AnnalesdeToxicologie Analytique, vol.
XVII,
n° 1,2005A rapid and sensitive liquid chromatography - tandem mass spectrometry method for the
determination of amphetamine and related designer drugs in urine
Une méthode par chromatographie liquide cou¬
plée à la spectrométrie de masse en tandem rapide et sensible pour le dosage de l'amphéta¬
mine et de drogues de synthèse dans les urines
Charlotte MATTHYS
»,Alain VERSTRAETE
(W)*(1) Department
Clinical
Chemistry andToxicology, GhentUniversity Hospital -Belgium (2)Departmentof
Clinical Biology, MicrobiologyandImmunology,Facultyof
Medicine,GhentUniversity-Belgium*Author for
correspondence: Prof. Dr.Alain
VERSTRAETE, Laboratoryof Clinical Biology
-Toxicology, Ghent University HospitalDe Pintelaan 185 -B-9000GENTBELGIUM
Phone+ 32 9 240 3407 -Fax+ 329 240 49 85 -janvier
2005; acceptéaprèsmodificationsle 7avril
2005)SUMMARY
A method
for
the directanalysisof
six amphetamine com¬poundsin urinewasdeveloped usingliquidchromatography tandemmassspectrometry (LC-MS/MS). Weadded 90
pi of
a solution
of
internal standards (1 pg/mLof
d5-AMP, d5- MET,d5-MDA,d5-MDMA, d5-MDEAandd5-MBDB)to 10pi of
urinefollowed, by vortex-mixing and centrifugation. The samplesolutions were analyzed byLC-MS/MSin theMRM mode after separation on a reversed-phase C18 column usinggradientelution. Separationanddetectionof
allcom¬pounds was accomplished within eight minutes. Linearity was established
for
all compounds,, from 78 to 100000 ng/mL. Correlation coefficientsfor all
analytes exceeded 0.998.Thelowerlimit of
quantificationwas 10ng/mLfor all
compounds, except
for
AMPandMDA (78 ng/mL). Within- day imprecision(CV%) and between-day CVs (78, 625andRESUME
Uneméthodepourl'analysedirectedesix amphétaminiques dans les urines a étédéveloppée en utilisant la chromato¬
graphieliquidecoupléeàlaspectrométrie demasse entan¬
dem (LC-MS/MS). Nous avons ajouté90
pi
d'un mélange d'étalons internes (1 pg/mL ded5-AMP, de d5-MET, de d5- MDA, dedyMDMA, de ds-MDEA et ded5-MBDB) à 10pi
d'urine, mélangéparvortex et centrifugé. Leséchantillons ontétéanalysésparLC-MS/MSenmodeMRMaprèssépa¬ration surune colonne Cl8à phase inverse enutilisantun gradient d'élution. Laséparation etladétection detous les composésontétéaccompliesenhuitminutes. Lalinéaritéa étéétabliepourtous les composés, de 78à 100000 ng/mL.
Les coefficients de corrélation étaient supérieurs à 0.998.
Leslimitesdequantificationétaientinférieures à10ng/mL, saufpourl'amphétamine et laMDA (78 ng/mL). Larépéta- 65 AnnalesdeToxicologie Analytique, vol.
XVII,
n° 1,2005A rapid and sensitive liquid chromatography - tandem mass spectrometry method for the
determination of amphetamine and related designer drugs in urine
Une méthode par chromatographie liquide cou¬
plée à la spectrométrie de masse en tandem rapide et sensible pour le dosage de l'amphéta¬
mine et de drogues de synthèse dans les urines
Charlotte MATTHYS
»,Alain VERSTRAETE
(W)*(1) Department
Clinical
Chemistry andToxicology, GhentUniversity Hospital -Belgium (2)Departmentof
Clinical Biology, MicrobiologyandImmunology,Facultyof
Medicine,GhentUniversity-Belgium*Author for
correspondence: Prof. Dr.Alain
VERSTRAETE, Laboratoryof Clinical Biology
-Toxicology, Ghent University HospitalDe Pintelaan 185 -B-9000GENTBELGIUM
Phone+ 32 9 240 3407 -Fax+ 329 240 49 85 -janvier
2005; acceptéaprèsmodificationsle 7avril
2005)SUMMARY
A method
for
the directanalysisof
six amphetamine com¬poundsin urinewasdeveloped usingliquidchromatography tandemmassspectrometry (LC-MS/MS). Weadded 90
pi of
a solution
of
internal standards (1 pg/mLof
d5-AMP, d5- MET,d5-MDA,d5-MDMA, d5-MDEAandd5-MBDB)to 10pi of
urinefollowed, by vortex-mixing and centrifugation. The samplesolutions were analyzed byLC-MS/MSin theMRM mode after separation on a reversed-phase C18 column usinggradientelution. Separationanddetectionof
allcom¬pounds was accomplished within eight minutes. Linearity was established
for
all compounds,, from 78 to 100000 ng/mL. Correlation coefficientsfor all
analytes exceeded 0.998.Thelowerlimit of
quantificationwas 10ng/mLfor all
compounds, except
for
AMPandMDA (78 ng/mL). Within- day imprecision(CV%) and between-day CVs (78, 625andRESUME
Uneméthodepourl'analysedirectedesix amphétaminiques dans les urines a étédéveloppée en utilisant la chromato¬
graphieliquidecoupléeàlaspectrométrie demasse entan¬
dem (LC-MS/MS). Nous avons ajouté90
pi
d'un mélange d'étalons internes (1 pg/mL ded5-AMP, de d5-MET, de d5- MDA, dedyMDMA, de ds-MDEA et ded5-MBDB) à 10pi
d'urine, mélangéparvortex et centrifugé. Leséchantillons ontétéanalysésparLC-MS/MSenmodeMRMaprèssépa¬ration surune colonne Cl8à phase inverse enutilisantun gradient d'élution. Laséparation etladétection detous les composésontétéaccompliesenhuitminutes. Lalinéaritéa étéétabliepourtous les composés, de 78à 100000 ng/mL.
Les coefficients de corrélation étaient supérieurs à 0.998.
Leslimitesdequantificationétaientinférieures à10ng/mL, saufpourl'amphétamine et laMDA (78 ng/mL). Larépéta- 65
10000ng/mL) rangedfrom2.62to 16.26%andfrom0.86to 11.98%, respectively. Accuracy (bias%) lay between 0.16 and 7.17 %. Thepeak areas
of
theamphetamines added to urinefell intherange85-115% comparedtostandardsolu¬tions in methanol/water; exceptfor AMPand MDA. Carry¬
overwasnegligibleandstability afterstorageatroomtem¬
peraturefor up to 24h was acceptable. In conclusion, the presented method allows the accurate, precise and rapid determinationofsix amphetamine compoundsin urineover awideanalyticalrange.
KEY-WORDS
Amphetamines, MDMA, liquid chromatography, tandem massspectrometry, urine.
bilité (CV%) etla reproductibilitévariaient respectivement entre 2.62 et 16.26% et entre 0.86 et 11.98%. En ce qui concernel'exactitude, lepourcentagedebiaisà 78et10000 ng/mLvariaitentre0.16 et7.17%. Lasurfacedespics des amphétaminesajoutéesà del'urine variaitentre85 et 100%
decelledesamphétamines dissous dansunmélanged'eauet de methanol, exceptépour l'amphétamine et le MDA. Le carry-over était négligeable eï la stabilité (156 et 5000ng/mL) après stockage àla température ambiantepen¬
dant 24h étaitacceptable. En conclusion, la méthodepré¬
sentéepermet la détermination exacte, précise etrapidede six amphétaminiques dans les urines sur une plage de concentration large.
MOTS-CLÉS
Amphétamines, MDMA, chromatographie enphaseliquide, spectrométriede masse entandem, urine.
Introduction
In the last few decades, amphetamine designer drugs have gained popularity as recreational drugs and they areusedmainly
for
their stimulatingeffects, especially ingatherings known asraves and inthe dancing scene (1,2).Monitoring of
amphetamines anddesigner drugs in human urine is successfully usedfor clinical
and forensic applications.For most
clinical
and forensic applications,initial
screening is performed by an immunoassay, and pre¬sumptive positive samples are confirmed by a more specific method. Todate,theconfirmation
of
ampheta¬mines in urine samples is mainly performed by gas
chromatography-mass spectrometry
(GC-MS)(3).
Despite the many advantages
of
GC-MS, such as the highsensitivityandspecificityandits widespreadavai¬lability,
it
doeshavelimitations. Oneof
them,linkedto amphetamines, is that the compoundswith
the amphe¬tamine core structure have base peaks at low masses, resulting in interference
from
biological background.Thiscan beovercome by theuse
of
extractionfrom
the biologicalfluid,
followed by derivatisation, astep also neededfor
improving the GC-propertiesof
the com¬pounds.
A
majordrawbackof
derivatisation,specifical¬ly in a routine laboratory with a large number
of
samples to be analysedin ashorttime, is that thepro¬cedurebecomeslaborious andtime-consuming.
Headspacesolidphasemicro-extraction (SPME)isone potential solution to minimizethe timespent by tech¬
nical staff preparing samples
for
GC-MS analysis (4).The disadvantages, on the other hand,are the need
for
special equipment, the carry-over effect and the needfor
conditioningof
the fibre before use. These limita¬tions
of
GC-MS led to investigate alternative approachesfor
analysing amphetamines in biological fluids.Inthelastfewyears,
liquid
chromatography coupledto mass spectrometry (LC-MS) has developedrapidly in forensicandclinical
applications(5,6). SeveralLC-MS
interface typesaredescribed. Today,however,tworela¬tively robustLC-MS interfacetypesaremost frequent¬
ly
used, the atmospheric-pressure ionisation tech¬niques, electrospray (ESI) and atmospheric-pressure chemical ionisation (APCI).
LC-MS
offers a higher sensitivityand specificity and reduces sampleprepara¬tionrequiredwith GC-MSbecauserelativelynon-vola¬
tile
compoundscan beanalysed and no derivatisationis necessary.A
furtherdevelopmentis the combinationof
two mass spectrometerswith
an interposedcollision
cell. This characterizes LC-tandem mass spectrometry(LC-
MS/MS), which generally provides superiorlimit of
quantification (LOQ), sensitivityandimproved selecti¬vity. An extra advantage
of
MS-MS, inrespectof
MS, is theability
to shorten thechromatographic run-time dramatically.This paper describesthe validation
of
aliquid
chrorna- tography-APCI-tandem mass spectrometry method(LC-APCI-MS/MS) for
simultaneous analysisof
six amphetamine compounds in urine. This method is based onthemethodof
Nordgren etal.(7).Materials and Methods
Chemicals and reagents
Standard solutions
of
amphetamine (1mg/mL),
methamphetamine(1 mg/mL), 3,4-methylenedioxyam- phetamine(MDA)
(1 mg/mL), 3,4- methylenedioxy- methamphetamine(MDMA)
(1 mg/mL), 3,4-methyle- nedioxyethylamphetamine(MDEA)
(1 mg/mL),N-
methy1-1-(3,4-methylenedioxyphenyl)-2-butanamine(MBDB) (lmg/mL),
and rf5-deuterated analogues (100 10000ng/mL) rangedfrom2.62to 16.26%andfrom0.86to11.98%, respectively. Accuracy (bias%) lay between 0.16 and 7.17 %. Thepeak areas
of
theamphetamines added to urinefell intherange85-115% comparedtostandardsolu¬tions in methanol/water; exceptfor AMPand MDA. Carry¬
overwasnegligibleandstability afterstorageatroomtem¬
peraturefor up to 24h was acceptable. In conclusion, the presented method allows the accurate, precise and rapid determinationofsix amphetamine compoundsin urineover awideanalyticalrange.
KEY-WORDS
Amphetamines, MDMA, liquid chromatography, tandem massspectrometry, urine.
bilité (CV%) etla reproductibilitévariaient respectivement entre 2.62 et 16.26% et entre 0.86 et 11.98%. En ce qui concernel'exactitude, lepourcentagedebiaisà 78et10000 ng/mLvariaitentre0.16 et7.17%. Lasurfacedespics des amphétaminesajoutéesà del'urine variaitentre85 et 100%
decelledesamphétamines dissous dansunmélanged'eauet de methanol, exceptépour l'amphétamine et le MDA. Le carry-over était négligeable eï la stabilité (156 et 5000ng/mL) après stockage àla température ambiantepen¬
dant 24h étaitacceptable. En conclusion, la méthodepré¬
sentéepermet la détermination exacte, précise etrapidede six amphétaminiques dans les urines sur une plage de concentration large.
MOTS-CLÉS
Amphétamines, MDMA, chromatographie enphaseliquide, spectrométriede masse entandem, urine.
Introduction
In the last few decades, amphetamine designer drugs have gained popularity as recreational drugs and they areusedmainly
for
their stimulatingeffects, especially ingatherings known asraves and inthe dancing scene (1,2).Monitoring of
amphetamines anddesigner drugs in human urine is successfully usedfor clinical
and forensic applications.For most
clinical
and forensic applications,initial
screening is performed by an immunoassay, and pre¬sumptive positive samples are confirmed by a more specific method. Todate,theconfirmation
of
ampheta¬mines in urine samples is mainly performed by gas
chromatography-mass spectrometry
(GC-MS)(3).
Despite the many advantages
of
GC-MS, such as the highsensitivityandspecificityandits widespreadavai¬lability,
it
doeshavelimitations. Oneof
them,linkedto amphetamines, is that the compoundswith
the amphe¬tamine core structure have base peaks at low masses, resulting in interference
from
biological background.Thiscan beovercome by theuse
of
extractionfrom
the biologicalfluid,
followed by derivatisation, astep also neededfor
improving the GC-propertiesof
the com¬pounds.
A
majordrawbackof
derivatisation,specifical¬ly in a routine laboratory with a large number
of
samples to be analysedin ashorttime, is that thepro¬cedurebecomeslaborious andtime-consuming.
Headspacesolidphasemicro-extraction (SPME)isone potential solution to minimizethe timespent by tech¬
nical staff preparing samples
for
GC-MS analysis (4).The disadvantages, on the other hand,are the need
for
special equipment, the carry-over effect and the needfor
conditioningof
the fibre before use. These limita¬tions
of
GC-MS led to investigate alternative approachesfor
analysing amphetamines in biological fluids.Inthelastfewyears,
liquid
chromatography coupledto mass spectrometry (LC-MS) has developedrapidly in forensicandclinical
applications(5,6). SeveralLC-MS
interface typesaredescribed. Today,however,tworela¬tively robustLC-MS interfacetypesaremost frequent¬
ly
used, the atmospheric-pressure ionisation tech¬niques, electrospray (ESI) and atmospheric-pressure chemical ionisation (APCI).
LC-MS
offers a higher sensitivityand specificity and reduces sampleprepara¬tionrequiredwith GC-MSbecauserelativelynon-vola¬
tile
compoundscan beanalysed and no derivatisationis necessary.A
furtherdevelopmentis the combinationof
two mass spectrometerswith
an interposedcollision
cell. This characterizes LC-tandem mass spectrometry(LC-
MS/MS), which generally provides superiorlimit of
quantification (LOQ), sensitivityandimproved selecti¬vity. An extra advantage
of
MS-MS, inrespectof
MS, is theability
to shorten thechromatographic run-time dramatically.This paper describesthe validation
of
aliquid
chrorna- tography-APCI-tandem mass spectrometry method(LC-APCI-MS/MS) for
simultaneous analysisof
six amphetamine compounds in urine. This method is based onthemethodof
Nordgren etal.(7).Materials and Methods
Chemicals and reagents
Standard solutions
of
amphetamine (1mg/mL),
methamphetamine(1 mg/mL), 3,4-methylenedioxyam- phetamine(MDA)
(1 mg/mL), 3,4- methylenedioxy- methamphetamine(MDMA)
(1 mg/mL), 3,4-methyle- nedioxyethylamphetamine(MDEA)
(1 mg/mL),N-
methy1-1-(3,4-methylenedioxyphenyl)-2-butanamine(MBDB) (lmg/mL),
and rf5-deuterated analogues (100AnnalesdeToxicologie Analytique,vol.
XVII,
n° 1,2005pg/mL) used as internai standards (IS) in methanol were obtained
from Cerilliant (Austin,
Texas).Methanol (absolute) and water
for LC-MS
were pur¬chased
from Biosolve
(Valkenswaard, The Netherlands).Ammonium
acetate (p.a) was supplied by Sigma-Aldrich (Bornem, Belgium).Instrumentation and MS/MS conditions
An Agilent
1100 seriesHPLC
system(Agilent
Technologies) consistingof
a pump, column oven, autosampler and degasser wereusedfor
solventdelive¬ry and sample introduction. The injected volume was 20 pi. Analytes were separated at40°C on a 2.1 x 30 mm Zorbax SB
-CI
8. Rapid Solution column(Agilent
Technologies).Thecolumnwaseluted ataflow
rateof
0.3
mL/min
anddevelopedwith
gradientelutionasfol¬lows:0-0.2min,95%A/5%B; 1.2-4.5min,5%A/95%B and4.8-8 min 95%A/5%B
(A:
H20 +2mM
ammoniu- macetate, B:MeOH
+2mM
ammoniumacetate).The
LC-MS/MS
system consistedof
anAPI 2000
triple-quadrupolemass spectrometer equippedwith
anAPCI
interface(Applied
Biosystems/MDS Sciex, Langen, Germany) usedin the positive-ion mode. The sixamphetamine compounds were detectedinthemul¬tiple-reaction monitoringmode. Two
MRM
transitionsfor
eachsubstanceweremonitoredtoprovidesufficient identificationof
the amphetamine compounds. The chosenMRM
transitionsfor
each amphetamine com¬pound and d5-deuterated analogue are summarised in table 1. The entrance potential varied
from
5.5V
toTable
I
:Retentiontime,parentionandthechosendaughter ionsfor
each amphetamine compound and d5-deuterated analogueAmphétaiijinesRetention
M+l
MRM1 MRM2 , nine(min) ;:AMP 4.20 136.079 91.05 65.05
MET 4.36 150.078 91.05 119.15
MDA 4.29 180.119 135.15 133.05
MDMA 4.38 194.085 163.05 105.05
MDEA 4.50 208.068 163.05 105.15
MBDB 4.61 208.068 135.05 177.15
d5-AMP 4.17 141.112 93.35
ds-MET 4.35 155.09 92.35
ds-MDA 4.26 185.114 110.15 ds-MDMA 4.35 199.136 165.15 ds-MDBA 4.48 213.075 163.05
*-MBDB 4.60 199.136 165.15
9 V,thecollision cellentrancepotentialvaried
from
14 to 20V
andthecell exit potentialwas setat2V
or4V, accordingtothe analyte.Analyst
Software (Ver. 1.3.1;Applied
Biosystems/MDS Sciex) was usedfor
HPLC system control,dataacquisition, anddataprocessing.Calibration standards and internal stan¬
dard mix-solution
Calibration standards were prepared
in
drug-freeurinefrom
methanolicstocksolutions,containingallamphe¬tamine compounds ataconcentration
of
1 mg/mL. The concentrationsof
thecalibration
standards were 78.125, 156.25, 312.5, 625, 1250,5000, 10000, 20000, 30000,40000, 50000, 60000, 70000, 80000, 90000and 100000ng/mL.Forrecovery testing, standards
with
concentrationsof
78.125, 625 and 10000ng/mL
were made in H20/MeOH (50/50) solutionfrom
thesame stocksolu¬tions (1mg/mL).
The internal standard-mix solution (1 pg/mL
of
each amphetamine) was prepared bydilution of
lOplfrom
eachdj-deuterated analogue in 10
mL
distilled water.All
standards were stored at4°C and were allowed to cometoroomtemperature, vortex-mixed and centrifu- gedprior
to analysis.Sample preparation
Sample preparation was minimal and consisted
of
adding 90 pi
of
theinternal standard-mix solution to 10pi
of
sample(calibrationstandards andstandardsmade in H20/MeOH (50/50) solution).After
vortex-mixing and centrifugation (2 min at 13000g), 85 piof
the sample solution was pipetted into crimp-cap autosam¬pler vials and placedinthe autosampler.
Validation experiment
Method validation, including studies
of
imprecision (within-day and between-day), accuracy, linearity, sta¬bility,
carry-over, recovery and thedeterminationof
thelimit of
detection (LOD) and quantification (LOQ)of
the
LC-MS/MS
method was performed according to theFDA
recommendations (8).Results and discussion
Figure
I
shows atypicalLC-MS/MS
chromatogramof
apatient urine sample containing 11400 ng/mL AMP, 5633 ng/mL
MDMA
and 281 ng/mLMDA.
AnnalesdeToxicologie Analytique,vol.
XVII,
n° 1,2005pg/mL) used as internai standards (IS) in methanol were obtained
from Cerilliant (Austin,
Texas).Methanol (absolute) and water
for LC-MS
were pur¬chased
from Biosolve
(Valkenswaard, The Netherlands).Ammonium
acetate (p.a) was supplied by Sigma-Aldrich (Bornem, Belgium).Instrumentation and MS/MS conditions
An Agilent
1100 seriesHPLC
system(Agilent
Technologies) consistingof
a pump, column oven, autosampler and degasser wereusedfor
solventdelive¬ry and sample introduction. The injected volume was 20 pi. Analytes were separated at40°C on a 2.1 x 30 mm Zorbax SB
-CI
8. Rapid Solution column(Agilent
Technologies).Thecolumnwaseluted ataflow
rateof
0.3
mL/min
anddevelopedwith
gradientelutionasfol¬lows:0-0.2min,95%A/5%B; 1.2-4.5min,5%A/95%B and4.8-8 min 95%A/5%B
(A:
H20 +2mM
ammoniu- macetate, B:MeOH
+2mM
ammoniumacetate).The
LC-MS/MS
system consistedof
anAPI 2000
triple-quadrupolemass spectrometer equippedwith
anAPCI
interface(Applied
Biosystems/MDS Sciex, Langen, Germany) usedin the positive-ion mode. The sixamphetamine compounds were detectedinthemul¬tiple-reaction monitoringmode. Two
MRM
transitionsfor
eachsubstanceweremonitoredtoprovidesufficient identificationof
the amphetamine compounds. The chosenMRM
transitionsfor
each amphetamine com¬pound and d5-deuterated analogue are summarised in table 1. The entrance potential varied
from
5.5V
toTable
I
:Retentiontime,parentionandthechosendaughter ionsfor
each amphetamine compound and d5-deuterated analogueAmphétaiijinesRetention
M+l
MRM1 MRM2 , nine(min) ;:AMP 4.20 136.079 91.05 65.05
MET 4.36 150.078 91.05 119.15
MDA 4.29 180.119 135.15 133.05
MDMA 4.38 194.085 163.05 105.05
MDEA 4.50 208.068 163.05 105.15
MBDB 4.61 208.068 135.05 177.15
d5-AMP 4.17 141.112 93.35
ds-MET 4.35 155.09 92.35
ds-MDA 4.26 185.114 110.15 ds-MDMA 4.35 199.136 165.15 ds-MDBA 4.48 213.075 163.05
*-MBDB 4.60 199.136 165.15
9 V,thecollision cellentrancepotentialvaried
from
14 to 20V
andthecell exit potentialwas setat2V
or4V, accordingtothe analyte.Analyst
Software (Ver. 1.3.1;Applied
Biosystems/MDS Sciex) was usedfor
HPLC system control,dataacquisition, anddataprocessing.Calibration standards and internal stan¬
dard mix-solution
Calibration standards were prepared
in
drug-freeurinefrom
methanolicstocksolutions,containingallamphe¬tamine compounds ataconcentration
of
1 mg/mL. The concentrationsof
thecalibration
standards were 78.125, 156.25, 312.5, 625, 1250,5000, 10000, 20000, 30000,40000, 50000, 60000, 70000, 80000, 90000and 100000ng/mL.Forrecovery testing, standards
with
concentrationsof
78.125, 625 and 10000ng/mL
were made in H20/MeOH (50/50) solutionfrom
thesame stocksolu¬tions (1mg/mL).
The internal standard-mix solution (1 pg/mL
of
each amphetamine) was prepared bydilution of
lOplfrom
eachdj-deuterated analogue in 10
mL
distilled water.All
standards were stored at4°C and were allowed to cometoroomtemperature, vortex-mixed and centrifu- gedprior
to analysis.Sample preparation
Sample preparation was minimal and consisted
of
adding 90 pi
of
theinternal standard-mix solution to 10pi
of
sample(calibrationstandards andstandardsmade in H20/MeOH (50/50) solution).After
vortex-mixing and centrifugation (2 min at 13000g), 85 piof
the sample solution was pipetted into crimp-cap autosam¬pler vials and placedinthe autosampler.
Validation experiment
Method validation, including studies
of
imprecision (within-day and between-day), accuracy, linearity, sta¬bility,
carry-over, recovery and thedeterminationof
thelimit of
detection (LOD) and quantification (LOQ)of
the
LC-MS/MS
method was performed according to theFDA
recommendations (8).Results and discussion
Figure
I
shows atypicalLC-MS/MS
chromatogramof
apatient urine sample containing 11400 ng/mL AMP, 5633 ng/mL
MDMA
and 281 ng/mLMDA.
w«^*c5S»î^^w?w^ «WÇ^ôw&ww&wwt&S.w»
le
XICof+MRM(28pairs):136.1/91.1amufrom Sample1(Sample chai) of 0B0405.wiff (Heated Nebulizer) 4.33
I
Max.3.0e6cps.
3.0e8 2.8e6 2.ee8 2.4eB 2.2e8
2.0eS 1.8e8 1.6eB
1.4e8
1.2e8 1.0e8 8.0e5
6.0e5 4.0e5 2.0e5
0.0
5.8 8.0!
KWSSSWWSS
Figure1:LC-MS/MS chromatogram
of
apatienturinesamplecontainingAMP(11400ng/mL),MDMA(5633ng/mL)andMDA (281ng/mL). Theretention times (RT)of
theinternalstandardsare4.33min (d5-AMP), 4.43min(dyMET), 4.38min (d5-MDA), 4.43min(dyMDMA),4.54min (d5-MDEA), and 4.63 min(dyMBDB).Calibration curves
To construct calibration curves, a set
of
eight urine samples spikedwith
the amphetamine compounds at concentrationsrangingfrom
78 ng/mLto 10000ng/mL were used. The calibrators were measuredfor
five consecutive days.All
calibration curves showedlinea¬rity for
all amphetamine compounds and correlation coefficientsexceeded0,998.LOD and LOQ
The
limit of
detection(LOD),
defined as a signal-to- noise ratioof
3, and thelimit of
quantification (LOQ), defined asasignal-to-noiseratioof
10, werecalculated by ascriptintheAnalystSoftware. TheLOD
andLOQ
were 4.9 ng/mL and 9.8 ng/mL, respectivelyfor
each analyte exceptfor
amphetamine andMDA
(table H).The LOQ's are lower than the current recommended urine confirmation
cut-off
levels (9). Howeverif for
certain applications higher sensitivity is necessary some modifications can be tried out: increasing the injected volume, increasing the sample/internal stan¬dardratio,loweringthenumber
of
MRM-transitions or anotherprocedure (e.g.with
extraction)canbeapplied.Table
II
:Limitof
detectionand limitof
quantificationfor
eachamphetamine compound.
Amphétamines AMP MET MDA MDMA
MDEA MBDB
LOD(ng/mL) 39.1
4.9 39.1 4.9 4.9 4.9
LOQ(ng/mL) 78.1
9.8 78.1
9.8 9.8 9.8
Imprecision
Imprecision (CV%) was evaluated by analysing three calibrator samples
with
a low (78 ng/mL), medium (625 ng/mL)andhigh(10000 ng/mL) concentrationof
each amphetamine on the same day
in
five replicates (within-day imprecision) and over five consecutive days (between-day reproducibility). The within-day CVs rangedfrom
2.62 to 16.26%, the between-day CVsfrom
0.86 to 11.98% (table HT). So, datafor
imprecision werewithin
requiredlimits of
20% at thew«^*c5S»î^^w?w^ «WÇ^ôw&ww&wwt&S.w»
le
XICof+MRM(28pairs):136.1/91.1amufrom Sample1(Sample chai) of 0B0405.wiff (Heated Nebulizer) 4.33
I
Max.3.0e6cps.
3.0e8 2.8e6 2.ee8 2.4eB 2.2e8
2.0eS 1.8e8 1.6eB
1.4e8
1.2e8 1.0e8 8.0e5
6.0e5 4.0e5 2.0e5
0.0
5.8 8.0!
KWSSSWWSS
Figure1:LC-MS/MS chromatogram
of
apatienturinesamplecontainingAMP(11400ng/mL),MDMA(5633ng/mL)andMDA (281ng/mL). Theretention times (RT)of
theinternalstandardsare4.33min (d5-AMP), 4.43min(dyMET), 4.38min (d5-MDA), 4.43min(dyMDMA),4.54min (d5-MDEA), and 4.63 min(dyMBDB).Calibration curves
To construct calibration curves, a set
of
eight urine samples spikedwith
the amphetamine compounds at concentrationsrangingfrom
78 ng/mLto 10000ng/mL were used. The calibrators were measuredfor
five consecutive days.All
calibration curves showedlinea¬rity for
all amphetamine compounds and correlation coefficientsexceeded0,998.LOD and LOQ
The
limit of
detection(LOD),
defined as a signal-to- noise ratioof
3, and thelimit of
quantification (LOQ), defined asasignal-to-noiseratioof
10, werecalculated by ascriptintheAnalystSoftware. TheLOD
andLOQ
were 4.9 ng/mL and 9.8 ng/mL, respectivelyfor
each analyte exceptfor
amphetamine andMDA
(table H).The LOQ's are lower than the current recommended urine confirmation
cut-off
levels (9). Howeverif for
certain applications higher sensitivity is necessary some modifications can be tried out: increasing the injected volume, increasing the sample/internal stan¬dardratio,loweringthenumber
of
MRM-transitions or anotherprocedure (e.g.with
extraction)canbeapplied.Table
II
:Limitof
detectionand limitof
quantificationfor
eachamphetamine compound.
Amphétamines AMP MET MDA MDMA
MDEA MBDB
LOD(ng/mL) 39.1
4.9 39.1 4.9 4.9 4.9
LOQ(ng/mL) 78.1
9.8 78.1
9.8 9.8 9.8
Imprecision
Imprecision (CV%) was evaluated by analysing three calibrator samples
with
a low (78 ng/mL), medium (625 ng/mL)andhigh(10000 ng/mL) concentrationof
each amphetamine on the same day
in
five replicates (within-day imprecision) and over five consecutive days (between-day reproducibility). The within-day CVs rangedfrom
2.62 to 16.26%, the between-day CVsfrom
0.86 to 11.98% (table HT). So, datafor
imprecision werewithin
requiredlimits of
20% at theAnnalesdeToxicologieAnalytique, vol.
XVII,
n° 1,2005Tableau
III:
Imprecision(CV%) bias (%) arid recovery (%) determinedfor
threecalibrationstandardswith low (78ng/mL) medium(625ng/mL)andhigh(10000ng/mL)concentrationof
eachamphetaminecompound, respectively(n=5).AMP MET MDA MDMA
MDEA MBDB
O- : V;-
y Impr^
\Vithin-dày CV
78 ng/mL
16.26 3.39 3.42 6.80 8.17 10.76
625 ng/mL
4.76 2.94 4.60 2.77 3.53 5.68
10000 ng/mL 3.86 3.35 3.41 4.42 5.36 2.62
Bètwêén-dayCVV 78
ng/mL
7.07 3.83 10.42 11.98 10.98 2.40625
ng/mL
4.39 4.14 3.05 3.57 3.91 2.39
10000
ng/mL
0.86 1.29 1.89 1.61 1.29 2.20Accuracy
(b78
ng/mL
2.5 2.4 7.7 4.1 3.0 1.1625
ng/mL
1.0 0.5 1.3 0.2 0.2 3.4
ias%)
10000 ng/mL 1.6 2.4 0.2 1.4 0.8 0.4
Recovery
(%)
78
ng/mL
-61 110
94 108 101 102
625 ng/mL
78 109
93 108 102 105
10000 ng/mL
70 99 75 95 92 98
lowestconcentrationandbelow 15% athigherconcen¬
trations.
Accuracy
The accuracy
of
this methodfor
each amphetamine compound was obtained by analyzing the same three calibration standards as mentioned in the paragraph 'imprecision' over five consecutivedays. As indicated in tablein,
the calculated concentrationof
each com¬pound agreed
well with
the expected values.Recovery
The recoveries were obtained by comparing the peak areas
of
spikedurinewith
thoseof
the same concentra¬tions
of
the analytes inH20/MeOH
(50/50) solution.Three concentrations were tested (78 ng/mL; 625 ng/mL; 10000ng/mL) in five-fold.The results arepre¬
sented in table
m.
We observed good agreement (< 15%deviation)for
most analytes, exceptfor
amphe¬tamine(all concentrations) and
MDA
(onlythehighest concentration).Carry-over
Carry-over was evaluated by injecting a blank urine specimencontainingthe internal standardsimmediate¬
ly after asample that contained 10 000ng/mL
of
each amphetamine compound. Carry-over was less than 0.32% and the results are shown intable IV. Although the carry-over is low, the confirmationcut-off of
200 ng/mL could be reached after a sample containing 60000ng/mL of
an amphetamine, which occurs occa¬sionally.
TableTV: Carry-over(%)inablankurine sampleanalyzed, aftera calibration standardwitha concentration
of
10000 ng/mLAmphétamines AMP MET MDA MDMA
MDEA MBDB
Carry-over(%)
0 0.22 0.06 0.22 0.32 0.26
Linearity above 10000 ng/mL
Standards with concentrations between 20000 ng/mL and 100000 ng/mL, made in drug-free urine samples, wereused to determine linearity above 10000 ng/mL.
The linearity was evaluated by
dividing
the observed valueof
each standard by the expected valueof
each standard to determine the percentageof
the expected resultfor
each concentration. The percentagesof
the expected resultsfor
the amphetamines were betv/een 91%and 107% (tableV).
Stability
Forstability studies,two calibration standards(calibra¬
tionstandard 2with aconcentration
of
156 ng/mLand calibrator7with aconcentrationof
5000ng/mL) were each split into 10 aliquots, with five aliquots assayed immediately and the otherfivestoredfor
upto 24 h at room temperature. The meansof
the five determina¬tions
for
each calibrator, beforeand after storagewere thencompared. The dataaregiven intableVI.
AnnalesdeToxicologieAnalytique, vol.
XVII,
n° 1,2005Tableau
III:
Imprecision(CV%) bias (%) arid recovery (%) determinedfor
threecalibrationstandardswith low (78ng/mL) medium(625ng/mL)andhigh(10000ng/mL)concentrationof
eachamphetaminecompound, respectively(n=5).AMP MET MDA MDMA
MDEA MBDB
O- : V;-
y Impr^
\Vithin-dày CV
78 ng/mL
16.26 3.39 3.42 6.80 8.17 10.76
625 ng/mL
4.76 2.94 4.60 2.77 3.53 5.68
10000 ng/mL 3.86 3.35 3.41 4.42 5.36 2.62
Bètwêén-dayCVV 78
ng/mL
7.07 3.83 10.42 11.98 10.98 2.40625
ng/mL
4.39 4.14 3.05 3.57 3.91 2.39
10000
ng/mL
0.86 1.29 1.89 1.61 1.29 2.20Accuracy
(b78
ng/mL
2.5 2.4 7.7 4.1 3.0 1.1625
ng/mL
1.0 0.5 1.3 0.2 0.2 3.4
ias%)
10000 ng/mL 1.6 2.4 0.2 1.4 0.8 0.4
Recovery
(%)
78
ng/mL
-61 110
94 108 101 102
625 ng/mL
78 109
93 108 102 105
10000 ng/mL
70 99 75 95 92 98
lowestconcentrationandbelow 15% athigherconcen¬
trations.
Accuracy
The accuracy
of
this methodfor
each amphetamine compound was obtained by analyzing the same three calibration standards as mentioned in the paragraph 'imprecision' over five consecutivedays. As indicated in tablein,
the calculated concentrationof
each com¬pound agreed
well with
the expected values.Recovery
The recoveries were obtained by comparing the peak areas
of
spikedurinewith
thoseof
the same concentra¬tions
of
the analytes inH20/MeOH
(50/50) solution.Three concentrations were tested (78 ng/mL; 625 ng/mL; 10000ng/mL) in five-fold.The results arepre¬
sented in table
m.
We observed good agreement (< 15%deviation)for
most analytes, exceptfor
amphe¬tamine(all concentrations) and
MDA
(onlythehighest concentration).Carry-over
Carry-over was evaluated by injecting a blank urine specimencontainingthe internal standardsimmediate¬
ly after asample that contained 10 000ng/mL
of
each amphetamine compound. Carry-over was less than 0.32% and the results are shown intable IV. Although the carry-over is low, the confirmationcut-off of
200 ng/mL could be reached after a sample containing 60000ng/mL of
an amphetamine, which occurs occa¬sionally.
TableTV: Carry-over(%)inablankurine sampleanalyzed, aftera calibration standardwitha concentration
of
10000 ng/mLAmphétamines AMP MET MDA MDMA
MDEA MBDB
Carry-over(%)
0 0.22 0.06 0.22 0.32 0.26
Linearity above 10000 ng/mL
Standards with concentrations between 20000 ng/mL and 100000 ng/mL, made in drug-free urine samples, wereused to determine linearity above 10000 ng/mL.
The linearity was evaluated by
dividing
the observed valueof
each standard by the expected valueof
each standard to determine the percentageof
the expected resultfor
each concentration. The percentagesof
the expected resultsfor
the amphetamines were betv/een 91%and 107% (tableV).
Stability
Forstability studies,two calibration standards(calibra¬
tionstandard 2with aconcentration
of
156 ng/mLand calibrator7with aconcentrationof
5000ng/mL) were each split into 10 aliquots, with five aliquots assayed immediately and the otherfivestoredfor
upto 24 h at room temperature. The meansof
the five determina¬tions
for
each calibrator, beforeand after storagewere thencompared. The dataaregiven intableVI.
TableauV:Linearityobtained bydividingtheobservedvalue
of
eachstandardby the expectedvalueof
eachstandardandmul¬tipliedby 100. Thedeviations(%)werebelow10%.
Amphetamine
AMP
METMDA MDMA
MDEA MBDB
20000 ng/mL 107 107 91.8 104 109 107
Linearity
30000ng/mL
104 106 95 106 101 105
asdeviatk 40000
ng/mL
103 104 96 103 106 99.2
m (%)
of I
50000 hg/mL 103 106 103 106 107 104lie
observ60000;
ng/mL
101 104 102 99.7 99.1 104
edvaluet(
70000
ng/mL
98,7100 104 102 101 101
5theexpeç 80000 ^
Tig/mL 101 100 105 102 99.7 102
:tedvalue 90000
L
ng/iûL
97,2 97.4 104 100 100 102100000
ng/mL
92,3 95.4 97.4 94 94.8 92.7
Table VI : Stability of two calibration standards with a concentration of156ng/mL and 5000 ng/mL, respectively aftera24hstorageatroom temperature. Themeansoffive determinationsfor eachcompound, beforeandafterstorage were subtractedanddividedby the meanofthecorrespon¬
ding results obtained beforestorage.
Références
îphétamine
AMP MET MDA MDMA
MDEA
Stability(%)
156ng/mL 6.8 1.43
-3.31 -3.45 -11.99
5000ng/mL -1.78 -0.26 -14.66
0.07 -5.04
MBDB -15.87 -13.57
Conclusion
We have developed and validated a
LC-MS/MS
methodfor
the simultaneous determinationof
six amphetaminecompoundsin urinesamples.The sample pre-treatmentis fast and simple, requiring no derivati¬sation. TheLOQs are much lower than recommended urine confirmation
cut-off
levels, i.e. this method is sensitive enoughfor
routine confirmation. Accuracy andimprecisionfulfil
thecriteriaof
<20%ataconcen¬tration equal to the
LOQ
and < 15% athigherconcen¬trations. Good recoveriesandlinearityoverawideana¬
lytical range were obtained. Carry-over is minimal.
Separation anddetection
of
allcompoundswas accom¬plished
within
eight minutes. The main advantagesof
thepresentmethodlie initssimple sample preparation, reliable results and short analysistime.
Acknowledgments
Wethank Fien Vander Heyden
for
assistanceduringthe developmentof
this method.1. European Monitoring centre for drugs and drug addic¬
tion. Annualreport 2004: the stateofthedrugsproblem in the European Union and Norway. Luxembourg:
Bureau for the official publications of the European Union,2004; 1-113.
2. United nations office on drugs and crime. 2004 World Drug report. United Nations. Geneva: United Nations Publications, 2004; 1-427.
3. KraemerT., MaurerH.H. Determination ofamphetami¬
ne, methamphetamine and amphetamine-derived desi¬
gner drugs or medicaments in blood and urine. J.
Chromatogr.BBiomed. Sci. Appl. 1998 ;713 : 163-87.
4. Jurado C, Gimenez M.P., Soriano T., Menendez M., RepettoM.Rapid analysisofamphetamine,methamphe¬
tamine, MDA, and MDMA in urine using solid-phase microextraction,directon-fiberderivatization,and analy¬
sisbyGC-MS.J.Anal. Toxicol. 2000;24 : 11-6.
5. Marquet P. Progress
of
liquid chromatography-mass spectrometry in clinical and forensic toxicology. Ther.DrugMonit.2002; 24: 255-76.
6. Marquet P., Lachatre G. Liquid chromatography-mass spectrometry: potentialinforensicandclinicaltoxicolo¬
gy.J.Chromatogr.BBiomed. Sci.Appl. 1999 ;733:93- 118.
7. Nordgren H.K., Beck O. Direct screening of urine for MDMA and MDA by liquid chromatography-tandem massspectrometry.J.Anal. Toxicol. 2003 ;27 : 15-9.
8. USDepartmentofHealthandHuman Services Foodand Drug Administration - Center for Drug Evaluation and Research (CDER). GuidanceforIndustry, Bioanalytical Method Validation. 2001 ; 1-25
9. Substance abuse and mental health services administra¬
tion Proposed revisions to mandatory guidelinesforfede¬
ral workplace drug testing programs. Federal Register 2004; 69 : 19673-732.
TableauV:Linearityobtained bydividingtheobservedvalue
of
eachstandardby the expectedvalueof
eachstandardandmul¬tipliedby 100. Thedeviations(%)werebelow10%.
Amphetamine
AMP
METMDA MDMA
MDEA MBDB
20000 ng/mL 107 107 91.8 104 109 107
Linearity
30000ng/mL
104 106 95 106 101 105
asdeviatk 40000
ng/mL
103 104 96 103 106 99.2
m (%)
of I
50000 hg/mL 103 106 103 106 107 104lie
observ60000;
ng/mL
101 104 102 99.7 99.1 104
edvaluet(
70000
ng/mL
98,7100 104 102 101 101
5theexpeç 80000 ^
Tig/mL 101 100 105 102 99.7 102
:tedvalue 90000
L
ng/iûL
97,2 97.4 104 100 100 102100000
ng/mL
92,3 95.4 97.4 94 94.8 92.7
Table VI : Stability of two calibration standards with a concentration of156ng/mL and 5000 ng/mL, respectively aftera24hstorageatroom temperature. Themeansoffive determinationsfor eachcompound, beforeandafterstorage were subtractedanddividedby the meanofthecorrespon¬
ding results obtained beforestorage.
Références
îphétamine
AMP MET MDA MDMA
MDEA
Stability(%)
156ng/mL 6.8 1.43
-3.31 -3.45 -11.99
5000ng/mL -1.78 -0.26 -14.66
0.07 -5.04
MBDB -15.87 -13.57
Conclusion
We have developed and validated a
LC-MS/MS
methodfor
the simultaneous determinationof
six amphetaminecompoundsin urinesamples.The sample pre-treatmentis fast and simple, requiring no derivati¬sation. TheLOQs are much lower than recommended urine confirmation
cut-off
levels, i.e. this method is sensitive enoughfor
routine confirmation. Accuracy andimprecisionfulfil
thecriteriaof
<20%ataconcen¬tration equal to the
LOQ
and < 15% athigherconcen¬trations. Good recoveriesandlinearityoverawideana¬
lytical range were obtained. Carry-over is minimal.
Separation anddetection
of
allcompoundswas accom¬plished
within
eight minutes. The main advantagesof
thepresentmethodlie initssimple sample preparation, reliable results and short analysistime.
Acknowledgments
Wethank Fien Vander Heyden
for
assistanceduringthe developmentof
this method.1. European Monitoring centre for drugs and drug addic¬
tion. Annualreport 2004: the stateofthedrugsproblem in the European Union and Norway. Luxembourg:
Bureau for the official publications of the European Union,2004; 1-113.
2. United nations office on drugs and crime. 2004 World Drug report. United Nations. Geneva: United Nations Publications, 2004; 1-427.
3. KraemerT., MaurerH.H. Determination ofamphetami¬
ne, methamphetamine and amphetamine-derived desi¬
gner drugs or medicaments in blood and urine. J.
Chromatogr.BBiomed. Sci. Appl. 1998 ;713 : 163-87.
4. Jurado C, Gimenez M.P., Soriano T., Menendez M., RepettoM.Rapid analysisofamphetamine,methamphe¬
tamine, MDA, and MDMA in urine using solid-phase microextraction,directon-fiberderivatization,and analy¬
sisbyGC-MS.J.Anal. Toxicol. 2000;24 : 11-6.
5. Marquet P. Progress
of
liquid chromatography-mass spectrometry in clinical and forensic toxicology. Ther.DrugMonit.2002; 24: 255-76.
6. Marquet P., Lachatre G. Liquid chromatography-mass spectrometry: potentialinforensicandclinicaltoxicolo¬
gy.J.Chromatogr.BBiomed. Sci.Appl. 1999 ;733:93- 118.
7. Nordgren H.K., Beck O. Direct screening of urine for MDMA and MDA by liquid chromatography-tandem massspectrometry.J.Anal. Toxicol. 2003 ;27 : 15-9.
8. USDepartmentofHealthandHuman Services Foodand Drug Administration - Center for Drug Evaluation and Research (CDER). GuidanceforIndustry, Bioanalytical Method Validation. 2001 ; 1-25
9. Substance abuse and mental health services administra¬
tion Proposed revisions to mandatory guidelinesforfede¬
ral workplace drug testing programs. Federal Register 2004; 69 : 19673-732.