A Rapid and Selective MS/MS Method for quantification of light sensitive chlortetracycline drug in wastewater and wastewater sludge samples using Laser Diode Thermal Desorption (LDTD) technique.

N/A
N/A
Protected

Academic year: 2021

Partager "A Rapid and Selective MS/MS Method for quantification of light sensitive chlortetracycline drug in wastewater and wastewater sludge samples using Laser Diode Thermal Desorption (LDTD) technique."

Copied!
1
0
0

Texte intégral

(1)

INTRODUCTION

METHOD

RESULTS

DISCUSSION AND CONCLUSIONS

OVERVIEW

Purpose:

 The development of an accurate LDTD-MS/MS method for a light sensitive pharmaceutical ligand is challenging

 LDTD represents a technological revolution that omits the chromatographic step and significantly increases the analytical throughput for the quantitation of chlortetracycline (CTC)

Method:

 Chlortetracycline spiked in wastewater (WW) and wastewater sludge (WWS) samples

 Solid phase extraction (SPE) of wastewater samples using C18 cartridges

 Calibration range: 5 – 200 µg/L  LDTD-APCI-MS/MS analysis

Chlortetracycline antibiotic is a potential WW contaminant. It is readily soluble in water. Its characteristic complexion property with metals makes it difficult to retain it on cartridge during clean up process and also desorption of CTC from LazWellTM plate (96-well plate). To overcome this problem, Na2H2EDTA was added to WW sample to make CTC free from metal complex and to improve

chlortetracycline desorption from plate. A MS interface LDTD combined with atmospheric

pressure chemical ionization (APCI), has shown great potential to reduce analysis time from minutes to seconds by removing the chromatographic step and introducing charged analytes directly into the mass spectrometer.

Figure 1: Schematics of Path of ionization of sample in LDTD

Sample Preparation

- Centrifugation of sludge sample - Lyophilization of solid residue - Microwave extraction of

lyophilized solid

Clean up

- Adjust the pH of sample to 3.5 – 4.0

- Add Na2H2EDTA to the sample to make CTC free from metal complex

- SPE of sample using C18 cartridges and methanol as eluent

- Dry the extract under a gentle stream of N2

- Dried residue was reconstituted with 1 mL of

methanol/water mixture (75:25, v/v) before

analysis

Instrumentation:

- LDTD S-960, Phytronix Technologies - API 5500 QTrap , AB SCIEX

LDTD Parameters

- Laser power pattern :

The laser power is ramped from 0 to 65% in 6.0 s

and hold at 65% for 2.0 s

- Carrier gas (Air) temperature -- 30 oC and flow rate –3 L/min

- The optimum sample volume deposited into a well

is

4 µL

Mass Spectrometry parameters

- Ion source --- APCI (+)

- Mode --- MRM

- Corona discharge current --- 3 µA

R² = 0.9989

y = 0.002x+0.00363

Area ratio = Area/IS area

0 0.05 0.1 0.15 0.2 0.25 0.3 0.35 0.4 0.45 0 25 50 75 100 125 150 175 200 Concentration ratio Area rati o Compound Precursor ion Mass, [M + H]+, (m/z) Product ion Mass Collision energy (volts) CTC 479 211 45.00 Clomiphene 406 100 35.00

Sample ID R2 Intra-day precision

(%RSD) Inter-day precision (%RSD) IDL (µg/L) MDL (µg/L) LOD LOQ Wastewater 0.998 4.22 7.53 1.67 5.13 17.54 Wastewater sludge 0.998 9.18 12.39 2.49 7.49 25.92 Sample ID Concentration ±%RSD (n=3) (µg/L) Spike Recovery Sample ID Concentration ±%RSD (n=3) (µg/g) Spike Recovery

Influent 61.46±3.84 92.46 Primary sludge 0.804±4.9 85.32

Screened

Sample 57.63±2.91 97.56 Secondary sludge 0.638±6.7 90.42

Effluent 7.28±1.71 95.81 Mixed sludge 0.469±5.4 87.32

A simple, rapid and sensitive method was developed for the quantification of CTC in wastewater samples. The applicability of this method in environmental samples was confirmed by analyzing CTC in WW and WWS and its recovery values. Since there is no chromatographic step involved in the analysis process, the analysis time was reduced from minutes to seconds compared to traditional LC-MS/MS system.

Instrumental parameters were optimized to give the maximal signal for CTC. Laser power and sample volume spotted were optimized and combined with appropriate MRM conditions for accurate quantification. The ultra-fast analysis time of 10 seconds/sample, combined with the high specificity power of mass spectra instrument makes this method suitable alternative for rapid environmental screening of emerging contaminants.

Table 2: Chlortetracycline method validation parameters

Table 3: Concentrations of chlortetracycline in wastewater samples collected from Quebec Urban Community (QUC) WWTP (n = 3)

Table 1: Precursor and product ions and collision energy (CE) for the pharmaceutical compound CTC

OH O OH O O NH2 OH CH3 Cl N C H3 CH3 OH OH H H Cl H

Figure 4: LDTD intensity peaks for CTC

calibration curve (multiple sample desorption in one file)

Figure 3: Calibration curve of CTC

A Rapid and Selective MS/MS Method for quantification of light sensitive chlortetracycline drug

in wastewater and wastewater sludge samples using Laser Diode Thermal Desorption (LDTD)

technique

Rama Pulicharla

1

, S. Auger

2

, S.K. Brar

1

, P. Drogui

1

, R.Y. Surampalli

3

1

INRS-ETE, Québec, Canada;

2

Phytronix Technologies, Québec, Canada;

3

University of Nebraska-Lincoln, Lincoln; USA

Figure 2: CTC chemical structure

1) Laser continuously illuminated on the back of the bottom surface of the well plate (LazWell).

2) LazWell plate is designed to hold only a sample of 1-10uL and for maximum heat transfer for quick sample evaporation.

3) Transfer tube to deliver vaporized samples without loss. 4) Sample ionization by corona discharge at the transfer

tube exit.

62nd ASMS Conference on Mass Spectrometry

and Allied Topics. June 15 - 19, 2014.

Baltimore Convention Center, Baltimore, MD 21201, USA.

Références

Documents relatifs

Purification by flash chromatography (hexane:ether =.. The 1H NMR spectrum matched the published spectrum.. Absolute Configuration Assignment for Table 2, entry 3:

Comparabilité, composant relationnel, état relationnel, imprécision, non ordre, pseudo-ordre, pouvoir discriminant, quasi-ordre, preference, rassembleur, relation,

En novembre 1969, en exposant le projet universitaire dahoméen à la mission française dirigée par le recteur Renaud Paulian, Jean Roger Ahoyo, porte-parole de la commission

The fact that the generation rates can be favorably compared as a function of liquid zinc velocity validated the use of the stirred crucible experiments in

In Maspero-Robert [31], the authors study linear Schr¨ odinger operators, obtain global well- posedness results and prove very precise polynomial bounds on the possible growth

J’ai pu grâce à lui passer d’un sujet vaste, le « neuromarketing », à l’étude plus spécifique des données émotionnelles de la start-up Datakalab et à la manière dont

Il est donc possible que nous ayons inclus dans ce groupe des patients gros consommateurs de cannabis, ce qui pourrait expliquer cette concentration élevée.. Ce biais

Starting from the above considerations, this Special Issue aims to report the latest advances and trends concerning the advanced machine learning techniques in regard to time