4)

Where:

x

lab

: Reported result by the participating in the ILC laboratory (express as the mean of multiple determination)

x

ass

: Assigned value

p

: Target standard deviation

u

xlab

: Standard uncertainty reported by the participating in the ILC laboratory u

ref

: Standard uncertainty of assigned value

The interpretation of laboratory’s performance was evaluated according to the following internationally accepted limits [4]:

│ z or Zeta│ ≤2 Satisfactory 2< │ z or Zeta│ <3 Questionable

│ z or Zeta│3 Unsatisfactory

5 RESULTS AND DISCUSSION

5.1 OVERVIEW OF THE RESULTS

Forty-nine sets of data were submitted comprising 568 analytical results for 21 elements.

z-

scores and Zeta-scores were evaluated only for 16 elements.

75% of participants reported results for at least half part of requested analytes. For Zn, Fe and Cu more than 75% of participants reported their obtained measurement results. On the other hand, only 12% of laboratories (6) reported results for methyl mercury mass fraction in the fish sample, while 35 could reported only results total mercury mass fraction.

Graphical presentations of reported results and Kernel density plots [2] (in the case were more than 8 measurement results were reported) are presented in Appendix I. z-score, Zeta-scores and summary of statistical evaluation for the assessed elements are presented in the Appendix I.

All reported by participants in this ILC measurement results, are compiled in the Appendix II.

(For editorial purposes some results have been rounded to appropriate number of significant figures).

5.2 LABORATORY PERFORMANCES 5.2.1

z-scores:

Table 2 shows the overall performance (z-scores) of laboratories element by element. Figures 1 and 3 summarize

z-scores of the participating in the ILC laboratories by element and by

participating laboratory.

The

z-score compares the bias of the reported result from the assigned value with the target

standard deviation (p for proficiency assessment). p is defined by the ILC organizer as the

(14)

5 maximum acceptable standard deviation (25%) of the assigned value for investigated trace elements.

In total from 463 z-scores calculated, 81% were with │z│ ≤ 2, 87% with │z│ < 3, and 13%

were considered as unsatisfactory with │z│ > 3. Among 49 datasets, 13 (27%) were with all

│z│ ≤ 2 and 23 (47%) with │z│ < 3. On the other hand, 6 datasets (12%) have more than half of their results considered as unsatisfactory with │z│ ≥ 3.

As shown on Figure 1 the proportion of the acceptable results per element was 50% or higher.

Silver, cadmium and lead appeared to be the elements reported with the highest number of unsatisfactory z-scores. This result shows that the accurate determination of those elements in biological material is challenging, particularly at low concentration levels. For cadmium and silver most of unsatisfactory results, 86% and 73% were related with overestimation of the respective mass fractions. As these analytes have the lowest mass fractions, the observe results are probably arising from the contamination during sample preparation (e.g., digestion) or instrumental step. The laboratories concerned should carefully check their analytical procedures (e.g., quality of purified water and reagents applied) and try to improve their working laboratory environment. Laboratories should also develop an effective scheme for cleaning of the lab ware and regularly control this process.

Erroneous calibration standards could be another source of measurement bias. Only standards (CRM) with stated SI traceability should be used for calibration purposes. It is important to note that losses related with the working standard solutions at low concentration levels, are leading to the overestimation of the concentrations of elements in the samples (e.g., standard solutions should not be stored for an extended period).

Laboratories with questionable and unacceptable results should carefully check the respective laboratory procedures and working instructions, related with the analytical method used in this ILC.

5.2.2 Zeta scores:

Table 5 shows the overall performance (Zeta-scores) of laboratories for all studied trace elements. Figure 2 and 4 summarizes Zeta-scores of the participating laboratories in the ILC by trace element and by participating laboratory.

The Zeta-score shows if the laboratory result agrees with the assigned value within the respective uncertainties. The denominator in the Eq. (4) is the combined uncertainty of the assigned value and the measurement uncertainty as reported by the participating laboratory.

Nine laboratories (18%) didn’t provide the uncertainty of their reported results and Zeta-scores could not have been calculated.

As it can be seen on Figures 1 and 2, the comparison of measurement performances evaluated with

z-score and Zeta score clearly indicate that the number of unsatisfactory Zeta-scores is

slightly higher than the number of unsatisfactory z-scores (13% for z-scores and 19% for Zeta- scores). Only 3 laboratories (6%) could report 100% of their results with │z│ and │Zeta│ ≤ 2.

As the Zeta-score is the evaluation parameter, reflecting all parts of the measurement process, laboratories with unsatisfactory Zeta-scores should invest additional efforts in the proper evaluation of measurement uncertainty. Obtained results show that they are still remaining problems with the realistic estimation of measurement uncertainty.

If should be mentioned here that an unsatisfactory Zeta-score can be also caused by an

inappropriate estimation of the mass fraction of the respective trace element.

(15)

6

│z│ 3, 2< │z│ <3, │z│ ≤ 2

FIG. 1. The z-scores of results reported by the participants per element.

│Zeta│ 3, 2 < │Zeta│ < 3, │Zeta│ ≤ 2

FIG. 2. The Zeta-scores of results reported by the participants per element.

(16)

7

│z│ 3, 2< │z│ <3, │z│ ≤ 2 FIG. 3. The z-scores of results reported by the participants per laboratory. │Zeta│ 3, 2< │Zeta│ <3, │Zeta│ ≤ 2 FIG. 4. The Zeta-scores of results reported by the participants per laboratory.

(17)

8

TABLE 4. OVERALL ASSESSMENT OF LABORATORIES PERFORMANCE (z-SCORE) BY ELEMENTS Lab. Code Ag AsCa Cd Cu FeHgK MeHgMgMnPb SeSrVZn 1 -4.26 -5.97 -2.59 -7.99 -0.38 -7.62 -0.35 2 0.53 -3.57 -3.29 -0.05 -0.77 -1.79 -1.69 -1.30 -1.61 -2.32 -0.96 -0.97 -1.16 -0.05 4 -0.08 5 0.03 -0.58 -0.32 -1.24 0.00 -0.21 -0.74 -0.32 6 0.25 0.80 -0.23 -0.41 -0.20 0.70 1.10 1.21 0.82 -0.72 1.50 0.75 1.95 1.59 7 -0.81 -2.35 2.71 -0.17 8 0.10 104.12 -0.13 0.62 -1.30 -1.34 -0.28 -1.53 -0.06 0.72 0.48 0.10 -0.63 0.32 2.12 -0.95 110.48 0.68 -0.91 -0.58 -0.64 0.24 0.74 0.66 0.31 0.39 -0.10 0.65 -2.77 -0.45 12-1.28 0.65 0.93 3.58 0.10 0.24 0.73 0.68 3.36 -0.54 130.48 0.85 0.60 -0.98 0.09 -0.65 1.27 -0.30 0.30 -0.35 1.06 0.02 140.58 -0.04 0.71 -0.13 0.53 0.82 1.17 0.63 2.05 0.42 15-0.07 -1.41 -1.95 -2.34 0.10 -0.14 -2.65 -5.18 -0.74 3.31 13.44 1.39 1614.46 16.31 -2.63 1.96 -7.22 -6.61 1.77 -0.50 21.33 17-3.04 7.24 -0.87 187.32 1.54 1.13 36.44 0.76 190.52 -2.13 -0.22 1.59 0.14 -5.73 0.44 -1.08 0.00 -0.89 2058.67 -1.20 -0.54 -0.15 -1.51 2.10 -1.17 -0.34 0.98 -7.03 0.01 0.67 21-1.04 -2.51 1.64 0.28 0.81 0.71 0.00 -6.69 0.65 0.35 22-2.31 -1.88 1.96 0.61 -0.11 0.05 -1.64 0.09 23-0.06 0.33 -0.31 0.15 -0.18 0.57 1.02 -0.64 0.18 1.98 0.27 250.48 -0.07 -0.56 -0.72 -0.18 1.69 -0.43 -2.80 -0.10 26-7.87 224.42-2.21 -7.59 -7.88 -7.98 11.79 -7.65 27-0.58 1.37 -1.28 -0.91 -0.73 0.47 -1.13 28-3.96 -0.43 0.49 66.41 -3.49 1.59 -1.21 -7.19 1.15 -0.39 -2.63 -0.15 -0.43 1.04 -0.06 29-1.22 5.49 -0.30 -0.70 -1.60 -0.20

(18)

9 TABLE 4. OVERALL ASSESSMENT OF LABORATORIES PERFORMANCE (z-SCORE) BY ELEMENTS (cont.) Lab. Code Ag AsCa Cd Cu FeHgK MeHgMgMnPb SeSrVZn 30-1.40 1.23 -0.25 0.18 -7.99 -7.99 -0.79 0.08 -0.17 310.16 0.09 0.37 0.02 1.85 0.34 1.65 -0.01 -0.35 0.44 0.93 -0.43 0.63 -0.82 0.06 35-1.21 2.32 -0.43 -0.21 -0.28 0.61 -0.670.56 0.39 0.36 -0.59 0.43 0.76 1.29 0.68 36-1.64 0.08 -0.25 -1.18 -1.34 0.12 -0.98 0.16 -0.45 0.37 -1.27 -0.16 -0.01 -0.55 380.28 -0.05 3.16 420.69 -0.32 4.24 1.15 1.07 0.91 0.08 0.65 0.24 -0.78 0.19 434.93 0.31 1.72 -0.31 -2.34 -0.49 0.41 1.50 0.83 -0.75 -2.59 0.22 -0.33 -0.37 443.08 0.34 0.47 0.25 45-0.50 1.88 7.43 0.71 0.90 -2.47 -2.18 -1.40 460.04 0.68 0.09 -0.72 -0.08 0.89 1.21 0.20 1.28 -0.27 -0.02 -2.71 -0.42 470.54 -1.88 1.17 -0.78 0.13 -1.11 -1.39 0.58 0.07 48-3.04 -0.49 -2.29 49-0.68 -0.33 -0.37 7340 -0.36 -0.87 -1.14 -0.49 -0.68 52-0.04 44.09 -2.50 -1.46 -0.32 -1.83 4.24 -1.71 5322.30 18.51 -1.00 140 25.51 1.78 1.33 -1.32 -0.60 3.97 108 10.40 -1.45 4652.50 1.40 540.47 -0.81 -0.24 -0.51 1.32 0.27 0.16 5584.93 0.34 0.83 48.22 -0.20 -1.24 -0.13 -0.51 -0.79 1.02 0.12 0.69 -3.48 0.05 568112 8450.52 57-0.95 0.99 0.46 -0.77 -0.91 0.46 0.59 0.74 0.82 0.97 0.25 1.19 0.52 0.04 0.16 1.61 58-1.01 -0.27 1.00 -1.63 1.18 -0.24 0.93 2.13 0.77 1.50 -1.93 -0.12 59-0.81 1.35 -0.43 0.02 -0.50 0.13 -0.69 0.11 0.80 0.99 0.03 0.70 1.59 -0.10 -2.68 0.76 60-1.74 2.17 0.34 0.55 -0.27 0.48 -0.39 -0.27 62-1.39 -0.98 0.49

(19)

10

TABLE 5. OVERALL ASSESSMENT OF LABORATORIES PERFORMANCE (ZETA-SCORE) BY ELEMENTS Lab. Code Ag AsCa Cd Cu FeHg K MeHg MgMnPb SeSrVZn 1 2 0.20 -4.41 -2.98 -0.05 -0.84 -2.72 -1.82 -1.34 -2.13 -2.40 -1.21 -1.00 -0.88 -0.07 4 5 0.11 -0.84-0.50 -3.28 0.00 -0.41 -1.02 -1.22 6 0.83 0.93 -0.13-0.41 -0.27 2.66 1.21 1.23 1.16 -1.17 3.55 1.04 0.93 4.04 7 -0.16 -3.74 4.87 -0.08 8 102.19 -0.12 0.32 -1.89 -1.53 -0.16 -0.77 -0.07 0.74 0.54 0.06 -0.55 0.29 2.46 -1.73 110.71 0.66 -1.53-0.80 -0.93 0.19 0.28 0.65 0.41 0.24 -0.16 0.70 -2.98 -0.72 12-4.59 0.91 1.72 5.43 0.17 0.52 1.07 0.94 3.74 -1.40 131.32 1.13 0.98 -1.21 0.15 -0.87 1.34 -0.50 0.54 -0.57 1.44 0.03 141.74 -0.08 0.78 -0.35 0.69 1.26 2.13 1.55 2.13 1.10 15-0.16 -2.06 -1.50 -4.02 0.15 -0.18 -3.03 -5.77 -0.72 4.67 5.25 3.13 162.16 1.64 -4.56 3.44 -31.37-16.582.74 -0.73 45.13 17 18 190.75 -1.93 -0.35 1.71 0.17 -5.02 0.50 -1.84 0.01 -2.16 2015.99 -3.83 -0.75 -0.06 -1.72 2.42 -1.63 -0.28 1.26 -15.160.02 2.56 21-1.57 -7.75 2.37 0.47 1.34 1.47 0.00 -15.020.81 0.68 22-9.38 -2.12 2.37 0.84 -0.34 0.08 -1.69 0.25 23-0.09 0.42 -0.35 0.18 -0.20 0.71 1.22 -0.94 0.20 1.87 0.40 25 26-11.6037.58 -3.77 -13.64-11.57-11.2918.20 -33.72 27-0.68 0.97 -1.87 -1.08 -0.81 1.13 -2.80 28-3.03 -1.83 0.66 2.68 -5.90 2.39 -3.50 -10.471.55 -0.59 -5.26 -0.35 -0.62 0.80 -0.27 29-0.92 2.69 -0.44 -1.17 -3.12 -0.33

(20)

11 TABLE 5. OVERALL ASSESMENT OF LABORATORIES PERFORMANCE (ZETA-SCORE) BY ELEMENTS (cont.) Lab. Code Ag AsCa Cd Cu FeHg K MeHg MgMnPb SeSrVZn 30-1.86 1.82 -0.33 0.39 -11.73-11.30-1.13 0.10 -0.21 310.27 0.23 0.49 0.03 2.51 0.49 2.66 -0.01 -0.44 0.77 1.68 -0.83 0.84 -0.94 0.16 35-2.16 5.85 -0.64 -0.29 -0.35 1.08 -1.59 0.75 0.54 0.88 -1.30 1.15 1.01 1.26 2.98 36 380.33 -0.09 3.96 420.39 -1.10 3.84 1.52 1.61 2.73 0.20 1.02 0.34 -0.94 0.81 439.09 0.15 1.07 0.00 -3.09 -0.38 0.49 0.71 0.56 -0.67 -3.30 0.11 -0.24 -0.46 443.48 0.57 0.57 0.56 45-1.08 2.77 3.58 1.17 1.32 -2.23 -3.21 -5.10 460.02 0.46 0.09 -0.62 -0.06 0.47 0.78 0.16 1.02 -0.13 -0.01 -2.58 -0.53 470.41 -1.71 0.77 -0.64 0.11 -0.89 -1.29 0.42 0.06 48-1.23 -1.41 -3.94 49-1.06 -0.45 -0.57 108.81-1.83 -1.39 -0.72 -2.84 52-0.05 8.85 -4.16 -2.55 -0.46 -2.57 6.45 -7.02 536.08 11.24 -1.47 7.09 12.47 2.02 1.19 -1.92 -0.85 4.52 9.63 6.00 -1.55 11.50 1.91 542.01 -1.49 -0.39 -2.29 2.95 0.58 0.71 55 566.11 12.54 57 58-1.67 -1.12 1.43 -3.02 1.61 -0.35 1.37 2.99 1.92 2.39 -4.71 -0.47 59-0.92 1.62 -0.56 0.02 -0.57 0.13 -0.91 0.10 0.50 1.27 0.03 0.99 1.16 -0.12 -3.04 0.55 60-2.14 4.01 0.45 0.96 -1.08 0.60 -0.75 -0.61 62-1.81 -2.39 0.43

(21)

12

5.3 ANALYTICAL METHODS

Analytical methods used by the participating in this ILC laboratories are presented on Figure 5. Generally, they can be divided to three groups: nondestructive techniques (NAA, XRF);

plasma spectrometric methods (ICP-MS and ICP-OES) and atomic absorption spectrometry methods, representing 10%, 60% and 20% respectively. Abbreviation used in the figure 5 and appendix II are shown in Table 6.

FIG. 5. Graphical distribution of instrumental techniques.

TABLE 6. ABBREVIATIONS OF INSTRUMENTAL TECHNIQUES USED IN THE INTERLABORATORY COMPARISON

Method code Instrumental technique

AAS Atomic Absorption Spectrometry

AFS Atomic Fluorescence Spectrometry

GC Gas Chromatography

ICP-MS Inductively Coupled Plasma Mass Spectrometry

ICP-OES Inductively Coupled Plasma Optical Emission Spectrometry

NAA Neutron Activation Analysis

XRF X-Ray Fluorescence Spectrometry

(22)

13 5.4 SAMPLE TREATMENT, IMPACT OF WATER CONTENT, USE OF CRM AND

RECOVERY CORRECTIONS

Eleven participants (22%) didn’t report the obtained for the CRM results as requested in the reporting form. Eight participating laboratories didn’t provide QC results for all requested elements. An important principle for the selection of reference material by laboratories was the principle of matrix and concentration range matching. CRMs used in this ILC were appropriately selected as most of laboratories used biota matrices of marine origin (i.e. IAEA 407 Tuna fish and IAEA 461 Clam from the IAEA, SRM 2976 mussel from the NIST).

Sixteen participants (32%) claimed to be accredited, but only part of them were accredited for the determination of trace element in fish sample matrix. 4 participants (25%) from accredited laboratories didn’t report results for QC sample. In general, results reported by accredited laboratories were comparable with results reported by non-accredited laboratories. Only one of accredited laboratory (6%) and 6 of non-accredited laboratories (36%) reported 100% of their results with │z│ ≤ 2.

Only 6 laboratories (12%) claimed implementing correction for recovery for at least part of their reported results, but 25 participants didn’t provide results for recoveries. Most of recoveries reported were in the range of 100 ± 25%. High proportion of the laboratories that didn’t perform correction for recovery have obtained satisfactory scorings, meaning that the laboratories have correctly estimated that the recovery achieved was not significantly different from 100%.

The ILC fish sample was subjected to freeze drying as part of its preparation process. At the time of bottling, the moisture content of the material was around 6.5%. Depending on local storage conditions and humidity levels the ILC sample might absorb moisture from the environment. Consequently, users were advised to make a separate determination of the moisture content of the material. As the moisture is operationally dependent parameter [5]

procedure on moisture content determination in the test sample was provided to all participating laboratories in the accompanying letter. Only 27 participants (55%) claimed to report results corrected for moisture, 14 of them have used effectively the prescribed by the organizers protocol (85°C). Other 13 laboratories didn’t provide their methodology for moisture determination. Inadequate determination of moisture can be a source of bias, especially for biological matrices, where the moisture content is often more than 5%. The moisture content reported by the laboratories that applied a correction factor was in the range from 0.05 to 11%.

It should be noted that out of 49 data sets received, 12 participants didn’t fill the questionnaire as requested.

6 RECOMMENDATIONS

Participants are recommended to review their data element by element, appraising whether the

z-scores and Zeta-scores are less than or equal to 2. The use of z-scores and Zeta-scores will

help to identify systematic errors in the measurement results (e.g. from calibration, reagent contamination or incomplete digestion) and should ultimately improve data quality.

Laboratories should investigate all unsatisfactory scores (i.e. |z| or |Zeta| > 3) and put in place

the necessary corrective actions in order to prevent the problem reoccurring. This is also a

requirement for accredited according to the ISO/IEC 17025 standard laboratories. [6].

(23)

14

Participants reporting total mercury in fish sample are strongly suggested to look at the implementation of methodologies for the measurement of mercury speciation’s, when the obtained measurement results are used for risk assessment purposes.

Some laboratories still need to improve quality assurance / quality control procedures, to implement regular analysis of certified reference materials and the use of quality control chart in their daily laboratory practice. This way of work provides continuous feedback to the analyst and is an essential tool for the monitoring of data quality and for the production of reliable measurement results, used in the monitoring and risk assessment studies.

A full catalogue of available IAEA reference materials is published regularly and can be consulted on the IAEA website: http://www.iaea.org/programmes/aqcs.

7 CONCLUSIONS

Although the overall performance of participating in this ILC laboratories is quite satisfactory, it must be pointed that for elements present at low to ultra-low concentration levels (i.e Cd, Pb) in fish matrix, there is clear danger of contamination problems. It is a subject of concern as these elements are toxics and the accurate determination of their content in seafood is of crucial importance.

The implementation of Minamata convention, especially article 19, should lead to increasing number of laboratories involved in the monitoring of mercury but also on mercury species, especially in fish samples. The need for improvement in this analytical field is obvious, as only 12% of participants in the ILC are in a position to report obtained measurement results for CH

3

Hg.

An extra effort is needed for relevant evaluation of measurement uncertainties, associated with measurement results. There is still almost 20% of participants that do not report results with associated uncertainties. As a result, the number of unsatisfactory Zeta-scores was systematically higher than the number of unsatisfactory z-scores for the same trace elements.

The uncertainty associated with measurement results is of paramount importance in the frame of different regulations and international agreements, and it is important for any analytical laboratory to report a complete uncertainty statement.

In general, it should be noted that uncertainty evaluations based only on the precision of

measurement results are very often underestimated. In many cases, they do not include other

major contributors, coming from uncertainties on the determination of recovery, procedural

blank, moisture content etc.

(24)

15 APPENDIX I

PERFORMANCE EVALUATION BY ELEMENT IN

IAEA-MESL-ILC-TE-BIOTA-2017

(25)

16

Evaluation of Reported data for Ag

Kernel density Plot

Summary of results:

Satisfactory Questionable Unsatisfactory

z-score 59% 0% 41%

Zeta-score 46% 23% 31%

Xass: 0.066 mg kg-1

Uass (k=2): 0.008 mg kg-1

2p: 0.017 mg kg-1

Number of results: 17

Number of method: 3

Reported results and expanded uncertainties:

Performance evaluation:

z-score Zeta-score

___ Xass ; Xlab ± Ulab; ---- Xass± 2p ;---- Xass± Uass(k=2)

(26)

17 Evaluation of Reported data for As

Kernel density Plot

Summary of results:

Satisfactory Questionable Unsatisfactory

z-score 87% 10% 3%

Zeta-score 69% 8% 23%

Xass: 19.9 mg kg-1

Uass (k=2): 1.1 mg kg-1

2p: 5.0 mg kg-1

Number of results: 30

Number of method: 6

Reported results and expanded uncertainties:

Performance evaluation:

z-score Zeta-score

___ Xass ; Xlab ± Ulab; ---- Xass± 2p ;---- Xass± Uass(k=2)

Références

Documents relatifs

The radiation protection and safety principles established in the IAEA safety standards (see Refs [1, 3]), on the basis of the recommendations of the International Commission

With a view to ensuring the protection of people and the environment from harmful effects of ionizing radiation, the IAEA safety standards establish fundamental safety

[1.5] INTERNATIONAL ATOMIC ENERGY AGENCY, Sediment Distribution Coefficients and Concentration Factors for Biota in the Marine Environment, Technical Reports Series No.. 422,

All participating laboratories have been requested to use validated methods for the determination of requested trace elements in the tuna fish flesh homogenate sample and to

Based on the results of the proficiency test presented in the report each participating laboratory should assess its analytical performance results by using the specified

Sill [85] published a very thorough and detailed study about the application of a lead sulphate co-precipitation for enrichment and separation of radium and a barium

As part of these activities, a new PT for the Helsinki Commission’s Project for Monitoring Radioactive Substances in the Baltic Sea (HELCOM-MORS) was organised to test the

The proficiency testing scoring system applied by the Chemistry Unit in the Agency’s laboratories takes into consideration the trueness and the precision of the reported data and it