• Aucun résultat trouvé

Combination results in 3-D data providing both quantitative and qualitative information.

N/A
N/A
Protected

Academic year: 2022

Partager "Combination results in 3-D data providing both quantitative and qualitative information."

Copied!
8
0
0

Texte intégral

(1)

Combination results in 3-D data providing both quantitative and qualitative information.

time

response

One critical aspect of GC-MS and LC- MS is the interface.

Both chromatographic approaches have relatively large flows.

GC - 1 - 50 ml / min LC - 0.1 - 5 ml / min

- 50-5000 ml/min as a gas The mass spec requires a vacuum of

about 10-5 - 10-6 torr - gas phase.

flow from chromatograph

MS

waste

Simplest approach would be to split the flow.

Only allow what the MS can handle to actually enter.

This can result in only 1/1000 or less of the sample to enter the MS - not good for trace work

(2)

Vacuum source from

GC

to MS

Most popular approach when packed columns must be used.

Based on concept that larger molecules will diffuse more slowly so more will reach the MS entry jet.

Relative simple and inexpensive approach.

from GC to MS

He in He out

flow restrictor liner

Somewhat similar to a jet separator. The MS pulls in about 1 ml/min through the flow restrictor.

If column flow is above that - excess is vented.

If flow is <1 ml/min, He from external source is pulled in.

Best for sources that have flows close to 1 ml/min like capillary columns.

(3)

end of column

MS ionization source

If we limit interface to capillary columns, the MS can actually use all column effluent.

Now that we have an interface, how do we justify have a detector that costs more than the chromatograph?

This is not a course in MS so we’ll limit our discussion to using the MS as a

chromatographic detector.

I’m assuming you know the basics of MS instrumentation and interpretation.

If your scan rate is too high, you don’t obtain enough scans to properly define your peak. This results in a loss in precision.

(4)

Maximum sensitivity and precision is obtained under conditions when minimum qualitative information is collected.

We need some way to maximize both types of information

- Do multiple runs

- Target compound analysis

A simple approach to obtain both quantitative and qualitative analysis.

Works best if you are only dealing with a limited number of analytes.

You also want to have an isotope labeled internal standard.

(5)

Example

Analysis of a drug in urine

time

response

145 179 214

142 176 211

trideutrated internal standard

Sample

•  Sample is heated and allowed to rapidly expand into a vacuum.

•  Solvent is quickly pulled from component drops and a ‘static’

charge is produced.

•  Charged particles enter into the MS via a skimmer.

MS

to vacuum column

heaters vapor

jet

A charge is transferred to the droplets.

They are then sprayed into a vacuum chamber where the volatile solvent is

‘pulled off’.

At some point, sample ions are ejected from the drop and enter the MS.

to vacuum nitrogen

nitrogen

to MS high voltage

from LC

(6)

Thermospray and Electrospray both result in a solvent mist that is electrostatically charged.

As the drops are pulled towards the skimmer, solvent is evaporated.

This increases the charge density, causes the drop to further disperse and ultimately transfer the charge to the analyte.

e-

e-

e- e-

e- e-

e-

e- e-

e-

From Source

Towards Analyzer

e-

e- e- e-

e- e-

e- e- e-

e-

e-

e- e- e-

e- e-

e- e- e-

e- Evaporation

e-

e- e- e-

e- e-

e- e- e-

e-

e- e-

e- e-

e- e- e-

e- e- e- Breakup

e- e-

e- e-

e- e- e-

e- e- e-

e- e-

e- e- e- e- e-

e- e- e-

(7)

FTIR GC

alternate detector injection port

detector IR source

interferometer

flow cell

Two types of data are produced IR spectra for each scan Gram-Schmidt Chromatogram

Full FT transformation of data is slow where as GS calculation is fast - it only results in one point per scan.

This permits real time display of your data.

This approach relies on doing plasma emission on your sample.

Hewlett-Packard offers a small plasma emission detector based on a microwave plasma source.

Relative expensive and sensitivity is very dependent on element and lines used

(8)

vent

reagent and makeup gas

column

window water

in

water out

plasma

plasma source

condensing mirror

holographic grating

movable photodiode array slit

order sorter

•  Produces emission spectra for each element.

•  A reagent gas may be needed to insure proper atomization and excitation.

•  Can obtain elemental analysis and empirical formula.

•  Limited at this point - can only look at a limited number of lines (due to small range of photodiode).

Références

Documents relatifs

Après avoir montré l’intérêt des diatomées littorales pour l’évaluation écologique en microscopie sur quelques zones restreintes du Léman (Rimet et al., 2016a),

We approach the task from information retrieval perspec- tive using a relatively small number of labeled videos in the de- velopment set to mine query expansion terms

Our main goal is to introduce the architecture of prototype of such p2p network and search request routing/data storing/indexing protocols for it, that allows any node to

If a stochastic model of the physical randomness source is available, it can be used in combination with the raw signal to estimate the entropy and the bias depending on random

More precisely, any interaction effect can be expressed as the product of the additive and dominance effects among the genotypic effects at each locus times the

CONCLUSION GÉNÉRALE Nous avons tenté dans le cadre de ce travail d’approfondir la relation entre le développement du secteur financier, la supervision bancaire et la

In Section 3, we present an algorithm for the problem of minimizing the total number of communications in the UET-bound model, for the special case of a tree and an unlimited number

Multiple high-dose albendazole was clearly unsuitable as a treatment regimen for bancroftian filariasis, but given the macrofilaricidal activity and microfilaraemic