BE CAREFUL WHEN WORKING IN MRM MODE

POTENTIAL FALSE POSITIVE RESULTS FOR 2-AMINOINDANE

M. DEVILLE

1

,N. DUBOIS

1

, C.CHARLIER

1

1. Laboratory of Clinical, Forensic, and Environmental Toxicology, Center for Interdisciplinary Research on Medicines (CIRM), University of Liege, Belgium

INTRODUCTION

UHPLC-MS-MS (Waters)

UPLC Acquity + Quattro Premier Column: BEH C18, 1.7 μm, 50 X 2.1 mm (Waters)

Mobile Phase A : 0.1% formic acid in water Mobile Phase B: 0.1% formic acid in methanol

Gradient mode: 15 min runtime

UHPLC-TOF-MS (Sciex)

Eksigent LC 100 XL + TripleTOF 4600

Column: Kinetex 2.6 C18, 100 Å, 50 x 3.00 mm (Phenomenex) Mobile Phase A : 10 mM NH4 formate

Mobile Phase B: ACN/MeOH (50/50) Gradient mode: 15.5 min runtime

MATERIAL AND METHODS

RESULTS

DISCUSSION

Conclusion

A patient came for the third time at the emergency department for ear-nose-throat symptoms, and finally died of a virulent group A Streptococcus pyogenes. As chemsex was evocated by the family, a screening for new psychoactive substances was undertaken.

Na2CO3 Liquid/liquid Extraction

IS = Prazepam

Even if MRM mode is often preferred in order to increase the sensitivity, identification criteria - including retention time and two transitions - are sometimes insufficient to be 100% sure of the compound identification, especially if the runtime is short. Calculating ion ratios is indispensable and not optional if identification is to be based on MRM data.

Keep in

mind

TOF MRM 2 1

All criteria are met for 2-aminoindane (2-AI) but:

- Unknown retention time - Parent ion

• present in the sample • absent in the library

2AI MS Spectrum in the sample

2AI MS spectrum in the library

UHPLC-TOF-MS: after reception of reference standard

Neat standard vs extracted serum:  0.3 min time shift

 Matrix effect?

 Sample was spiked with 2-AI

2 peaks

Perfect match

UHPLC-TOF-MS : before reception of reference standard

UHPLC-MS-MS: after reception of reference standard

Neat standard vs extracted serum:

 0.1 min time shift

BUT

Ion Ratio 133.9 > 116.9 / 133.9 > 115.1 Standard : 6.62 ≠ Sample : 2.37

 Knowledge of the retention time is usefull to identify a compound but reference standard is essential to do so  Delivery time are sometimes long  potential stability issue if a case is open

 Some matrix effects can generate a shift in the retention time

 LC-MSMS method developped to be as fast as possible: retention time shift even smaller  risk of false positive result if ions ratios aren’t checked

 Same interference in 6 out of 900 other samples but origin still unknown

2-Aminoindane

NPS – stimulant effect Without reference standard

After reception of reference standard 1

Références

Documents relatifs

It includes patients known to have died, stopped treatment for medical reasons and registered as so in the clinic, in addition to patients LTFU whose outcome is unknown..

This revised version of the manuscript "How robust are cross-population signatures of polygenic adaptation in humans?" addressed, in my opinion, all points raised by

As the characteristics of national health delivery systems are fixed by legislation, those involved in such legislation should increasingly recognize that the promotional,

We say that (s, τ) is decorated if each vertex v of s is associated with an elementary star F v , and a “gluing” application that maps the half-edges of s adjacent to v to the

In this article, extending previous work on topological sensitivity, a methodology for expanding to order O(ε 4 ) a generic cost function under the nucleation of a small inclusion

As we are concerned about investigating fairness, it makes sense to think of two types of transparency corresponding to those for fairness: process transparency (how much we

Figure 2: Illustrating the interplay between top-down and bottom-up modeling As illustrated in Figure 3 process modelling with focus on a best practice top-down model, as well

New Brunswick is increasing its training of medi- cal students with the opening of the Saint John Campus of Dalhousie University’s medical school.. At the same time,